Literature DB >> 25466166

Adverse reactions to metal debris in metal-on-polyethylene total hip arthroplasty using a titanium-molybdenum-zirconium-iron alloy stem.

Manish Kiran1, Petros J Boscainos1.   

Abstract

We report a series of three patients who underwent uncemented total hip arthroplasty with a modular titanium-molybdenum-zirconium-iron stem and a cobalt-chrome-molybdenum head on an ultra-high molecular weight highly cross-linked polyethylene liner bearing. All three cases subsequently developed pain and adverse reaction to metal debris, leading to revision of the implants within thirty-six months. They were subsequently found to have hypersensitivity to cobalt or chromium. However where tested, blood metal ion levels were within MHRA guideline limits. Corrosion was noted at the taper-trunnion junction. It is possible, that the multi alloy head-neck combination may lead to corrosion. Hypersensitivity to metal ions may result to ARMD at lower metal ion levels. The use of ceramic heads may help avoid this risk.
Copyright © 2014 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  adverse reaction to metal debris; corrosion; fretting; modular stem; total hip replacement

Mesh:

Substances:

Year:  2014        PMID: 25466166     DOI: 10.1016/j.arth.2014.06.030

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  4 in total

1.  Metal-on-metal total hip arthroplasty: is there still a role in 2016?

Authors:  Edward J Silverman; Blair Ashley; Neil P Sheth
Journal:  Curr Rev Musculoskelet Med       Date:  2016-03

2.  Histopathological characterization of corrosion product associated adverse local tissue reaction in hip implants: a study of 285 cases.

Authors:  Benjamin F Ricciardi; Allina A Nocon; Seth A Jerabek; Gabrielle Wilner; Elianna Kaplowitz; Steven R Goldring; P Edward Purdue; Giorgio Perino
Journal:  BMC Clin Pathol       Date:  2016-02-27

3.  Fretting and Corrosion in Modular Shoulder Arthroplasty: A Retrieval Analysis.

Authors:  Johannes A Eckert; Ulrike Mueller; Sebastian Jaeger; Benjamin Panzram; J Philippe Kretzer
Journal:  Biomed Res Int       Date:  2016-06-28       Impact factor: 3.411

4.  Atomic Scale Origin of Metal Ion Release from Hip Implant Taper Junctions.

Authors:  Shanoob Balachandran; Zita Zachariah; Alfons Fischer; David Mayweg; Markus A Wimmer; Dierk Raabe; Michael Herbig
Journal:  Adv Sci (Weinh)       Date:  2020-01-21       Impact factor: 16.806

  4 in total