Strut penetration: local complications, breakthrough pulmonary embolism, and retrieval failure in patients with Celect vena cava filters.

Aaron Bos1, Thuong Van Ha2, Darren van Beek2, Michael Ginsburg2, Steven Zangan2, Rakesh Navuluri2, Jonathan Lorenz2, Brian Funaki2.   

Abstract

PURPOSE: To investigate strut penetration in patients with Celect filters, specifically local complications and association with breakthrough pulmonary embolism (PE) or retrieval failure.
MATERIALS AND METHODS: A retrospective single-center study was conducted to evaluate patients who received Celect filters between January 2007 and May 2013. A total of 595 filters were placed during the study period. Primary indications included thromboembolic disease (93%) and primary surgical prophylaxis (7%). Complications and retrieval data were assessed by computed tomography (CT) and electronic medical records.
RESULTS: A total of 193 patients underwent follow-up abdominal CT at a mean follow-up interval of 176.2 days (range, 0-1,739 d). The rate of strut penetration more than 3 mm outside the caval wall was 28.5% (n = 55). One patient had CT evidence of clinically major strut penetration (1.8%) with strut compression of the right ureter causing hydronephrosis. Indwelling filter time longer than 100 days was associated with strut penetration (P < .001). Age, sex, and history of thromboembolic disease were not associated with strut penetration (P = .51, P = .81, and P = .89). Sixty-three patients presented for follow-up CT pulmonary angiography at a mean of 128.1 days (range, 1-895 d). The rate of breakthrough PE was 12.7%. The overall retrieval success rate was 96.7% (n = 150). Strut penetration was not associated with breakthrough PE or retrieval failure (P = .49 and P = .22).
CONCLUSIONS: Although strut penetration is a common complication with Celect filters, there is no association with breakthrough PE or retrieval failure. CT evidence of local complications associated with strut penetration is rare.
Copyright © 2015 SIR. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2014        PMID: 25446424     DOI: 10.1016/j.jvir.2014.09.010

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  3 in total

Review 1.  A Critical Review of Available Retrievable Inferior Vena Cava Filters and Future Directions.

Authors:  Jennifer P Montgomery; John A Kaufman
Journal:  Semin Intervent Radiol       Date:  2016-06       Impact factor: 1.513

Review 2.  Inferior vena cava filter use and patient safety: legacy or science?

Authors:  William Geerts; Rita Selby
Journal:  Hematology Am Soc Hematol Educ Program       Date:  2017-12-08

3.  Successful Percutaneous Retrieval of IVC Filter with Wide Retroperitoneal Penetration Presenting with Gastrointestinal Bleeding.

Authors:  Jordan M Bond; Erik Wayne; Amit J Dwivedi; Abindra Sigdel
Journal:  Ann Vasc Dis       Date:  2017-12-25
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.