| Literature DB >> 25317185 |
Syed Husain Hashemi Mousavi1, Farzad Kobarfard2, Syed Waqif Husain1, Mohammad Saber Tehrani1, Parviz Abroomand Azar1, Reza Ahmadkhaniha3, Amir Mehdizadeh4.
Abstract
A fast, accurate, sensitive, selective and reliable method using reversed-phase high performance liquid chromatography coupled to electrospray ionization ion trap mass spectrometry was developed and validated for the determination of finasteride in human plasma. After protein precipitation with perchloric acid, satisfactory separation was achieved on a Zorbax Eclipse(®) C8 analytical column using a mobile phase consisted of acetonitrile, 2 mM ammonium formate buffer (58:42, pH adjusted at 2.5 using formic acid); the flow rate was 0.25 mLmin(-1) and the column oven was set to 50°C. Tamoxifen citrate was used as internal standard. This method involved the use of [M +H](+) ions of finasteride and IS at m/z 373 and 372 respectively with the selected ion monitoring (SIM) mode. The calibration curve was linear over the range of 0.1-60 ng mL(-1). The limit of quantification for finasteride in plasma was 0.1 ng mL(-1). The intra-day and inter-day repeatability (precision) were 2.68-13.87% and 2.14-14.69% respectively. Intra-day and inter-day accuracy were 98-101.57% and 99.7-110%. The assay method has been successfully used to estimate the pharmacokinetics of finasteride after oral administration of a 5 mg tablet of finasteride in 12 healthy volunteers.Entities:
Keywords: Electrospray ionization (ESI); Finasteride; Ion trap; Liquid chromatography–mass spectrometry (LC–MS); Pharmacokinetic
Year: 2012 PMID: 25317185 PMCID: PMC3876561
Source DB: PubMed Journal: Iran J Pharm Res ISSN: 1726-6882 Impact factor: 1.696
Figure 1Chemical structures of finasteride and tamoxifen
Figure 2Ion trap LC/ESI-MS mass spectra of finasteride and tamoxifen (IS) in positive scan mode from m/z 100 to 850. Chromatographic conditions and MS parameters are described in Sections 2.3 and 2.4.
Figure 3Chromatograms of finasteride and tamoxifen (IS). (A) blank plasma, (B) Plasma spiked with standard (0.1 ng mL −1) and IS (100 ng mL −1), and (C) Plasma spiked with standard (30 ng mL −1) and IS (100 ng mL −1).
Intra-day precision and accuracy of quality control samples for the determination of finasteride in human plasma
|
|
|
|
|
|---|---|---|---|
| 0.1 | 0.098 | 98.00 | 13.87 |
| 30 | 30.47 | 101.57 | 5.75 |
| 60 | 59.71 | 99.52 | 2.68 |
Inter-day precision and accuracy of quality control samples for the determination of finasteride in human plasma
|
|
|
|
|
|---|---|---|---|
| 0.1 | 0.11 | 110 | 14.69 |
| 30 | 30.55 | 101.83 | 4.32 |
| 60 | 59.89 | 99.77 | 2.14 |