Radiation dose reduction at coronary artery calcium scoring by using a low tube current technique and hybrid iterative reconstruction.

Noriaki Matsuura1, Masaki Urashima, Wataru Fukumoto, Hiroshi Sunamori, Fuminari Tatsugami, Naoyuki Toyota, Kazuo Awai.   

Abstract

PURPOSE: The aim of this study was to compare the accuracy of coronary artery calcium scoring (CACS) on cardiac computed tomographic images using hybrid iterative reconstruction (hIR) and a low tube current as well as on images acquired with a filtered back projection (FBP) algorithm and a normal tube current. SUBJECTS AND METHODS: Patients (N = 77) with suspected coronary artery disease were subjected to 2 CACS evaluations based on their Agatston, volume, and mass scores. One CACS evaluation was performed on images obtained with a 364-mA tube current and reconstructed with FBP; the other was performed on images obtained with a 73-mA tube current and reconstructed with hIR at iDose4. All scans were performed with the prospective electrocardiogram-triggered method using a 256-slice computed tomographic scanner (Brilliance iCT; Philips). We assessed agreement between calcium scores obtained with FBP and with IR using the percentage difference and Bland-Altman analysis.
RESULTS: The effective radiation doses for CACS at 80 mA s with FBP and at 16 mA s with IR were 1.20 and 0.24 mSv, respectively (k = 0.014). The mean Agatston, volume, and mass scores at 80 mA s with FBP as well as at 16 mA s with IR were 390.7, 146.5, and 63.2 as well as 377.7, 142.5, and 62.2, respectively. The percentage difference between FBP and hIR for the Agatston, volume, and mass score was 20.7%, 20.7%, and 27.1%, respectively. Bland-Altman analysis showed that there was no systemic bias.
CONCLUSIONS: The radiation dose for CACS can be reduced at a low tube current and hIR without affecting the calcium score.

Entities:  

Mesh:

Year:  2015        PMID: 25319604     DOI: 10.1097/RCT.0000000000000168

Source DB:  PubMed          Journal:  J Comput Assist Tomogr        ISSN: 0363-8715            Impact factor:   1.826


  9 in total

1.  Coronary calcium screening with dual-source CT: reliability of ungated, high-pitch chest CT in comparison with dedicated calcium-scoring CT.

Authors:  Antoine Hutt; Alain Duhamel; Valérie Deken; Jean-Baptiste Faivre; Francesco Molinari; Jacques Remy; Martine Remy-Jardin
Journal:  Eur Radiol       Date:  2015-09-04       Impact factor: 5.315

Review 2.  Risk Stratification for Primary Prevention of Coronary Artery Disease: Roles of C-Reactive Protein and Coronary Artery Calcium.

Authors:  Waqas T Qureshi; Jamal S Rana; Joseph Yeboah; Usama Bin Nasir; Mouaz H Al-Mallah
Journal:  Curr Cardiol Rep       Date:  2015-12       Impact factor: 2.931

3.  Influence of iterative reconstruction on coronary calcium scores at multiple heart rates: a multivendor phantom study on state-of-the-art CT systems.

Authors:  N R van der Werf; M J Willemink; T P Willems; M J W Greuter; T Leiner
Journal:  Int J Cardiovasc Imaging       Date:  2017-12-28       Impact factor: 2.357

4.  Prospective evaluation of the influence of iterative reconstruction on the reproducibility of coronary calcium quantification in reduced radiation dose 320 detector row CT.

Authors:  Andrew D Choi; Eric S Leifer; Jeannie Yu; Sujata M Shanbhag; Kathie Bronson; Andrew E Arai; Marcus Y Chen
Journal:  J Cardiovasc Comput Tomogr       Date:  2016-07-27

5.  Accuracy of coronary artery calcium scoring with tube current reduction by 75%, using an adaptive iterative reconstruction algorithm.

Authors:  Reny Luhur; Joanne D Schuijf; Jürgen Mews; Jörg Blobel; Bernd Hamm; Alexander Lembcke
Journal:  Br J Radiol       Date:  2018-01-10       Impact factor: 3.039

6.  Iterative reconstruction can permit the use of lower X-ray tube current in CT coronary artery calcium scoring.

Authors:  Mark A Rodrigues; Michelle C Williams; Thomas Fitzgerald; Martin Connell; Nicholas W Weir; David E Newby; Edwin J R van Beek; Saeed Mirsadraee
Journal:  Br J Radiol       Date:  2016-06-08       Impact factor: 3.039

Review 7.  Low Radiation Dose Calcium Scoring: Evidence and Techniques.

Authors:  Kaitlin B Baron; Andrew D Choi; Marcus Y Chen
Journal:  Curr Cardiovasc Imaging Rep       Date:  2016-03-02

8.  In vitro evaluation of a new iterative reconstruction algorithm for dose reduction in coronary artery calcium scoring.

Authors:  Tobias Gassenmaier; Thomas Allmendinger; Andreas S Kunz; Maike Veyhl-Wichmann; Süleyman Ergün; Thorsten A Bley; Bernhard Petritsch
Journal:  Acta Radiol Open       Date:  2017-05-29

9.  Quantifying the impact of using Coronary Artery Calcium Score for risk categorization instead of Framingham Score or European Heart SCORE in lipid lowering algorithms in a Middle Eastern population.

Authors:  Hussain A Isma'eel; Mohamad M Almedawar; Bernard Harbieh; Wissam Alajaji; Laila Al-Shaar; Mukbil Hourani; Fadi El-Merhi; Samir Alam; Antoine Abchee
Journal:  J Saudi Heart Assoc       Date:  2015-05-23
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.