| Literature DB >> 25478180 |
David Hannon1, Judy M Bradley2, Ian Bradbury2, Nicholas Bell3, J Stuart Elborn1, Katherine O'Neill1.
Abstract
BACKGROUND: Lung clearance index (LCI) derived from sulfur hexafluoride (SF6) multiple breath washout (MBW) is a sensitive measure of lung disease in people with cystic fibrosis (CF). However, it can be time-consuming, limiting its use clinically. AIM: To compare the repeatability, sensitivity and test duration of LCI derived from washout to 1/30th (LCI1/30), 1/20th (LCI1/20) and 1/10th (LCI1/10) to 'standard' LCI derived from washout to 1/40th initial concentration (LCI1/40).Entities:
Keywords: Cystic Fibrosis; Equipment Evaluations
Year: 2014 PMID: 25478180 PMCID: PMC4212720 DOI: 10.1136/bmjresp-2014-000031
Source DB: PubMed Journal: BMJ Open Respir Res ISSN: 2052-4439
CF and healthy control participant characteristics
| People with CF | Healthy controls | p Value (CF vs controls) | |
|---|---|---|---|
| N | 30 | 30 | – |
| M/F | 14/16 | 16/14 | 0.61 |
| Age (years) | 20.7 (11.1) (6–51) | 20.8 (10.7) (7–44) | 0.93 |
| FEV1 (% predicted) | 79.3 (17.9) (46.0–116.0) | 92.9 (11.3) (68.0–116.0) | 0.009 |
| FRC1/40 (L) | 1.95 (0.76) | 2.23 (1.0) | 0.22 |
| LCI1/40 (number of turnovers) | 9.0 (2.4) | 6.4 (0.5) | <0.0001 |
| LCI1/40 CV% | 5.2 (2.8) | 4.3 (2.0) | 0.16 |
| LCI1/40 triplicate test duration (min) | 21.8 (9.7) (8.0–57.1) | 19.4 (6.5) (10.1–31.1) | 0.26 |
| FRC1/30 (L) | 1.93 (0.76) | 2.24 (1.02) | 0.18 |
| LCI1/30 (number of turnovers) | 7.7 (1.6) | 5.9 (0.4) | <0.0001 |
| LCI1/30 CV% | 4.6 (3.2) | 4.1 (2.8) | 0.53 |
| LCI1/30 triplicate test duration (min) | 20.9 (9.1) (7.3–53.2) | 18.9 (6.2) (9.9–29.9) | 0.34 |
| FRC1/20 (L) | 1.90 (0.75) | 2.24 (1.0) | 0.15 |
| LCI1/20 (number of turnovers) | 6.6 (1.2) | 5.2 (0.4) | <0.0001 |
| LCI1/20 CV% | 5.7 (3.4) | 4.6 (2.3) | 0.14 |
| LCI1/20 triplicate test duration (min) | 20.0 (8.9) (6.7–51.7) | 18.3 5.9 (9.6–28.3) | 0.38 |
| FRC1/10 (L) | 1.84 (0.73) | 2.22 (1.00) | 0.09 |
| LCI1/10 (number of turnovers) | 4.7 (0.6) | 4.0 (0.3) | <0.0001 |
| LCI1/10 CV% | 5.4 (3.8) | 4.3 (2.8) | 0.25 |
| LCI1/10 triplicate test duration (min) | 18.6 (8.1) (5.7–46.9) | 17.1 (5.5) (8.8–26.6) | 0.41 |
All values summarised as mean (SD)±(range).
CF, cystic fibrosis; CV, coefficient of variation, F, female; FEV1, forced expiratory volume in 1 s; FRC, functional residual capacity; LCI, Lung Clearance Index; M, male.
Figure 1(A) Lung Clearance Index (LCI)1/40; (B) LCI1/30; (C) LCI1/20; (D) LCI1/10 first and third test in people with cystic fibrosis (dotted horizontal lines represent the bias and 95% limits of agreement).
Figure 2Shortened Lung Clearance Index (LCI) versus standard LCI1/40 in people with cystic fibrosis.
Figure 3Forced expiratory volume in 1 s (FEV1)% predicted versus (A) LCI1/40; (B) LCI1/30; (C) LCI1/20 (D) LCI1/10 (dotted horizontal lines represent the limits of normal for FEV1% predicted (80% predicted) and LCI (LCI1/40:7.3; LCI1/30:6.7; LCI1/20:5.9; LCI1/10:4.6)).
Figure 4Receiver operating characteristic ROC curve of Lung Clearance Index (LCI)1/40, LCI1/30, LCI1/20, LCI1/10 and inverse forced expiratory volume in 1 s % predicted: sensitivity and specificity to the presence of cystic fibrosis CF.
AUCROC and 95% CI for LCI1/40, LCI1/30, LCI1/20 LCI1/10 and inverse FEV1% predicted (1.0 indicating best performance, 0.5 indicating poor performance)
| AUCROC | 95% CI | p Value | |
|---|---|---|---|
| LCI1/40 | 0.89 | 0.80 to 0.97 | <0.0001 |
| LCI1/30 | 0.87 | 0.77 to 0.96 | <0.0001 |
| LCI1/20 | 0.87 | 0.78 to 0.96 | <0.0001 |
| LCI1/10 | 0.83 | 0.72 to 0.94 | <0.0001 |
| FEV1% predicted (inverse) | 0.73 | 0.60 to 0.86 | 0.002 |
AUC, area under the curve; FEV1, forced expiratory volume in 1 s; LCI, Lung Clearance Index, ROC, receiver operating characteristic