False-positive reactions against HLA class II molecules detected in Luminex single-antigen bead assays.

Ji Won In1, Eun Youn Rho1, Sue Shin1, Kyoung Un Park1, Eun Young Song1.   

Abstract

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25187899      PMCID: PMC4151015          DOI: 10.3343/alm.2014.34.5.408

Source DB:  PubMed          Journal:  Ann Lab Med        ISSN: 2234-3806            Impact factor:   3.464


× No keyword cloud information.
Anti-donor HLA-specific antibodies (DSA) are associated with poor graft outcomes in renal transplantations [1, 2]. Panel reactive antibody (PRA) assays using the Luminex platform are commonly used to detect DSA. PRA assays are performed by using three different panels: 1) pooled antigen panels that use bead populations coated in affinity-purified HLA molecules obtained from multiple cell lines, which are used as screening tests; 2) phenotype panels that use bead populations bearing HLA proteins from a cell line derived from a single individual; and 3) single-antigen bead (SAB) assays that use beads coated in molecules representing a single cloned allelic HLA antigen [3]. Each bead population in phenotype panels features more than one type of HLA molecule, thus requiring expertise to interpret the results, whereas SAB assays provide accurate identification of HLA antibodies [3]. However, cloned HLA antigens are not in their native state. Purification and bead coating can lead to conformational changes that could cause binding of clinically irrelevant antibodies [4, 5, 6, 7, 8, 9]. Here, we report a case with suspected false reactions against denatured HLA class II molecules in SAB assays. A 74-yr-old male patient who was monitored after penetrating keratoplasty and corneal allograft (November 7, 2012) was referred for PRA screening for anti-HLA antibodies. PRA screening was performed using the LIFECODES LifeScreen Deluxe kit (Immucor, Stamford, CT, USA). The serum drawn on October 17, 2014 (serum No. #2001) was negative for anti-HLA, but showed a slight increase in median fluorescence intensity (MFI) value in one of the four cross-reactive class II group beads. On the following day, a SAB assay was performed on newly collected serum (serum No. #2004) by using the LIFECODES LSA Class II Single Antigen kit (Immucor, lot. 10102A). The following positive reactions with HLA antibody specificities were obtained: 1) moderate DR11 and 14; 2) weak DR1, 3, 4, 7, 9, 10, 12, 13, 15, and 16; and 3) DR52 and 53 (MFI >1,000) (Table 1; moderate: MFI 3,000-9,999; weak: MFI 1,000-2,999). Two months later, a follow-up test performed using the LIFECODES Class II ID kit (Immucor) was found to be negative (serum No. #2397). A repeat test was performed on serum #2004 by using the LIFECODES LSA Class II Single Antigen kit to rule out clerical errors; the results obtained were the same as those obtained using the SAB assay performed previously. LIFECODES Screen and Class II ID tests were performed on serum #2004 and were negative. Finally, the LIFECODES Screen test performed on serum #2397 showed negative results, and a LIFECODES LSA Single Antigen test performed on serum #2397 by using a kit with a different lot number (06113A) showed results consistent with those of the previous SAB assay.
Table 1

HLA Class II PRA and B-flow cytometric crossmatch results for patient (#2001, #2004, and #2397) and control (#0054) sera

*Result of the retest was same as that of the routine test.

Abbreviations: PRA, panel reactive antibody; Screen, LIFECODES LifeScreen Deluxe; B-FCXM, B cell-flow cytometric crossmatch; NT, not tested; MFI, median fluorescence intensity.

To rule out false-positive reactions in the SAB assays, flow cytometric crossmatches (FCXM) were performed on serum #2004 and serum from a second patient (#0054) with similar HLA antibody specificities as control: moderate anti-DR11 and 13; weak DR1, 3, 4, 12 + extras; and no HLA class I antibody in PRA ID tests. Lymphocytes from two donors with HLA-DR11, 13 and HLA-DR11, 14 were used for FCXM analysis. Serum #0054 showed positive B-cell (B)-FCXM to both donors, but serum #2004 was negative. In a previous study, we showed 90% of sera containing DSA with moderate MFI (3,000-9,999) in LIFECODES ID tests showed positive B-FCXM [10]. Because the LIFECODES LSA Single Antigen and LIFECODES ID MFI values were similar in our laboratory, the negative B-FCXM result on serum #2004 suggests moderate DSA is less likely. The patient showed a clear corneal allograft at the final follow-up on February 17, 2014. So-called "natural" antibodies reacting with dissociated forms of HLA class I antigens have been reported in healthy male blood donors [8]. These antibodies could be anti-HLA-E, which is frequently found in healthy people [11] and usually shows low MFI values (less than 1,000) in SAB assays [8]. However, unpredictable false-positive reactions to denatured HLA antigens in SAB assays can show moderate to strong reactivity to both HLA class I [4, 5, 7] and class II antigens [6], which can lead to unnecessary exclusion of organ donors or immunosuppressive therapy. A patient with negative FCXM but positive DSA to denatured anti-A*01:01 in SAB assays, who received a kidney transplantation, had no rejection episodes in the 5 yr following transplantation [4]. Moreover, a patient with DSA to denatured HLA-A*02:01 showed no rejection one year after heart transplantation [5]. Among 8,121 consecutive samples tested using the LABScreen Single Antigen kit (One Lambda, Canoga Park, CA, USA), 141 sera samples (1.4%) showed an atypical reactivity to denatured HLA-DRB1*09:01, DRB3*01:01, *02:02, *03:01, DPB1*20:01, and *28:01 [6]. While most previous studies were performed using the LABScreen Single Antigen kit [4, 5, 6], one study reported identification of antibodies to denatured HLA-B*44:02 by using the LIFECODES LSA Single Antigen test [7]. To the best of our knowledge, we are the first to report identification of suspected antibodies to denatured HLA class II antigens by using the LIFECODES LSA Single Antigen kit, which is widely used in Korea. The limitation of our study is that we did not confirm that the reactivity was specifically toward denatured, but not intact antigens. Further, other possible reasons for false-positive SAB assay reactions could not be ruled out. However, we suggest that SAB assays should be used in conjunction with other types of PRA tests that use extracted HLA molecules to avoid unnecessary barriers to transplantation and inappropriate use of immunosuppressive therapy.
  10 in total

