Literature DB >> 25276032

Effects of the Electrode Type on N100 and P300 in tDCS Applications.

Jeong-Woo Lee1, Se-Won Yoon1, Woong-Sik Park2, Myoung Heo3, Dong-Geol Lee4, Seong-Kwan Jeong5.   

Abstract

[Purpose] This study investigated the effects of types of electrode on N100 and P300 in transcranial direct current stimulation (tDCS) applications.
[Subjects and Methods] Thirty subjects were randomly assigned to two groups with 15 subjects in each group depending on the electrode types. A positive electrode on the primary motor area (C4) and a negative electrode on the left primary motor area (C3), and stimulation was applied for 20 minutes. Before and after tDCS, N100 and P300 were measured by attaching an electrode to Fp1 and Fp2.
[Results] In tDCS applications, N100 and P300 showed no significant interaction effects between time and group for either latency or amplitude in the Fp1 and Fp 2 areas, but there was a statistically significant difference in the main effect duration.
[Conclusion] The latencies of N100 and P300 were shortened and that their amplitudes increased in both the Fp1 and Fp2 areas, regardless of the type of electrode.

Entities:  

Keywords:  N100; P300; Transcranial direct current stimulation (tDCS)

Year:  2014        PMID: 25276032      PMCID: PMC4175253          DOI: 10.1589/jpts.26.1441

Source DB:  PubMed          Journal:  J Phys Ther Sci        ISSN: 0915-5287


INTRODUCTION

Transcranial direct current stimulation (tDCS) can be utilized as a key tool because it can affect the neural plasticity of the cerebral cortex. Furthermore, it can directly change the excitability of electrical potentials in a stimulated region, as well as indirectly affect the excitability of the same region at the other side of the cerebrum1). Such tDCS can also cause polarity-dependent changes in the excitability of the cerebral motor cortex2). Positive-polarity stimulation increases excitability as depolarization occurs. Negative-polarity stimulation, on the other hand, decreases excitability due to hyperpolarization occurring at the neurons or neural networks in corresponding regions. Moreover, a low intensity current affects excitability at the motor cortex3). The study conducted by Fregni et al.4) revealed that tDCS improved working memory performance in a consecutive sequential letter-matching test, while Antal et al.5) showed in their study that tDCS influenced visual cognitive function in the contrast sensitivity test, a visual function test tDCS-related studies have mostly used carbon rubber electrodes in applications involving scalp electrodes. Carbon rubber electrodes are very inconvenient to use and are disliked by subjects because the face must be wrapped using separate straps, to keep the electrodes in place. Furthermore, the available electrodes are too large to stimulate specific cerebral areas. Therefore, this study was conducted to investigate the effects of electrode type on N100 and P300 when tDCS was applied to a primary motor area.

SUBJECTS AND METHODS

This study selected 30 healthy adult females in their 20’s who had a non-dominant left hand. We obtained approval for this experiment from the research ethics committee of Kwangju Women’s University. The subjects were randomly divided into two groups comprised 15 women each. Group I were administered tDCS with carbon rubber electrodes, while round, self-adhesive electrodes were used for Group II. The general features of the subjects are presented in Table 1.
Table 1.

General characteristics of the subjects

Group I (n=10)Group II (n=10)
Age (years)20.6±0.521.5±1.1
Height (cm)161.3±3.7162.1±1.5
Weight (kg)55.1±8.656.6±7.1

Mean±SD. Group I: Carbon rubber electrode. Group II: A circular adhesive electrode

