| Literature DB >> 25481296 |
Yukihide Minoda1, Shigeru Nakagawa2, Ryo Sugama3, Tessyu Ikawa4, Takahiro Noguchi5, Masashi Hirakawa6.
Abstract
The relationship between the joint gap before and after implantation in 259 knees during the total knee arthroplasty was investigated using a tensor device which can attach the polyethylene insert trial. Patients were divided into following 3 groups according to the joint gap balance before implantation (flexion joint gap--extension joint gap); group 1: >1mm; group 2: -1 to 1mm, and group 3: <-1mm. Joint gap after implantation was loose at 30°, 60°, 90°, and 120° of flexion in group 1 and 2, but loose only at 30° of flexion in group 3 (p<0.01). This study showed that loose flexion joint gap before implantation increased the risk of joint gap laxity after implantation especially at midflexion ranges.Entities:
Keywords: joint gap after implantation; joint gap before implantation; midflexion laxity; tensor device; total knee arthroplasty
Mesh:
Substances:
Year: 2014 PMID: 25481296 DOI: 10.1016/j.arth.2014.11.011
Source DB: PubMed Journal: J Arthroplasty ISSN: 0883-5403 Impact factor: 4.757