Literature DB >> 25184817

Differential ion mobility spectrometry coupled to tandem mass spectrometry enables targeted leukemia antigen detection.

Udara Dharmasiri1, Samantha L Isenberg, Gary L Glish, Paul M Armistead.   

Abstract

Differential ion mobility spectrometry (DIMS) can be used as a filter to remove undesired background ions from reaching the mass spectrometer. The ability to use DIMS as a filter for known analytes makes DIMS coupled to tandem mass spectrometry (DIMS-MS/MS) a promising technique for the detection of cancer antigens that can be predicted by computational algorithms. In experiments using DIMS-MS/MS that were performed without the use of high-performance liquid chromatography (HPLC), a predicted model antigen, GLR (FLSSANEHL), was detected at a concentration of 10 pM (20 amol) in a mixture containing 94 competing model peptide antigens, each at a concentration of 1 μM. Without DIMS filtering, the GLR peptide was undetectable in the mixture even at 100 nM. Again, without using HPLC, DIMS-MS/MS was used to detect 2 of 3 previously characterized antigens produced by the leukemia cell line U937.A2. Because of its sensitivity, a targeted DIMS-MS/MS methodology can likely be used to probe for predicted cancer antigens from cancer cell lines as well as human tumor samples.

Entities:  

Keywords:  DIMS; Differential ion mobility spectrometry; FAIMS; cancer antigen; mass spectrometry

Mesh:

Substances:

Year:  2014        PMID: 25184817      PMCID: PMC4184456          DOI: 10.1021/pr500527c

Source DB:  PubMed          Journal:  J Proteome Res        ISSN: 1535-3893            Impact factor:   4.466


Introduction

Differential ion mobility spectrometry (DIMS) provides a gas-phase separation of ions prior to mass analysis. DIMS has the ability to separate isobaric and isomeric species and therefore can improve the specificity and detection limits of mass spectrometry-based analyses.[1−3] Low electric field ion mobility separations take advantage of the difference in the collisional cross-section of ions, which is directly proportional to the low-field ion mobility.[4,5] At low electric fields the mobility of an ion is independent of the field. However, at high electric fields (>104 V/cm), the mobility of an ion becomes dependent on the electric field in a complex function that is still not understood.[4,5] DIMS takes advantage of the change in ion mobility in high electric fields to separate ions based on the difference in ion mobility between low and high electric fields. Differential ion mobility spectrometers are composed of two parallel electrodes. The gas flow into the mass spectrometer carries ions through the gap between the DIMS electrodes. In “transparent” mode, both of the electrodes are held at the same potential, and all ions are allowed to pass through the device without an ion mobility separation. In “active” mode, an asymmetric waveform is applied to the electrodes. The DIMS waveform alternates between low and high electric fields of opposing polarity, such that the ions zigzag through the device. The V0-P of the DIMS waveform measured during the high electric field portion is called the dispersion voltage (DV). Because ions have a different mobility in the high field versus low field, they are displaced toward one or the other electrodes during transit through the device. To direct ions of a selected differential ion mobility into the mass spectrometer, a compensation voltage (CV) must be applied to one of the electrodes, which counter balances the displacement due to the differential mobility. Because high-field ion mobility is not yet well understood, the optimum CV and DV for a given analyte are not easily predicted. This makes DIMS currently most useful for targeted analyses, where optimal DIMS conditions for the selection of the analyte of interest can be determined prior to analysis of the real sample. While there is a loss in overall ion signal using DIMS, the background signal (chemical noise) is typically reduced to a much greater extent than the analyte signal because background ions are eliminated through collisions with the DIMS electrodes while target ions are not. The net result is an overall increase in signal-to-noise (S/N). One potential application for DIMS is the identification of cancer antigens. These antigens consist of a 9–10 amino acid peptide fragment, derived from a cancer associated protein, which is complexed to a class I human leukocyte antigen (HLA) molecule. The peptide/HLA complexes are assembled in the endoplasmic reticulum and subsequently transported to and presented on the cancer cell’s surface. A cytotoxic CD8+ T-cell expressing a T-cell receptor with high binding affinity for the specific peptide/HLA complex can kill the antigen-presenting cells (i.e., cancer cell) through release of cytotoxic granule proteins.[6−8] Because of the high activity and specificity of cancer-antigen specific CD8+ T-cells and their ability to persist in patients indefinitely, many forms of cancer immunotherapy have been developed to harness this approach in various cancers including melanoma, renal cell carcinoma, prostate cancer, and leukemia.[9−12] The identification of appropriate peptide antigens is a vital part of the development of new T-cell therapies because of the complexity and costs involved with the production and development of these therapeutics and the risks involved with directing T-cells to noncancer (i.e., normal tissue) targets.[13,14] The majority of cancer antigenic peptides have been confirmed using nontargeted reverse phase high-performance liquid chromatography coupled to tandem mass spectrometry (HPLC–MS) approaches on a complex mixture of antigenic peptides obtained from a cellular sample.[15−20] While this approach has proven extremely useful for peptide discovery, nontargeted HPLC–MS preferentially identifies the most prevalent peptides, which are not necessarily the most immunogenic.[21] In recent years computational algorithms have been developed to predict which peptides in a protein are likely to bind specific HLA molecules.[22−26] These algorithms, when combined with the significant advances in genomics and computational biology, can be used to predict the relatively small number of antigenic peptides in tumor extracts that are likely immunogenic and cancer-specific because they are produced as a result of cancer-specific processes such as somatic gene mutations, fusions, overexpression, and alternative splicing.[27,28] By employing a computationally predictive cancer-antigen discovery approach, targeted MS methodologies aimed at detecting cancer-specific peptide antigens would facilitate the discovery of high-value cancer peptide antigens. In this study, a targeted antigen-discovery strategy employing DIMS–MS/MS was used to identify known cancer-associated antigenic peptides from a human leukemia cell line without using HPLC.