Mean±SD. Group I: Carbon rubber electrode. Group II: A circular adhesive electrode tDCS was applied using an Endomed 482 (Enraf-Nonius B.V., Rotterdam, Netherlands). In accordance with the international 10–20 system placement method, the positive electrode was placed over the primary motor cortex at C4, and the negative electrode was placed at C3. The pulse duration was set to 2 ms, and the interpulse duration to 5 ms. The entire duration was set to last for a total of 20 minutes. Carbon rubber electrodes 4 × 6 cm2 (Daeyang Medical, Wonju, South Korea) were used for experimental group I with a current intensity of 0.058 mA/cm2, and 1.77 cm2 round, self-adhesive electrodes (SKINTACT, Leonhard Lang GmbH, Innsbruck, Australia) were used for experimental group II with a current intensity of 0.06 mA/cm2. To measure event-related brain potentials (ERP), an EEG-8 electroencephalograph (LXE5208, LAXTHA, Daejeon, South Korea) was used, and the TeleScan software (LAXTHA, Daejeon, South Korea) program was used to conduct the latency and amplitude analyses of N100 and P300. The sampling rate was set to 256 Hz, and band-pass filtered between 1~50 Hz. The attachment regions were wiped clean with alcohol to reduce regional resistance to less than 5 kΩ before the electrodes were attached. The Ag-AgCl electrodes were attached to the Fp1 and Fp2 areas; the active electrode was placed on the right mastoid, and the earth electrode was placed on the left mastoid, in accordance with the international 10–20 system placement method. The ERP, which is the average of brainwaves, appears after stimulating the targets, was analyzed based on the maximum negative potential value observed between 80 and 120 ms for N100 and on the maximum positive potential value between 250 and 500 ms for P300. A serial reaction time task was implemented. Arrow shapes were randomly displayed in two different colors on a monitor. The subjects were requested to move a blue-colored arrows in the arrow direction by their non-dominant left hand, and the red-colored arrows in the direction away from the arrow. Visual stimulation was presented a total of 50 times, among which 18 times were designated for target stimulation (red arrow) and 32 times for non-target stimulation (blue arrow). The temporal interval between each stimulation was set at 1 second, and each time duration was randomly presented between 2 to 4 seconds. Statistical analysis was performed with the SPSS 12.0 for windows. Repeated measures ANOVA was used to analyze the changes in the two groups according to the duration of treatment. A significance level of α=0.05 was chosen.

RESULTS

The results of the experiment show that the N100 latency and amplitude had statistically insignificant differences in the interactions between time and group in both the Fp1 and Fp2 areas. However, a statistically significant difference was found in the main effect duration (p<0.05) in both areas (Table 2). Thus, tDCS application reduced the N100 latency and increased its amplitude, regardless of the type of electrode. The P300 latency and amplitude had statistically insignificant differences in interactions between time and group in both the Fp1 and Fp2 areas. However, a statistically significant difference was found in main effect duration (p<0.05) in both areas (Table 3). Thus, tDCS application reduced the P300 latency and increased its amplitude, regardless of the electrode type.
Table 2.

Changes in N100

EEGGroupLatency (sec)Amplitude (µV)


PrePostPrePost
Fp1I0.14±0.040.13±0.04−6.37±6.91−7.28±7.58
II0.13±0.030.11±0.04−7.58±9.40−8.50±9.88
Fp2I0.14±0.040.13±0.03−6.71±7.23−7.91±7.07
II0.14±0.030.12±0.03−7.49±9.13−8.95±8.16

Mean ± SD. At Fp1 and Fp2, there were only significant changes depending on time in the latency (p<0.05) and amplitude (p<0.05), but there was no statistically significant difference in interaction effect. Group I: Carbon rubber electrode. Group II: A circular adhesive electrode.

Table 3.

Changes in P300

EEGGroupLatency (sec)Amplitude (µV)


PrePostPrePost
Fp1I0.34±0.060.32±0.0512.35±8.7013.51±10.45
II0.33±0.080.32±0.0913.87±15.0414.27±14.97
Fp2I0.34±0.040.31±0.0413.51±10.4517.52±14.05
II0.33±0.090.31±0.1015.27±15.1118.05±18.29

Mean ± SD. At Fp1 and Fp2, there were only significant changes depending on time in the latency (p<0.05) and amplitude (p<0.05), but there was no statistically significant difference in interaction effect. Group I: Carbon rubber electrode. Group II: A circular adhesive electrode.

Mean ± SD. At Fp1 and Fp2, there were only significant changes depending on time in the latency (p<0.05) and amplitude (p<0.05), but there was no statistically significant difference in interaction effect. Group I: Carbon rubber electrode. Group II: A circular adhesive electrode. Mean ± SD. At Fp1 and Fp2, there were only significant changes depending on time in the latency (p<0.05) and amplitude (p<0.05), but there was no statistically significant difference in interaction effect. Group I: Carbon rubber electrode. Group II: A circular adhesive electrode.

