Automatic noninvasive tear breakup time (TBUT) and conventional fluorescent TBUT.

Weizhong Lan1, Lixia Lin, Xiao Yang, Minbin Yu.   

Abstract

PURPOSE: To compare the effects of different volumes of fluorescein on tear breakup time (FTBUT) and to investigate if and to what extent the tear breakup time determined by an automated noninvasive instrument (NITBUT) differs from FTBUT.
METHODS: Twenty-four healthy volunteers were recruited to the study. Fluorescent tear breakup time was measured with different volumes of fluorescein solution delivered by either glass rod or objective directly by fluorescein strip. Noninvasive tear breakup time was measured with a noninvasive instrument (Oculus Keratograph 5M, Germany; K5). The average levels of the tear breakup times (TBUTs) and the variability of the successive recordings for each measurement were compared.
RESULTS: Increasing the volume of fluorescein delivered from 1 to 7 μl lengthened FTBUT by a mean ratio of 1.26-fold (p = 0.019) for the glass rod technique. No significant difference was detected in the FTBUT measured by the fluorescein strip technique when the delivered volume was increased from 4.5 to 7 μl. The variability of successive recordings was stable across the tested volumes for both techniques. Noninvasive tear breakup time determined by K5 was significantly longer than FTBUTs (mean difference: 3.90 seconds, p = 0.003 and 4.12 seconds, p = 0.002, respectively). Although no significant difference was detected in the average SD for successive recordings among K5 and the other two invasive techniques (p = 0.325), the variability of NITBUT was found significantly dependent on the observed TBUT (r = 0.532, p = 0.007).
CONCLUSIONS: The Oculus Keratograph 5M usually produced substantially longer TBUT compared with invasive techniques tested. The variability of readings by this novel method increased with the values of NITBUT. Given the small sample size in the current experiment, however, these findings need to be confirmed in a larger study.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25343685     DOI: 10.1097/OPX.0000000000000418

Source DB:  PubMed          Journal:  Optom Vis Sci        ISSN: 1040-5488            Impact factor:   1.973


  17 in total

1.  Agreement between Automated and Traditional Measures of Tear Film Breakup.

Authors:  Stephanie M Cox; Kelly K Nichols; Jason J Nichols
Journal:  Optom Vis Sci       Date:  2015-09       Impact factor: 1.973

2.  Repeatability of Non-invasive Keratograph Break-Up Time measurements obtained using Oculus Keratograph 5M.

Authors:  José Vicente García-Marqués; Noelia Martínez-Albert; Cristian Talens-Estarelles; Santiago García-Lázaro; Alejandro Cerviño
Journal:  Int Ophthalmol       Date:  2021-03-16       Impact factor: 2.031

3.  Agreement and repeatability of objective systems for assessment of the tear film.

Authors:  Joaquín Fernández; Manuel Rodríguez-Vallejo; Javier Martínez; Ana Tauste; Javier García-Montesinos; David P Piñero
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2018-04-18       Impact factor: 3.117

4.  A novel noninvasive ocular surface analyzer for the assessment of dry eye with Meibomian gland dysfunction.

Authors:  Yuanyuan Qi; Chen Zhang; Shaozhen Zhao; Yue Huang; Ruibo Yang
Journal:  Exp Ther Med       Date:  2017-04-19       Impact factor: 2.447

5.  Influence of short-term face mask wear on semiautomatically measured tear film parameters in dry eye patients and healthy volunteers.

Authors:  Marc Schargus; E M Zimzik; L Fuhrmann; G Geerling
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2022-10-21       Impact factor: 3.535

Review 6.  The Growing Need for Validated Biomarkers and Endpoints for Dry Eye Clinical Research.

Authors:  Neeta S Roy; Yi Wei; Eric Kuklinski; Penny A Asbell
Journal:  Invest Ophthalmol Vis Sci       Date:  2017-05-01       Impact factor: 4.799

Review 7.  A Review of Imaging Biomarkers of the Ocular Surface.

Authors:  William W Binotti; Betul Bayraktutar; M Cuneyt Ozmen; Stephanie M Cox; Pedram Hamrah
Journal:  Eye Contact Lens       Date:  2020-03       Impact factor: 3.152

8.  Correlation of Measures From the OCULUS Keratograph and Clinical Assessments of Dry Eye Disease in the Dry Eye Assessment and Management Study.

Authors:  John E Sutphin; Gui-Shuang Ying; Vatinee Y Bunya; Yinxi Yu; Meng C Lin; Kathleen McWilliams; Elizabeth Schmucker; Eric J Kuklinski; Penny A Asbell; Maureen G Maguire
Journal:  Cornea       Date:  2021-07-21       Impact factor: 3.152

9.  Effect of Proparacaine 0.375%-Sodium Fluorescein 0.25% Eye Drop Mixture and Fluorescein Strip on Anterior Segment Parameters.

Authors:  Mustafa Doğan; Mehmet Cem Sabaner; Mehmet Akif Erol
Journal:  J Ophthalmol       Date:  2018-09-05       Impact factor: 1.909

Review 10.  Tear instability importance, mechanisms, validity and reliability of assessment.

Authors:  Charles W Mcmonnies
Journal:  J Optom       Date:  2018-01-11
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.