Experimental Section

Model Peptide Pool

A panel of 95 “Flashpure” peptides (New England Peptide, Gardner MA) that had been computationally predicted to bind with high affinity to HLA-A*0201 (Supplementary Table S-1) were dissolved in 100% DMSO to a concentration of 20 mM and subsequently diluted to 1 μM using 50/49.9/0.1 ACN/H2O/FA (V/V/V) (Sigma-Aldrich, St. Louis, MO) (Supplementary Table S-1). To create a peptide pool, 94 of the 20 mM peptide solutions in DMSO were combined in equal volumes and diluted to a final concentration of 1 μM of each peptide in 50/49.9/0.1 ACN/H2O/FA (V/V/V). The 95th peptide, GLR (amino acid sequence FLSSANEHL), was further diluted using 50/49.9/0.1 ACN/H2O/FA (V/V/V) and spiked into the 94 peptide pool to make 1 μM, 100 nM, 10 nM, 100 pM, 10 pM, and 1 pM concentrations of GLR peptide solutions.

Leukemia Cell Line Derived Peptide Mixture

The U937 leukemia cell line stably transfected with HLA-A*0201 (called U937.A2) were grown under sterile conditions at 37 °C with 5% CO2 in a medium consisting of RPMI 1640 supplemented with 10% heat inactivated (56 °C for 30 min) fetal bovine serum (FBS), 1% penicillinstreptomycin, and 1% l-glutamine.[29] Four liters of U937.A2 cell culture suspension was incubated until the live cell density reached ∼7.5 × 105 cells/mL. The cell suspension (approximately 2 to 3 × 109 U937.A2 cells) was centrifuged at 4 °C for 30 min at 1250 rpm to obtain a cell pellet, and then the pelleted cells were lysed in 20 mM Tris-HCl, pH 8, 150 mM NaCl, 1% (mass/volume) 3-[(3-cholamidopropyl) dimethylammonio]-1-propanesulfonate (CHAPS), 1 mM (phenylmethylsulfonyl fluoride), 10 μg/mL pepstatin, 5 μg/mL aprotinin, and 10 μg/mL leupeptin (Sigma-Aldrich, St. Louis, MO). The mixture was gently rocked in a laboratory shaker for 15 min at 4 °C to break up the cells. The cell lysate was subjected to ultracentrifugation at 40,000 rpm for 1 h to remove cell debris.[30] All protein loading and elution steps were then performed using a Chemyx Fusion 100 syringe pump (Stafford, TX). The cell pellet consisted of ∼2 × 109 cells of U937.A2 with each cell estimated to have ∼10,000 HLA-A*0201 molecules.[31,32] For efficient isolation of nearly all the HLA-A*0201 molecules in the lysate, 100× excess molecules (3.3 × 10–9 moles) of the anti-HLA-A*0201 antibody, BB7.2, were introduced into a 5 mL GE Healthcare HiTRAP recombinant protein A column (Chicago, IL) for the immobilization. The BB7.2 antibody molecules were reconstituted in 200 mM Tris HCl (pH 8) (Sigma-Aldrich, St. Louis, MO). For effective antibody immobilization, first, the column was washed with 50 mM Tris (pH 8) for 5 min at 1 mL/min flow rate. Then, the reconstituted BB7.2 solution (1 mL) was injected into the column at 0.2 mL/min flow rate for 5 min. Excess antibody was washed out using 200 mM Tris HCl (pH 8) at 1 mL/min for 5 min. The HLA-A*0201/peptide