| Literature DB >> 25390881 |
Guodong Feng1, Yang Zhao1, Xu Tian1, Zhiqiang Gao1.
Abstract
This study aimed to establish a 3-dimensional dynamic quantitative facial motion analysis system, and then determine its accuracy and test-retest reliability. The system could automatically reconstruct the motion of the observational points. Standardized T-shaped rod and L-shaped rods were used to evaluate the static and dynamic accuracy of the system. Nineteen healthy volunteers were recruited to test the reliability of the system. The average static distance error measurement was 0.19 mm, and the average angular error was 0.29°. The measuring results decreased with the increase of distance between the cameras and objects, 80 cm of which was considered to be optimal. It took only 58 seconds to perform the full facial measurement process. The average intra-class correlation coefficient for distance measurement and angular measurement was 0.973 and 0.794 respectively. The results demonstrated that we successfully established a practical 3-dimensional dynamic quantitative analysis system that is accurate and reliable enough to meet both clinical and research needs.Entities:
Mesh:
Year: 2014 PMID: 25390881 PMCID: PMC4229263 DOI: 10.1371/journal.pone.0113115
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Figure 13-Dimensional Dynamic Quantitative Analysis System of Facial Motion.
Figure 2Observational points on face.
A/a tragus parallel to the upper wall of external acoustic canal, B/b central position above the eyebrow, C/c center of the upper eyelid, D/d center of the lower eyelid, E/e angulus oculi temporalis, F/f angulus oculi medialis, G/g ala of the nose, H/h corner of the mouth, I root of the columella nasi, J center of eyebrows, K bony–cartilaginous junction along the nasal dorsum, II philtrum, III center of the lower lip.
Figure 3Reference planes.
Figure 4Specialized helmet.
Standardized facial expressions.
| Standardized facial expressions |
| 1.Maximal brow lift |
| 2.Light eye closure |
| 3.Maximally tight eye closure |
| 4.Maximally frown |
| 5.Maximally showing teeth |
| 6.move right mouth corner to right-back |
| 7. move left mouth corner to left-back |
| 8. maximal jaw drop |
| 9.Smile(showing at least 8 upper teeth) |
| 10.Maximal pucker |
Figure 5A, online reconstruction; B, offline reconstruction.
Static measurement ICC.
| static measurement |
| 95% | |
| G-H (distance) | 0.995 | 0.986 | 0.998 |
| G-H (direction) X | 0.643 | 0.288 | 0.844 |
| G-H (direction) Z | 0.346 | −0.106 | 0.682 |
| g-h (distance) | 0.985 | 0.962 | 0.994 |
| g-h (direction) X | 0.814 | 0.586 | 0.924 |
| g-h (direction) Z | 0.706 | 0.39 | 0.875 |
| I-G (distance) | 0.986 | 0.965 | 0.995 |
| I-G (direction) X | 0.585 | 0.2 | 0.815 |
| I-G (direction) Z | 0.721 | 0.416 | 0.882 |
| I-g (distance) | 0.988 | 0.97 | 0.995 |
| I-g (direction) X | 0.11 | −0.343 | 0.525 |
| I-g (direction) Z | 0.727 | 0.427 | 0.885 |
| ∠CED | 0.985 | 0.962 | 0.994 |
| ∠ced | 0.819 | 0.596 | 0.926 |
| ∠EHh | 0.745 | 0.458 | 0.893 |
| ∠ehH | 0.627 | 0.263 | 0.837 |
| H-h(distance) | 0.982 | 0.956 | 0.993 |
| E-H(distance) | 0.986 | 0.964 | 0.994 |
| E-h(distance) | 0.987 | 0.967 | 0.995 |
| II-III (distance) | 0.951 | 0.874 | 0.98 |
X = X axis, Z = Z axis, ∠ = angle.