Spatial differences in corneal electroretinogram potentials measured in rat with a contact lens electrode array.

Yelena Krakova1, Hadi Tajalli, Sanitta Thongpang, Zahra Derafshi, Tamas Ban, Safa Rahmani, Ashley N Selner, Amani Al-Tarouti, Justin C Williams, John R Hetling.   

Abstract

PURPOSE: It has been known for several decades that the magnitude of the corneal electroretinogram (ERG) varies with position on the eye surface, especially in the presence of focal or asymmetric stimuli or retinal lesions. However, this phenomenon has not been well-characterized using simultaneous measurements at multiple locations on the cornea. This work provides the first characterization of spatial differences in the ERG across the rat cornea.
METHODS: A contact lens electrode array was employed to record ERG potentials at 25 corneal locations simultaneously following brief full-field flash stimuli in normally sighted Long-Evans rats. These multi-electrode electroretinogram (meERG) responses were analyzed for spatial differences in a-wave and b-wave amplitudes and implicit times.
RESULTS: Spatially distinct ERG potentials could be recorded reliably. Comparing relative amplitudes across the corneal locations suggested a slight non-uniform distribution when using full-field, near-saturating stimuli. Amplitudes of a- and b-waves were approximately 3 % lower in the inferior quadrant than in the superior quadrant of the cornea.
CONCLUSIONS: The present results comprise the start of the first normative meERG database for rat eyes and provide a basis for comparison of results from eyes with functional deficit. Robust measures of spatial differences in corneal potentials will also support optimization and validation of computational source models of the ERG. To fully utilize the information contained in the meERG data, a detailed understanding of the roles of the many determinants of local corneal potentials will eventually be required.

Entities:  

Mesh:

Year:  2014        PMID: 25266461      PMCID: PMC4219023          DOI: 10.1007/s10633-014-9459-5

Source DB:  PubMed          Journal:  Doc Ophthalmol        ISSN: 0012-4486            Impact factor:   2.379


  24 in total

1.  Recording of the human electroretinogram with the contact glass; influence of the position of the recording electrode on the b-potential.

Authors:  E SUNDMARK
Journal:  Acta Ophthalmol (Copenh)       Date:  1958

2.  ISCEV Standard for full-field clinical electroretinography (2008 update).

Authors:  M F Marmor; A B Fulton; G E Holder; Y Miyake; M Brigell; M Bach
Journal:  Doc Ophthalmol       Date:  2008-11-22       Impact factor: 2.379

3.  A computerized analysis of the entire retinal ganglion cell population and its spatial distribution in adult rats.

Authors:  M Salinas-Navarro; S Mayor-Torroglosa; M Jiménez-López; M Avilés-Trigueros; T M Holmes; R D Lund; M P Villegas-Pérez; M Vidal-Sanz
Journal:  Vision Res       Date:  2008-11-17       Impact factor: 1.886

Review 4.  ISCEV standard for clinical pattern electroretinography (PERG): 2012 update.

Authors:  Michael Bach; Mitchell G Brigell; Marko Hawlina; Graham E Holder; Mary A Johnson; Daphne L McCulloch; Thomas Meigen; Suresh Viswanathan
Journal:  Doc Ophthalmol       Date:  2012-10-17       Impact factor: 2.379

Review 5.  ISCEV standard for clinical multifocal electroretinography (mfERG) (2011 edition).

Authors:  Donald C Hood; Michael Bach; Mitchell Brigell; David Keating; Mineo Kondo; Jonathan S Lyons; Michael F Marmor; Daphne L McCulloch; Anja M Palmowski-Wolfe
Journal:  Doc Ophthalmol       Date:  2011-10-30       Impact factor: 2.379

6.  Cytoarchitecture of the retinal ganglion cells in the rat.

Authors:  John Danias; Fran Shen; David Goldblum; Bin Chen; Jerome Ramos-Esteban; Steven M Podos; Thom Mittag
Journal:  Invest Ophthalmol Vis Sci       Date:  2002-03       Impact factor: 4.799

7.  Three-dimensional electromagnetic model of the human eye: advances towards the optimisation of electroretinographic signal detection.

Authors:  H M Job; D Keating; A L Evans; S Parks
Journal:  Med Biol Eng Comput       Date:  1999-11       Impact factor: 2.602

8.  Detection of early hydroxychloroquine retinal toxicity enhanced by ring ratio analysis of multifocal electroretinography.

Authors:  Jonathan S Lyons; Matthew L Severns
Journal:  Am J Ophthalmol       Date:  2007-03-06       Impact factor: 5.258

9.  Using multifocal ERG ring ratios to detect and follow Plaquenil retinal toxicity: a review : Review of mfERG ring ratios in Plaquenil toxicity.

Authors:  Jonathan S Lyons; Matthew L Severns
Journal:  Doc Ophthalmol       Date:  2008-05-09       Impact factor: 2.379

10.  Automated quantification and topographical distribution of the whole population of S- and L-cones in adult albino and pigmented rats.

Authors:  Arturo Ortín-Martínez; Manuel Jiménez-López; Francisco M Nadal-Nicolás; Manuel Salinas-Navarro; Luis Alarcón-Martínez; Yves Sauvé; Maria Paz Villegas-Pérez; Manuel Vidal-Sanz; Marta Agudo-Barriuso
Journal:  Invest Ophthalmol Vis Sci       Date:  2010-01-13       Impact factor: 4.799

View more
  2 in total

1.  Fabrication and utility of a transparent graphene neural electrode array for electrophysiology, in vivo imaging, and optogenetics.

Authors:  Dong-Wook Park; Sarah K Brodnick; Jared P Ness; Farid Atry; Lisa Krugner-Higby; Amelia Sandberg; Solomon Mikael; Thomas J Richner; Joseph Novello; Hyungsoo Kim; Dong-Hyun Baek; Jihye Bong; Seth T Frye; Sanitta Thongpang; Kyle I Swanson; Wendell Lake; Ramin Pashaie; Justin C Williams; Zhenqiang Ma
Journal:  Nat Protoc       Date:  2016-10-13       Impact factor: 13.491

2.  Soft transparent graphene contact lens electrodes for conformal full-cornea recording of electroretinogram.

Authors:  Rongkang Yin; Zheng Xu; Ming Mei; Zhaolong Chen; Kai Wang; Yanlin Liu; Tao Tang; Manish Kr Priydarshi; Xuejuan Meng; Siyuan Zhao; Bing Deng; Hailin Peng; Zhongfan Liu; Xiaojie Duan
Journal:  Nat Commun       Date:  2018-06-13       Impact factor: 14.919

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.