Comparison between continuous non-invasive estimated cardiac output by pulse wave transit time and thermodilution method.

Ashish C Sinha1, Preet Mohinder Singh, Navneet Grewal, Mansoor Aman, Gerald Dubowitz.   

Abstract

AIMS AND
OBJECTIVES: Cardiac output (CO) measurement is essential for many therapeutic decisions in anesthesia and critical care. Most available non-invasive CO measuring methods have an invasive component. We investigate "pulse wave transit time" (estimated continuous cardiac output [esCCO]) a method of CO measurement that has no invasive component to its use.
MATERIALS AND METHODS: After institutional ethical committee approval, 14 adult (21-85 years) patients undergoing surgery and requiring pulmonary artery catheter (PAC) for measuring CO, were included. Postoperatively CO readings were taken simultaneously with thermodilution (TD) via PAC and esCCO, whenever a change in CO was expected due to therapeutic interventions. Both monitoring methods were continued until patients' discharge from the Intensive Care Unit and observer recording values using TD method was blinded to values measured by esCCO system.
RESULTS: Three hundred and one readings were obtained simultaneously from both methods. Correlation and concordance between the two methods was derived using Bland-Altman analysis. Measured values showed significant correlation between esCCO and TD ( r = 0.6, P < 0.001, 95% confidence limits of 0.51-0.68). Mean and (standard deviation) for bias and precision were 0.13 (2.27) L/min and 6.56 (2.19) L/min, respectively. The 95% confidence interval for bias was - 4.32 to 4.58 L/min and for precision 2.27 to10.85 L/min.
CONCLUSIONS: Although, esCCO is the only true non-invasive continuous CO monitor available and even though its values change proportionately to TD method (gold standard) with the present degree of error its utility for clinical/therapeutic decision-making is questionable.

Entities:  

Mesh:

Year:  2014        PMID: 25281622     DOI: 10.4103/0971-9784.142059

Source DB:  PubMed          Journal:  Ann Card Anaesth        ISSN: 0971-9784


  4 in total

Review 1.  [Meta-analyses on measurement precision of non-invasive hemodynamic monitoring technologies in adults].

Authors:  G Pestel; K Fukui; M Higashi; I Schmidtmann; C Werner
Journal:  Anaesthesist       Date:  2018-06       Impact factor: 1.041

2.  Estimated continuous cardiac output based on pulse wave transit time in off-pump coronary artery bypass grafting: a comparison with transpulmonary thermodilution.

Authors:  Alexey A Smetkin; Ayyaz Hussain; Evgenia V Fot; Viktor I Zakharov; Natalia N Izotova; Angelika S Yudina; Zinaida A Dityateva; Yanina V Gromova; Vsevolod V Kuzkov; Lars J Bjertnæs; Mikhail Y Kirov
Journal:  J Clin Monit Comput       Date:  2016-03-07       Impact factor: 2.502

3.  Continuous Estimation of Cardiac Output in Critical Care: A Noninvasive Method Based on Pulse Wave Transit Time Compared with Transpulmonary Thermodilution.

Authors:  Ulrike Ehlers; Rolf Erlebach; Giovanna Brandi; Federica Stretti; Richard Valek; Stephanie Klinzing; Reto Schuepbach
Journal:  Crit Care Res Pract       Date:  2020-07-20

Review 4.  Hemodynamic monitoring in the critically ill: an overview of current cardiac output monitoring methods.

Authors:  Johan Huygh; Yannick Peeters; Jelle Bernards; Manu L N G Malbrain
Journal:  F1000Res       Date:  2016-12-16
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.