Enhancement of holding strength of cannulated screw supported with PMMA: a biomechanical study on femoral head [corrected].

Reşad Zeynalov1, İsmail Ağır, Ahmet Hamdi Akgülle, Barış Kocaoğlu, Mithat Selim Yalçın.   

Abstract

OBJECTIVE: The aim of this study was to evaluate the holding strength of cannulated screw with multiple holes on threaded area, supported with PMMA in femoral head.
MATERIAL AND METHODS: A total of 48 human femoral heads were divided into two groups after mineral density measurement with Q-CT. Seven-millimeter cannulated screws with multiple holes on threaded area supported with PMMA were used in the study group, while in the control group standard 7-mm cannulated screws were used. Each group was divided into three subgroups with eight femoral heads. Mineral density of each subgroup was equal to the other. Groups were compared in terms of pull-out, maximum extraction torque and cut-out.
RESULTS: In pull-out group, maximum holding strength (N) was measured, while axial pull-out of 0.5 mm/sec applied with Instron. Results showed meaningful significant difference (p < 0.011) between two groups. In cut-out group, femoral heads were placed into Instron and loading was started from 5 N at 2 mm per minute at first, and it was continued until a failure, at least 5 mm, of implant was observed. Results showed significant difference (p < 0.05) between two groups. In maximum extraction group, 4° per second reverse torque (Nm) was applied with torque meter. Highest torque value was measured during extraction time, and results showed very significant difference (p < 0. 001) between two groups.
CONCLUSION: The results of our new design of cannulated screw augmented with PMMA provided background data to clinical application.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25208652     DOI: 10.1007/s00590-014-1525-x

Source DB:  PubMed          Journal:  Eur J Orthop Surg Traumatol        ISSN: 1633-8065


  20 in total

1.  Stability of internally fixed femoral neck fractures augmented with resorbable cement. A prospective randomized study using radiostereometry.

Authors:  P Mattsson; S Larsson
Journal:  Scand J Surg       Date:  2003       Impact factor: 2.360

2.  Calcium phosphate cement for augmentation did not improve results after internal fixation of displaced femoral neck fractures: a randomized study of 118 patients.

Authors:  Per Mattsson; Sune Larsson
Journal:  Acta Orthop       Date:  2006-04       Impact factor: 3.717

3.  The stability of femoral neck fractures and its influence on healing.

Authors:  L Rehnberg; C Olerud
Journal:  J Bone Joint Surg Br       Date:  1989-03

4.  Augmentation of femoral neck fracture fixation with an injectable calcium-phosphate bone mineral cement.

Authors:  C J Stankewich; M F Swiontkowski; A F Tencer; D N Yetkinler; R D Poser
Journal:  J Orthop Res       Date:  1996-09       Impact factor: 3.494

5.  Outcomes after displaced fractures of the femoral neck. A meta-analysis of one hundred and six published reports.

Authors:  G L Lu-Yao; R B Keller; B Littenberg; J E Wennberg
Journal:  J Bone Joint Surg Am       Date:  1994-01       Impact factor: 5.284

6.  Hip screw augmentation with an in situ-setting calcium phosphate cement: an in vitro biomechanical analysis.

Authors:  D C Moore; E P Frankenburg; J A Goulet; S A Goldstein
Journal:  J Orthop Trauma       Date:  1997-11       Impact factor: 2.512

Review 7.  Complications after intracapsular hip fractures in young adults. A meta-analysis of 18 published studies involving 564 fractures.

Authors:  D S Damany; Martyn J Parker; Adrian Chojnowski
Journal:  Injury       Date:  2005-01       Impact factor: 2.586

8.  Resorbable cement for the augmentation of internally-fixed unstable trochanteric fractures. A prospective, randomised multicentre study.

Authors:  P Mattsson; A Alberts; G Dahlberg; M Sohlman; H C Hyldahl; S Larsson
Journal:  J Bone Joint Surg Br       Date:  2005-09

9.  Biomechanical evaluation of a new augmentation method for enhanced screw fixation in osteoporotic proximal femoral fractures.

Authors:  P von der Linden; A Gisep; V Boner; M Windolf; A Appelt; N Suhm
Journal:  J Orthop Res       Date:  2006-12       Impact factor: 3.494

10.  Vertically oriented femoral neck fractures: mechanical analysis of four fixation techniques.

Authors:  Arash Aminian; Fan Gao; Wasyl W Fedoriw; Li-Qun Zhang; David M Kalainov; Bradley R Merk
Journal:  J Orthop Trauma       Date:  2007-09       Impact factor: 2.512

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.