Anatomical mapping for atrial fibrillation ablation: a head-to-head comparison of ultrasound-assisted reconstruction versus fast anatomical mapping.

Roberto Rordorf1, Enrico Chieffo, Simone Savastano, Alessandro Vicentini, Barbara Petracci, Valentina De Regibus, Adele Valentini, Catherine Klersy, Roberto Dore, Maurizio Landolina.   

Abstract

BACKGROUND: Accuracy in left atrial (LA) anatomical reconstruction is crucial to the safe and effective performance of catheter ablation of atrial fibrillation (AF). The aim of this study was to evaluate the accuracy of LA reconstruction performed with intracardiac echocardiography (ICE) as compared to fast anatomical mapping (FAM) both integrated in the CARTO mapping system (Biosense Webster, Diamond Bar, CA, USA).
METHODS: A multislice computed tomography (MSCT) was preacquired from 29 patients with AF who underwent catheter ablation and 3D-LA geometry was reconstructed using both ICE and FAM separately. The accuracy of the LA anatomical definition was evaluated by comparing LA volumes, LA and pulmonary vein (PV) diameters obtained using ICE and FAM versus MSCT (gold standard).
RESULTS: Anterior-posterior and superior-inferior LA diameters were shorter in ICE versus MSCT (32 ± 10 vs 46 ± 9 mm and 48 ± 7 vs 53 ± 7 mm, P < 0.01) but similar in FAM versus MSCT (45 ± 9 vs 46 ± 9 mm and 52 ± 10 vs 53 ± 7 mm). Latero-septal LA diameter was similar in ICE versus MSCT (63 ± 11 vs 63 ± 9 mm) but larger in FAM versus MSCT (69 ± 9 vs 63 ± 9 mm, P < 0.001). LA volume was lower in ICE versus MSCT (73 ± 30 mL vs 116 ± 45 mL, P < 0.0001) and slightly larger in FAM versus MSCT (132 ± 45 vs 116 ± 45 mL, P = 0.06). PV diameters were similar in FAM versus MSCT but significantly underestimated with ICE.
CONCLUSIONS: Overall accuracy in the LA and PV anatomical reconstruction was found to be superior with FAM compared to ICE-guided approach. ICE resulted in a significant underestimate of both LA and PV dimensions, while FAM slightly overestimated LA geometry.
© 2014 Wiley Periodicals, Inc.

Entities:  

Keywords:  CARTO system; atrial fibrillation ablation; fast anatomical mapping; intracardiac echocardiography

Mesh:

Year:  2014        PMID: 25469451     DOI: 10.1111/pace.12539

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  2 in total

1.  Reconstruction of the left atrium for atrial fibrillation ablation using the machine learning CARTO 3 m-FAM software.

Authors:  Arie Lorin Schwartz; Ehud Chorin; Tomer Mann; Yuval Levi M; Aviram Hochstadt; Gilad Margolis; Sami Viskin; Shmuel Banai; Raphael Rosso
Journal:  J Interv Card Electrophysiol       Date:  2021-08-12       Impact factor: 1.900

Review 2.  Intracardiac Echocardiography in Catheter Ablation for Atrial Fibrillation: It Is Better to See What You Are Doing?

Authors:  Taishi Kuwahara
Journal:  J Atr Fibrillation       Date:  2015-04-30
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.