Review 1.  The impact of donor-specific anti-HLA antibodies on late kidney allograft failure.

Authors:  Alexandre Loupy; Gary S Hill; Stanley C Jordan
Journal:  Nat Rev Nephrol       Date:  2012-04-17       Impact factor: 28.314

2.  Antibodies to HLA-E in nonalloimmunized males: pattern of HLA-Ia reactivity of anti-HLA-E-positive sera.

Authors:  Mepur H Ravindranath; Hugo Kaneku; Nadim El-Awar; Luis E Morales-Buenrostro; Paul I Terasaki
Journal:  J Immunol       Date:  2010-07-07       Impact factor: 5.422

3.  Detecting and monitoring human leukocyte antigen-specific antibodies.

Authors:  Andrea A Zachary; M Sue Leffell
Journal:  Hum Immunol       Date:  2008-08-08       Impact factor: 2.850

4.  Epitopes of human leukocyte antigen class I antibodies found in sera of normal healthy males and cord blood.

Authors:  Nadim El-Awar; Paul I Terasaki; Anh Nguyen; Nori Sasaki; Luis E Morales-Buenrostro; Hiroh Saji; Etsuko Maruya; Francesca Poli
Journal:  Hum Immunol       Date:  2009-07-04       Impact factor: 2.850

5.  Heart transplantation with donor-specific antibodies directed toward denatured HLA-A*02:01: a case report.

Authors:  Francesca Poli; Elena Benazzi; Annalisa Innocente; Angela Nocco; Nicoletta Cagni; Andrea Gianatti; Roberto Fiocchi; Mario Scalamogna
Journal:  Hum Immunol       Date:  2011-08-10       Impact factor: 2.850

6.  Positive virtual crossmatch with negative flow crossmatch results in two cases.

Authors:  Eapen K Jacob; Steve R De Goey; Manish J Gandhi
Journal:  Transpl Immunol       Date:  2011-05-23       Impact factor: 1.708

7.  Antibodies against denatured HLA class II molecules detected in luminex-single antigen assay.

Authors:  Patricia C Grenzi; Renato de Marco; Rosemeire Z R Silva; Erika F Campos; Maria Gerbase-DeLima
Journal:  Hum Immunol       Date:  2013-07-02       Impact factor: 2.850

Review 8.  Donor-specific antibodies adversely affect kidney allograft outcomes.

Authors:  Sumit Mohan; Amudha Palanisamy; Demetra Tsapepas; Bekir Tanriover; R John Crew; Geoffrey Dube; Lloyd E Ratner; David J Cohen; Jai Radhakrishnan
Journal:  J Am Soc Nephrol       Date:  2012-11-15       Impact factor: 10.121

9.  Challenging the golden standard in defining donor-specific antibodies: does the solid phase assay meet the expectations?

Authors:  Y M Zoet; S H Brand-Schaaf; D L Roelen; A Mulder; F H J Claas; I I N Doxiadis
Journal:  Tissue Antigens       Date:  2011-03

10.  Donor-specific antibody against denatured HLA-A1: clinically nonsignificant?

Authors:  Shalini Pereira; Susan Perkins; Jar-How Lee; Wayne Shumway; William LeFor; Mayra Lopez-Cepero; Cynthia Wong; Amy Connolly; Jane C Tan; F Carl Grumet
Journal:  Hum Immunol       Date:  2011-03-09       Impact factor: 2.850

  10 in total
  3 in total

Review 1.  Alloantibody Generation and Effector Function Following Sensitization to Human Leukocyte Antigen.

Authors:  Michelle J Hickey; Nicole M Valenzuela; Elaine F Reed
Journal:  Front Immunol       Date:  2016-02-04       Impact factor: 7.561

2.  Technical Limitations of the C1q Single-Antigen Bead Assay to Detect Complement Binding HLA-Specific Antibodies.

Authors:  Craig J Taylor; Vasilis Kosmoliaptsis; Jessie Martin; Graham Knighton; Dermot Mallon; J Andrew Bradley; Sarah Peacock
Journal:  Transplantation       Date:  2017-06       Impact factor: 4.939

3.  False Positive Class II HLA Antibody Reaction Due to Antibodies Against Denatured HLA Might Differ Between Assays: One Lambda vs. Immucor.

Authors:  Borae G Park; Younhee Park; Beom Seok Kim; Yu-Seun Kim; Hyon-Suk Kim
Journal:  Ann Lab Med       Date:  2020-09       Impact factor: 3.464

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.