Literature DB >> 25421058

Intraoperative findings on ECAP-measurement: normal or special case?

Alexander Müller1, Thomas Hocke, Parwis Mir-Salim.   

Abstract

OBJECTIVE: This study aimed to investigate the measurability and threshold level of electrically evoked compound action potentials (ECAPs) in order to provide a baseline for the development of diagnostic references to support aftercare procedures.
DESIGN: In this retrospective study, cochlear implant patients were grouped according to the electrode array type and preoperative diagnostic findings from radiological imaging. ECAP measurements were performed intraoperatively on 22 electrodes for each patient resulting in two comparisons: (1) normal vs. pathological findings based on imaging with same electrode array, (2) perimodiolar electrode vs. straight electrode array within the normal group. STUDY SAMPLE: The study sample consisted of 218 ears with a Nucleus(®) implant.
RESULTS: In the group with normal radiological imaging, the percentage of measurable ECAPs was higher than in the pathological group. The ECAP thresholds on 21 electrodes were significantly different between the two groups. Furthermore the thresholds were significantly lower for basal electrodes mainly for the two different electrode types.
CONCLUSION: The pathological changes in the inner ear, the type of electrode array, and the electrode position affect the ECAP threshold. Further consideration suggests that a receiver operating characteristic curve can be derived from the results.

Entities:  

Keywords:  Cochlear implant; NRT; evoked compound action potential; imaging diagnostic; neural response telemetry

Mesh:

Year:  2014        PMID: 25421058     DOI: 10.3109/14992027.2014.969410

Source DB:  PubMed          Journal:  Int J Audiol        ISSN: 1499-2027            Impact factor:   2.117


  5 in total

Review 1.  Cochlear Implantation.

Authors:  Stefan Dazert; Jan Peter Thomas; Andreas Loth; Thomas Zahnert; Timo Stöver
Journal:  Dtsch Arztebl Int       Date:  2020-10-09       Impact factor: 5.594

2.  Radiological evaluation of a new straight electrode array compared to its precursors.

Authors:  Manuel Christoph Ketterer; A Aschendorff; S Arndt; I Speck; A K Rauch; R Beck; F Hassepass
Journal:  Eur Arch Otorhinolaryngol       Date:  2020-10-22       Impact factor: 2.503

3.  Effect of inner ear malformations on intraoperative ECAP thresholds and postoperative auditory performance.

Authors:  Jeong-Seo Kim; Sung Hwa Hong; Il Joon Moon
Journal:  Laryngoscope Investig Otolaryngol       Date:  2022-06-15

4.  Relationship between Electrically Evoked Compound Action Potential Thresholds and Auditory, Language, and Speech Progress after Cochlear Implant Surgery.

Authors:  Masoud Motasaddi Zarandy; Navid Nourizadeh; Farzad Mobedshahi; Sadegh Jafarzadeh
Journal:  Iran J Otorhinolaryngol       Date:  2018-07

5.  Electrophysiological measurements with electrode types of different perimodiolar properties and the same cochlear implant electronics - a retrospective comparison study.

Authors:  A Perenyi; F Toth; B Dimak; R Nagy; P Schoerg; J Jori; J G Kiss; G Sprinzl; M Csanady; L Rovo
Journal:  J Otolaryngol Head Neck Surg       Date:  2019-09-06