| Literature DB >> 25391023 |
Joshua A Roberti1, Michael D SanClements2, Henry W Loescher2, Edward Ayres2.
Abstract
Even though fine-root turnover is a highly studied topic, it is often poorly understood as a result of uncertainties inherent in its sampling, e.g., quantifying spatial and temporal variability. While many methods exist to quantify fine-root turnover, use of minirhizotrons has increased over the last two decades, making sensor errors another source of uncertainty. Currently, no standardized methodology exists to test and compare minirhizotron camera capability, imagery, and performance. This paper presents a reproducible, laboratory-based method by which minirhizotron cameras can be tested and validated in a traceable manner. The performance of camera characteristics was identified and test criteria were developed: we quantified the precision of camera location for successive images, estimated the trueness and precision of each camera's ability to quantify root diameter and root color, and also assessed the influence of heat dissipation introduced by the minirhizotron cameras and electrical components. We report detailed and defensible metrology analyses that examine the performance of two commercially available minirhizotron cameras. These cameras performed differently with regard to the various test criteria and uncertainty analyses. We recommend a defensible metrology approach to quantify the performance of minirhizotron camera characteristics and determine sensor-related measurement uncertainties prior to field use. This approach is also extensible to other digital imagery technologies. In turn, these approaches facilitate a greater understanding of measurement uncertainties (signal-to-noise ratio) inherent in the camera performance and allow such uncertainties to be quantified and mitigated so that estimates of fine-root turnover can be more confidently quantified.Entities:
Mesh:
Year: 2014 PMID: 25391023 PMCID: PMC4229195 DOI: 10.1371/journal.pone.0112362
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Metrology terms and definitions.
| Variable | Definition (is the-) |
| Accuracy | closeness of agreement between a measured quantity value and a |
| true quantity value of a measurand; the concept ‘measurement | |
| Accuracy' is not a quantity and is not given a numerical quantity | |
| Value | |
| Precision | closeness of agreement between indications or measured quantity |
| values obtained by replicate measurements on the same or similar | |
| objects under specified conditions; the ‘specified conditions’ can | |
| be, for example, repeatability conditions of measurement etc. | |
| Relative uncertainty | standard measurement uncertainty divided by the absolute value of |
| the measured quantity value | |
| Repeatability condition | condition of measurement, out of a set of conditions that includes |
| the same measurement procedure, same operators, same | |
| measuring system, same operating conditions and same location, | |
| and replicate measurements on the same or similar objects over a | |
| short period of time | |
| Reproducibility condition | condition of measurement, out of a set of conditions that includes |
| different locations, operators, measurement systems, and | |
| replicate measurements on the same or similar objects | |
| Resolution | smallest change in a quantity being measured that causes a |
| perceptible change in the corresponding indication | |
| Resolvable limit | minimum distance at which neighboring objects of the same |
| intensity are resolved as two objects | |
| Sensitivity | quotient of the change in an indication of a measuring system and |
| the corresponding change in a value of a quantity being measured | |
| Standard uncertainty | measurement uncertainty expressed as a standard deviation |
| Trueness | closeness of agreement between the average of an infinite number |
| of replicate measured quantity values and a reference quantity | |
| Value | |
| Type A evaluation | evaluation of a component of measurement uncertainty by a |
| statistical analysis of measured quantity values obtained under | |
| defined measurement conditions, | |
| Type B evaluation | evaluation of a component of measurement uncertainty determined |
| by means other than a Type A evaluation |
All definitions are provided by JGCM [27] standards-holding body unless otherwise specified.