DISCUSSION

Utz et al.6) reported that generalized electrode sizes had yet to be established for tDCS; however, some researchers have recommended using small electrodes to effectively transmit current. Our present results demonstrate that tDCS shortens the latencies and increases the amplitudes of N100 and P300 at both the Fp1 and Fp2 areas, regardless of the electrode type. N100 reflects early attentiveness in a cerebral process7); therefore, it considered to be an indicator of selective attentiveness8). In addition, N100 latency represents the complexity and efficiency of synapses involved in information processing speed and responses9). Therefore, we consider tDCS affected N100 latency and amplitude and to have aided the neurophysiologic processes that are related to attentiveness, information processing speed, and the responses of the subjects to external stimuli. Hwang and Lee10) reported that tDCS delivered by carbon rubber electrodes to stroke patients’ primary motor cortex resulted in shorter P300 latency at both the Fp1 and 2 areas. Their result is similar to the result our present study, which used healthy people as subjects. The P300 amplitude represents the amount of cognitive activity, such as memory and attentiveness, and is related to the time it takes to evaluate and classify latencies in response to stimulation11). In addition, P300 also represents the information-gathering process of perceptual decisions that identify and determine stimulation from outside influences12, 13). Therefore, given that both N100 and P300 latencies were shortened, and that their amplitudes increased, regardless of the electrode type, we consider tDCS increased cognitive activities that execute tasks in response to target and non-target stimulation, regardless of the electrode type. Consequently, since tDCS positively influenced cognitive activities, regardless of the electrode type, we consider self-adhesive electrodes are a better treatment choice because of their user-friendliness and convenience in clinical applications of tDCS compared with carbon rubber electrodes.
  10 in total

1.  External modulation of visual perception in humans.

Authors:  A Antal; M A Nitsche; W Paulus
Journal:  Neuroreport       Date:  2001-11-16       Impact factor: 1.837

Review 2.  Clinical application of the P300 event-related brain potential.

Authors:  John Polich
Journal:  Phys Med Rehabil Clin N Am       Date:  2004-02       Impact factor: 1.784

3.  Anodal transcranial direct current stimulation of prefrontal cortex enhances working memory.

Authors:  Felipe Fregni; Paulo S Boggio; Michael Nitsche; Felix Bermpohl; Andrea Antal; Eva Feredoes; Marco A Marcolin; Sergio P Rigonatti; Maria T A Silva; Walter Paulus; Alvaro Pascual-Leone
Journal:  Exp Brain Res       Date:  2005-07-06       Impact factor: 1.972

4.  Contralateral and ipsilateral motor effects after transcranial direct current stimulation.

Authors:  Bradley W Vines; Dinesh G Nair; Gottfried Schlaug
Journal:  Neuroreport       Date:  2006-04-24       Impact factor: 1.837

5.  Shaping the effects of transcranial direct current stimulation of the human motor cortex.

Authors:  M A Nitsche; S Doemkes; T Karaköse; A Antal; D Liebetanz; N Lang; F Tergau; W Paulus
Journal:  J Neurophysiol       Date:  2007-01-24       Impact factor: 2.714

6.  Short and long duration transcranial direct current stimulation (tDCS) over the human hand motor area.

Authors:  Toshiaki Furubayashi; Yasuo Terao; Noritoshi Arai; Shingo Okabe; Hitoshi Mochizuki; Ritsuko Hanajima; Masashi Hamada; Akihiro Yugeta; Satomi Inomata-Terada; Yoshikazu Ugawa
Journal:  Exp Brain Res       Date:  2007-10-17       Impact factor: 1.972

7.  Electrical signs of selective attention in the human brain.

Authors:  S A Hillyard; R F Hink; V L Schwent; T W Picton
Journal:  Science       Date:  1973-10-12       Impact factor: 47.728

8.  Evoked-potential correlates of stimulus uncertainty.

Authors:  S Sutton; M Braren; J Zubin; E R John
Journal:  Science       Date:  1965-11-26       Impact factor: 47.728

Review 9.  Electrified minds: transcranial direct current stimulation (tDCS) and galvanic vestibular stimulation (GVS) as methods of non-invasive brain stimulation in neuropsychology--a review of current data and future implications.

Authors:  Kathrin S Utz; Violeta Dimova; Karin Oppenländer; Georg Kerkhoff
Journal:  Neuropsychologia       Date:  2010-06-11       Impact factor: 3.139

10.  Morphological and latency abnormalities of the mid-latency auditory evoked responses in schizophrenia: a preliminary report.

Authors:  Nashaat N Boutros; Oleg Korzyuko; Glen Oliwa; Alan Feingold; Duane Campbell; Dennis McClain-Furmanski; Frederick Struve; Ben H Jansen
Journal:  Schizophr Res       Date:  2004-10-01       Impact factor: 4.939

  10 in total
  2 in total

1.  Effects of galvanic vestibular stimulation on event related potentials.

Authors:  Jeong-Woo Lee; Woong-Sik Park; Se-Won Yoon
Journal:  J Phys Ther Sci       Date:  2016-09-29

2.  Effect of Anodal-tDCS on Event-Related Potentials: A Controlled Study.

Authors:  Ahmed Izzidien; Sriharasha Ramaraju; Mohammed Ali Roula; Peter W McCarthy
Journal:  Biomed Res Int       Date:  2016-11-10       Impact factor: 3.411

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.