The evaluation of data completeness and image quality in multiplexing multi-pinhole SPECT.

Karen Van Audenhaege, Christian Vanhove, Stefaan Vandenberghe, Roel Van Holen.   

Abstract

Multi-pinhole collimators are often used in pre-clinical SPECT systems because they have a better resolution-sensitivity tradeoff than parallel hole collimators when imaging small objects. Most multi-pinhole collimators are designed to allow no or only a limited amount of overlap between the different pinhole projections because the ambiguity introduced by multiplexing pinholes can result in artifacts. The origin of these artifacts is still not fully understood, but previous research has already shown that data incompleteness could be part of the explanation. Therefore, we developed a method to investigate data completeness in multiplexing multi-pinhole systems and showed that a certain activity distribution can be successfully reconstructed when the nonmultiplexed data is complete or when the overlap can be sufficiently de-multiplexed. We validated this method using computer simulated phantom data of different multiplexing systems. We also studied contrast-to-noise and nonprewhitening matched filter signal-to-noise ratio (NPW-SNR) to compare the image quality in a single pinhole system with multiplexing systems. We found that our method can indeed be used to evaluate data completeness in multiplexing systems and found no artifacts in the systems that had complete data. Sensitivity increased significantly with multiplexing but we found only small, nonsignificant differences in contrast-to-noise ratio. However, the NPW-SNR did slightly improve in the multiplexing setups. We conclude that more multiplexing does not necessarily result in more artifacts and that even a high amount of multiplexing can still result in artifact-free images if the nonmultiplexed data is complete or when the overlap can be sufficiently de-multiplexed.

Mesh:

Year:  2014        PMID: 25291791     DOI: 10.1109/TMI.2014.2361051

Source DB:  PubMed          Journal:  IEEE Trans Med Imaging        ISSN: 0278-0062            Impact factor:   10.048


  4 in total

Review 1.  Review of SPECT collimator selection, optimization, and fabrication for clinical and preclinical imaging.

Authors:  Karen Van Audenhaege; Roel Van Holen; Stefaan Vandenberghe; Christian Vanhove; Scott D Metzler; Stephen C Moore
Journal:  Med Phys       Date:  2015-08       Impact factor: 4.071

2.  Quantification of Tomographic Incompleteness in Cone-Beam Reconstruction.

Authors:  Rolf Clackdoyle; Frédéric Noo
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2019-05-22

3.  Design of a Multi-Pinhole Collimator for I-123 DaTscan Imaging on Dual-Headed SPECT Systems in Combination with a Fan-Beam Collimator.

Authors:  Michael A King; Joyeeta M Mukherjee; Arda Könik; I George Zubal; Joyoni Dey; Robert Licho
Journal:  IEEE Trans Nucl Sci       Date:  2016-02-03       Impact factor: 1.679

4.  Accurate molecular imaging of small animals taking into account animal models, handling, anaesthesia, quality control and imaging system performance.

Authors:  Christian Vanhove; Jens P Bankstahl; Stefanie D Krämer; Eric Visser; Nicola Belcari; Stefaan Vandenberghe
Journal:  EJNMMI Phys       Date:  2015-11-11
  4 in total