Specifications and manufacturer recommended default settings of each minirhizotron
| Specification/Settings | CI-600 | AMR-A |
| Tube length (cm) | 182.0 (Others available) | 156.8 (including handle and fin) |
| Tube diameter (cm) | 6.4 | 10.8 |
| Tube circumference (cm) | 20.1 | 33.9 |
| Operating range (°C) | Dependent on laptop | −12 to 45 |
| Power requirements | Powered by laptop | 5A @ 120 VAC |
| Removable camera | Yes | No |
| Operations | On site; field technician | Remotely; automated |
| Image dimensions (cm) | 21.6×19.6 | 0.226×0.301 |
|
| 42.3×42.3 (∼24 pixels (p) mm−1) | 4.7×4.7 (∼212 p mm−1) |
| Pixels/image | 2.356×107 | 3.07×105 |
| Images/full-tube scan | 8 (with 182.0 cm tube) | 3.2928×104 |
| Software | CI-600 Root Scanner v.3.1.22 | Root View v3.0.0.7 |
|
| Brightness = 0; (50%) | Camera Exposure = 8; (60%) |
| Contrast = 0; (50%) | LED camera lights = 192; (100%) | |
| Threshold = 0; (50%) | Dwell time = 100 ms |
manufacturer recommended resolutions (B Meyer, CID-BioScience Inc., pers. comm.; M Taggart RhizoSystems, LLC, pers. comm.)
settings provided with percentage relative to the maximum range of the particular setting [24], [25]
Description of camera position tests.
| Test | Condition of measurement | Description |
| A1 | Reproducibility | AMR-A: Camera homed between subsequent scans; |
| deviation from home location | ||
| manufacturer recommended 20 µm | ||
| A2 | Reproducibility | AMR-A: Camera homed between subsequent scans; |
| deviation from home location | ||
| the manufacturer recommended 20 µm | ||
| A3 | Repeatability | AMR-A: Camera not homed between subsequent scans |
| C1 | Reproducibility | CI-600: Removal and replacement of camera |
| C2 | Repeatability | CI-600: No removal and replacement of camera |
Each test had = 30.
Line widths and expanded uncertainties as measured by the NIST traceable micrometer.
| Line Width (µm) | ±U95 (µm) |
| 409.63 | 2.22 |
| 348.21 | 1.64 |
| 303.53 | 1.64 |
| 268.48 | 1.64 |
| 242.62 | 1.64 |
| 194.06 | 1.64 |
| 161.95 | 1.38 |
| 120.27 | 1.38 |
| 100.15 | 1.38 |
| 73.81 | 1.38 |
| 57.99 | 1.38 |
| 43.66 | 1.38 |
| 39.29 | 1.38 |
Figure 1Conceptual diagram depicting the resolution of two objects (smooth contours) with equal intensities at varying distances.
The dashed lines denote two conceptual objects of equal intensity, and the shaded, gray area represents what the camera resolves; (a) fully resolvable, (b) just resolvable (Rayleigh Criterion), and (c) unresolvable, i.e., both objects are resolved as a single object.
Mean chromaticity coordinates (and expanded uncertainties) converted from reflectance data measured by the densitometer.
| Color | x | ±U95(x) | y | ±U95(y) |
| Yellow | 0.463 | 0.001 | 0.494 | 0.001 |
| Amber | 0.424 | 0.001 | 0.397 | 0.001 |
| White | 0.349 | 0.001 | 0.363 | 0.002 |
Figure 2Experimental design to determine heat dissipation.
A cross-section schematic of the experimental design used to assess heat dissipation (not to scale) for the AMR-A. The three depths: Top, Middle, and Bottom, correspond to depths of 5, 48, and 86 cm below the sand surface, respectively. The locations of the PRTs residing 0.5 cm from the minirhizotron are denoted by letters A through F, and correspond to the results shown in Figure 5. The LED lights reside on the body of the camera. The 60° angle refers to the overall installation relative to the horizontal (plane).
Figure 5Time series of temperature change at six profile locations.
Temperature measurements were made at the three depths (5 [top], 48 [middle], and 86 [bottom] cm) below the surface of the sand and at 0.5 cm away from the AMR-A's tube, the locations of the PRTs A through F are shown in Figure 2. Temperature change at each location was quantified relative to the temperature of the respective PRT at the start time of each full-tube scan.
Precision of minirhizotron camera location for each sampling method; values are provided as one standard deviation.
| Precision | ||
| Test | Radial (degrees) | Vertical (µm) |
| A1 | 0.12 | 24.87 |
| A2 | 0.01 | 4.32 |
| A3 | 0.01 | 38.37 |
| C1 | 13.72 | 1942.23 |
| C2 | 0.22 | 84.06 |
Figure 3Line widths as measured by both minirhizotrons.
