Hypoxic guard systems do not prevent rapid hypoxic inspired mixture formation.

Sofie De Cooman1, Caroline Schollaert, Jan F A Hendrickx, Philip J Peyton, Tom Van Zundert, Andre M De Wolf.   

Abstract

Because a case report and theoretical mass balances suggested that hypoxic guard systems may not prevent the formation of hypoxic inspired mixtures (FIO2 ≤ 21 %) over the clinically used fresh gas flow (FGF) range, we measured FIO2 over a wide range of hypoxic guard limits for O2/N2O and O2/air mixtures. After IRB approval, 16 ASA I-II patients received sevoflurane in either O2/N2O (n = 8) or O2/air (n = 8) using a Zeus(®) anesthesia machine in the conventional mode. After using an 8 L/min FGF with FDO2 = 25% for 10 min, the following hypoxic guard limits were tested for 4 min each, expressed as [total FGF in L/min; FDO2 in %]: [0.3;85], [0.4;65], [0.5;50], [0.7;36], [0.85;30], [1.0;25], [1.25;25], [1.5;25], [2;25], [3;25], [5;25], and [8;25]. In between these [FGF;FDO2] combinations, 8 L/min FGF with 25% O2 was used for 4 min to return to the same baseline FIO2 (25%) before the start of the next combination. This sequence was studied once in each patient receiving O2/air (n = 8), but twice in each patient who received O2/N2O (n = 8) to examine the effect of decreasing N2O uptake over time, resulting in three groups: early O2/N2O, late O2/N2O, and O2/air group. The [FGF;FDO2]-FIO2 relationship was examined. The overall [FGF;FDO2]-FIO2 relationship in the three groups was similar. In all 1, 1.25, and 1.5 L/min FGF groups, FIO2 decreased below 21% in all but one patient; this occurred within 1 min in at least one patient. In the 0.7 L/min O2/air group and the 3 L/min late O2/N2O and O2/air groups, FIO2 decreased below 21% in one patient. Current hypoxic guard systems do not reliably prevent a hypoxic FIO2 with O2/N2O and O2/air mixtures, particularly between 0.7 and 3 L/min.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25270987     DOI: 10.1007/s10877-014-9626-y

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  1 in total

1.  Air-oxygen mixtures in circle systems.

Authors:  J F Hendrickx; S De Cooman; D M Vandeput; J Van Alphen; J Coddens; T Deloof; A M De Wolf
Journal:  J Clin Anesth       Date:  2001-09       Impact factor: 9.452

  1 in total
  3 in total

Review 1.  Journal of Clinical Monitoring and Computing 2015 end of year summary: anesthesia.

Authors:  Jan F A Hendrickx; Andre De Wolf; Stanley Skinner
Journal:  J Clin Monit Comput       Date:  2015-12-26       Impact factor: 2.502

2.  More on the danger of inhaling air at reduced fresh gas flows: a self-experiment.

Authors:  Stellan Eriksson; Bredbacka Sixten
Journal:  J Clin Monit Comput       Date:  2015-06-23       Impact factor: 2.502

3.  Performance of an active inspired hypoxic guard.

Authors:  Idris E Ghijselings; Sofie De Cooman; Rik Carette; Philip J Peyton; Andre M De Wolf; Jan F A Hendrickx
Journal:  J Clin Monit Comput       Date:  2015-03-11       Impact factor: 2.502

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.