Literature DB >> 25232087

The relation between titanium taper corrosion and cobalt-chromium bearing wear in large-head metal-on-metal total hip prostheses: a retrieval study.

F Witt1, B H Bosker2, N E Bishop1, H B Ettema2, C C P M Verheyen2, M M Morlock1.   

Abstract

BACKGROUND: Revision of hip implants due to adverse tissue reactions to metal debris has been associated with wear and corrosion of the metal-on-metal bearing articulation and the modular taper interface. Bearing articulation wear is increased in conditions of poor lubrication, which can also lead to high friction moments that may cause corrosion at the taper interface. This suggests that wear of the bearing and increased corrosion of the taper interface should occur simultaneously, which was investigated in this study.
METHODS: Forty-three large-diameter cobalt-chromium bearings of the same design, implanted with a titanium stem using a titanium adapter, were retrieved at revision at a single center. Retrievals were grouped according to visual inspection of the female taper surface of the adapter into slight and severe corrosion groups. Volume change of bearing and taper surfaces was assessed using a coordinate measurement machine. Serum ion concentrations were determined for forty-three patients, whereas tissue metal concentration was measured for twelve patients.
RESULTS: Severe taper corrosion was observed in 30% of the retrievals. Corrosion was observed either as material deposition or wear. The overall bearing wear rate was significantly higher in the group with severe taper corrosion than in the group with slight corrosion (7.2 ± 9.0 mm(3)/yr versus 3.1 ± 6.8 mm(3)/yr, respectively; p = 0.023) as were the serum cobalt (40.5 ± 44.9 μg/L versus 15.2 ± 23.9 μg/L, respectively; p = 0.024) and chromium ion concentrations (32.7 ± 32.7 μg/L versus 12.0 ± 15.1 μg/L, respectively; p = 0.019). Serum metal ion concentrations were more consistent indicators of wear than tissue metal concentrations.
CONCLUSIONS: The increased bearing articulation wear and serum metal ion concentrations in cases with taper interface corrosion support the hypothesis that increased friction in the joint articulation is one of the factors responsible for simultaneous articulation and taper damage. However, independent taper or bearing damage was also observed, suggesting that other factors are involved in the process.
Copyright © 2014 by The Journal of Bone and Joint Surgery, Incorporated.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25232087     DOI: 10.2106/JBJS.M.01199

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  14 in total

1.  Low failure rate at short term for 40 mm heads and second generation triple annealed HCLPE liners in hybrid hip replacements.

Authors:  Rajkumar Thangaraj; Jan Kuiper; Ralph D Perkins
Journal:  J Clin Orthop Trauma       Date:  2018-01-03

2.  Primary hip replacement stem taper fracture due to corrosion in 3 patients.

Authors:  Michael M Morlock; Dennis Bünte; Harmen Ettema; Cees C Verheyen; Åke Hamberg; Jeremy Gilbert
Journal:  Acta Orthop       Date:  2016-01-11       Impact factor: 3.717

3.  The innovation trap.

Authors:  Nils P Hailer
Journal:  Acta Orthop       Date:  2016-02-06       Impact factor: 3.717

4.  Histopathological characterization of corrosion product associated adverse local tissue reaction in hip implants: a study of 285 cases.

Authors:  Benjamin F Ricciardi; Allina A Nocon; Seth A Jerabek; Gabrielle Wilner; Elianna Kaplowitz; Steven R Goldring; P Edward Purdue; Giorgio Perino
Journal:  BMC Clin Pathol       Date:  2016-02-27

5.  Clinical and Wear Analyses of 9 Large Metal-on-Metal Total Hip Prostheses.

Authors:  M C Koper; N M C Mathijssen; F Witt; M M Morlock; S B W Vehmeijer
Journal:  PLoS One       Date:  2016-10-06       Impact factor: 3.240

Review 6.  Bearings in Hip Arthroplasty: Joint Registries vs Precision Medicine: Review Article.

Authors:  Mark J Pearson; Liam M Grover; Janet M Lord; Simon W Jones; Edward T Davis
Journal:  HSS J       Date:  2016-11-14

Review 7.  Sequelae of large-head metal-on-metal hip arthroplasties: Current status and future prospects.

Authors:  Christiaan P van Lingen; Luigi M Zagra; Harmen B Ettema; Cees C Verheyen
Journal:  EFORT Open Rev       Date:  2017-03-13

8.  Head size in primary total hip arthroplasty.

Authors:  Georgios Tsikandylakis; Maziar Mohaddes; Peter Cnudde; Antti Eskelinen; Johan Kärrholm; Ola Rolfson
Journal:  EFORT Open Rev       Date:  2018-05-21

9.  Quantification of the Contact Area at the Head-Stem Taper Interface of Modular Hip Prostheses.

Authors:  Florian Witt; Julian Gührs; Michael M Morlock; Nicholas E Bishop
Journal:  PLoS One       Date:  2015-08-17       Impact factor: 3.240

10.  Three-dimensional friction measurement during hip simulation.

Authors:  Robert Sonntag; Steffen Braun; Loay Al-Salehi; Joern Reinders; Ulrike Mueller; J Philippe Kretzer
Journal:  PLoS One       Date:  2017-09-08       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.