Estimated line widths as determined by both minirhizotrons cameras against those of the NIST traceable micrometer. Results of the AMR-A are shown in (a) using the 250 pixel intensity threshold (), and (b) using the FWHM threshold (). Results of the CI-600 are shown in (c) using the 250 pixel intensity threshold (), and (d) using the FWHM threshold (). A 1∶1 line is shown on each graph. Inset plots are provided at the 194.06±1.64 µm calibration point to emphasize the magnitude of random uncertainty and show that calibration can correct for bias, but not random uncertainty. Error bars are given as ±2 SE for uncalibrated data and ± for calibrated data.
Mean sRGB values (and expanded uncertainties) as measured by the AMR-A.
| Color | R | ±2SE (R) | G | ±2SE (G) | B | ±2SE (B) |
|
| 0 | N/A | 107.529 | 0.059 | 244.23 | 0.026 |
|
| 206.423 | 0.035 | 123.219 | 0.048 | 200.561 | 0.041 |
|
| 232.684 | 0.022 | 251.412 | 0.019 | 17.079 | 0.127 |
|
| 231.734 | 0.024 | 234.742 | 0.011 | 202.499 | 0.033 |
|
| 130.17 | 0.048 | 0 | N/A | 0 | N/A |
|
| 0 | N/A | 89.485 | 0.061 | 23.089 | 0.089 |
|
| 0 | N/A | 0 | N/A | 193.347 | 0.054 |
|
| 245.703 | 0.025 | 255 | N/A | 255 | N/A |
Expanded uncertainties represent the repeatability of the camera for the specific color patch. Uncertainties for values at the limits of the sRGB spectrum are not provided. per color patch.
Mean sRGB values (and expanded uncertainties) as measured by the CI-600.
| Color | R | ±2SE (R) | G | ±2SE (G) | B | ±2SE (B) |
|
| 90.406 | 0.026 | 179.614 | 0.013 | 217.403 | 0.013 |
|
| 255 | N/A | 167.971 | 0.012 | 200.864 | 0.011 |
|
| 255 | N/A | 249.617 | 0.013 | 177.415 | 0.019 |
|
| 255 | N/A | 232.644 | 0.025 | 214.657 | 0.033 |
|
| 231.103 | 0.015 | 95.250 | 0.027 | 95.827 | 0.029 |
|
| 87.799 | 0.024 | 171.210 | 0.011 | 119.531 | 0.015 |
|
| 78.508 | 0.047 | 128.516 | 0.024 | 188.291 | 0.014 |
|
| 255 | N/A | 255 | N/A | 255 | N/A |
Expanded uncertainties represent the repeatability of the camera for the specific color patch. Uncertainties for values at the limits of the sRGB spectrum are not provided. per color patch.
Mean, characterized chromatic coordinates (and expanded uncertainties) of each minirhizotron camera.
| AMR-A | CI-600 | |||||||
| Color | x | ±U95(x) | y | ±U95(y) | x | ±U95(x) | y | ±U95(x) |
|
| 0.455 | 0.004 | 0.488 | 0.004 | 0.464 | 0.003 | 0.493 | 0.002 |
|
| 0.425 | 0.004 | 0.396 | 0.004 | 0.424 | 0.003 | 0.397 | 0.002 |
|
| 0.349 | 0.004 | 0.362 | 0.004 | 0.349 | 0.002 | 0.363 | 0.002 |
Figure 4Characterized CIE chromaticity coordinates.
CIE chromaticity coordinates of the (a) three characterization points relative to the entire chromaticity coordinate space (as denoted by the horseshoe shaped outline), (b) a zoomed version of the same data in (a) that also depicts the expanded uncertainties of the three characterization points. Error bars are given as ± This figure is presented in addition to Table 9 to demonstrate the spatial (color-space) limitations of the linear characterization performed in this analysis. The overlaid color gamut is for illustrative purposes only.