Akbar Khalilpour1, Seyed Mahmoud Sadjjadi2, Zohreh Kazemi Moghadam1, Muhammad Hafiznur Yunus1, Nor Dyana Zakaria1, Sabariah Osman1, Rahmah Noordin2.
Abstract
Cystic echinococcosis (CE) caused by infection with Echinococcus granulosus is of major concern for humans in many parts of the world. Antigen B was prepared from E. granulosus hydatid fluid, and Western blots confirmed eight batches showing a band corresponding to the 8-/12-kDa subunit with positive serum and no low-molecular mass band (< 15 kDa) with negative serum. The batches were pooled and used to prepare lateral flow immunoglobulin G4 (IgG4) and IgG dipsticks. Diagnostic sensitivity was determined using serum samples from 21 hydatidosis patients, and diagnostic specificity was established using sera from 17 individuals infected with other parasites and 15 healthy people. IgG4 dipstick had a diagnostic sensitivity of 95% (20 of 21) and a specificity of 100% (32 of 32). The IgG dipstick had a sensitivity of 100% (21 of 21) and a specificity of 87.5% (28 of 32). Thus, both IgG and IgG4 dipsticks had high sensitivities, but IgG4 had greater specificity for the diagnosis of human CE. © The American Society of Tropical Medicine and Hygiene.Entities:
Mesh:
Substances:
Year: 2014 PMID: 25200268 PMCID: PMC4228898 DOI: 10.4269/ajtmh.14-0170
Source DB: PubMed Journal: Am J Trop Med Hyg ISSN: 0002-9637 Impact factor: 2.345