Literature DB >> 25419104

A 10-mm MR-Conditional Unidirectional Pneumatic Stepper Motor.

Yue Chen1, Christopher D Mershon1, Zion Tsz Ho Tse1.   

Abstract

Magnetic resonance (MR) conditional robotic devices facilitate accurate interventional procedures under MR imaging (MRI) guidance. For this purpose, a compact (10-mm diameter) MR-conditional stepper motor is presented. The device features seven key components, which contribute to a dense and easy to fabricate design. Alternating bursts of pressurized air and vacuum can drive the motor in 60° per step to achieve a maximum torque of 2.4 mNm. The relationship between torque and angular speed was investigated to demonstrate motor performance under different loading conditions. The stepper motor was tested in a GE 3T MRI scanner to verify its MR-compatibility. A maximum artifact width of 3 mm was measured in MRI images and a maximum signal-to-noise ratio reduction of 2.49% was recorded.

Entities:  

Keywords:  MR imaging (MRI); Magnetic resonance (MR)-conditional; pneumatic stepper motor

Year:  2015        PMID: 25419104      PMCID: PMC4235678          DOI: 10.1109/TMECH.2014.2305839

Source DB:  PubMed          Journal:  IEEE ASME Trans Mechatron        ISSN: 1083-4435            Impact factor:   5.303


  7 in total

1.  Towards MRI guided surgical manipulator.

Authors:  K Chinzei; K Miller
Journal:  Med Sci Monit       Date:  2001 Jan-Feb

2.  Mutual interferences and design principles for mechatronic devices in magnetic resonance imaging.

Authors:  Ningbo Yu; Roger Gassert; Robert Riener
Journal:  Int J Comput Assist Radiol Surg       Date:  2010-09-02       Impact factor: 2.924

3.  Multi-imager compatible actuation principles in surgical robotics.

Authors:  D Stoianovici
Journal:  Int J Med Robot       Date:  2005-01       Impact factor: 2.547

4.  Fusion of multislice computed tomography imaging with three-dimensional electroanatomic mapping to guide radiofrequency catheter ablation procedures.

Authors:  Laurens F Tops; Jeroen J Bax; Katja Zeppenfeld; Monique R M Jongbloed; Hildo J Lamb; Ernst E van der Wall; Martin J Schalij
Journal:  Heart Rhythm       Date:  2005-10       Impact factor: 6.343

5.  Coronary artery imaging with intravascular high-frequency ultrasound.

Authors:  B N Potkin; A L Bartorelli; J M Gessert; R F Neville; Y Almagor; W C Roberts; M B Leon
Journal:  Circulation       Date:  1990-05       Impact factor: 29.690

6.  A New Type of Motor: Pneumatic Step Motor.

Authors:  Dan Stoianovici; Alexandru Patriciu; Doru Petrisor; Dumitru Mazilu; Louis Kavoussi
Journal:  IEEE ASME Trans Mechatron       Date:  2007-02-01       Impact factor: 5.303

7.  Outcome of MRI-guided breast biopsy.

Authors:  Boo-Kyung Han; Mitchell D Schnall; Susan G Orel; Mark Rosen
Journal:  AJR Am J Roentgenol       Date:  2008-12       Impact factor: 3.959

  7 in total
  5 in total

1.  An MR-conditional high-torque pneumatic stepper motor for MRI-guided and robot-assisted intervention.

Authors:  Yue Chen; Ka-Wai Kwok; Zion Tsz Ho Tse
Journal:  Ann Biomed Eng       Date:  2014-06-24       Impact factor: 3.934

2.  System Integration and Preliminary Clinical Evaluation of a Robotic System for MRI-Guided Transperineal Prostate Biopsy.

Authors:  Niravkumar A Patel; Gang Li; Weijian Shang; Marek Wartenberg; Tamas Heffter; Everette C Burdette; Iulian Iordachita; Junichi Tokuda; Nobuhiko Hata; Clare M Tempany; Gregory S Fischer
Journal:  J Med Robot Res       Date:  2018-05-15

3.  MR-conditional steerable needle robot for intracerebral hemorrhage removal.

Authors:  Yue Chen; Isuru S Godage; Saikat Sengupta; Cindy Lin Liu; Kyle D Weaver; Eric J Barth
Journal:  Int J Comput Assist Radiol Surg       Date:  2018-09-01       Impact factor: 2.924

4.  Piezoelectrically Actuated Robotic System for MRI-Guided Prostate Percutaneous Therapy.

Authors:  Hao Su; Weijian Shang; Gregory Cole; Gang Li; Kevin Harrington; Alexander Camilo; Junichi Tokuda; Clare M Tempany; Nobuhiko Hata; Gregory S Fischer
Journal:  IEEE ASME Trans Mechatron       Date:  2015-08       Impact factor: 5.303

5.  Design and Fabrication of MR-Tracked Metallic Stylet for Gynecologic Brachytherapy.

Authors:  Yue Chen; Wei Wang; Ehud J Schmidt; Ka-Wai Kwok; Akila N Viswanathan; Robert Cormack; Zion Tsz Ho Tse
Journal:  IEEE ASME Trans Mechatron       Date:  2015-11-24       Impact factor: 5.303

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.