complex containing supernatant was introduced into the BB7.2 antibody immobilized column at 1 mL/min for immunoaffinity capture of HLA-A*0201/peptide complexes. Then, the column was washed using 50 mM Tris (pH 8) solution at 0.5 mL/min for 5 min to remove excess cell lysate in the column. Finally, 0.2 M acetic acid solution (pH 3) was introduced at a flow rate of 0.5 mL/min for 10 min to elute the immobilized HLA-A*0201/peptide complexes from the column surface. The output was collected into 1.5 mL low-protein binding microcentrifuge tubes. Glacial acetic acid (17.5 M original) was added to the eluate to increase the acetic acid percentage to 10% to dissociate the HLA-A*0201 from the peptides. Finally, the peptides were separated from the high molecular weight proteins by centrifugation at 3220 rpm in 4 °C for 40 min on an Amicon Ultra-3 kDa device (Millipore, Billerica, MA). The peptide extract was lyophilized for ∼8 h at 40 °C. The lyophilized peptide samples were stored in −80 °C until they were analyzed by DIMS–MS/MS.

Mass Spectrometry

A Bruker HCT ion trap mass spectrometer was used for these experiments. The lyophilized samples obtained from the cell line were reconstituted for electrospray ionization (ESI) in 60 μL of 50/49.9/0.1 ACN/H2O/FA (V/V/V) and were directly infused for electrospray ionization (ESI) at a flow rate of 2 μL/min. The scan rate was 26,000 (m/z)/s for all mass spectra. All MS/MS spectra represent an average of ten to 11 spectra.

Differential Ion Mobility Spectrometry

The ESI emitter was held at ground potential, and a voltage of −4.25 kV was applied to the custom built planar DIMS electrode assembly and housing (Figure 1). The two parallel stainless steel 4 × 10 mm electrodes are separated by a 0.3 mm gap. The assembly slides onto the glass transfer capillary of the source region of the mass spectrometer. A housing is then oriented around the assembly to reroute the desolvation gas, which also serves as the carrier gas through the device. The temperature of the desolvation gas is set to 300 °C in the instrument software. The residence time for ions in the device is 0.2 ms.
Figure 1

AutoCAD drawing of DIMS assembly.

AutoCAD drawing of DIMS assembly. Ideally, a square wave should be used for DIMS, alternating between low and high electric fields of opposing polarity. However, because of the high power requirements of high-voltage, high-frequency square waves, most DIMS waveforms are bisinusoidal, approximating a square wave (Figure 2).[33] In these experiments the waveform generator output a sinusoidal wave at 2 MHz and a lower amplitude sinusoidal wave at 4 MHz, phase shifted about 90°. Each sinusoidal wave is applied to one of the electrodes, and thus, they are capacitatively summed across the gap to form the bisinusoidal DIMS waveform at frequency of 2 MHz.[34] DIMS spectra were acquired by scanning the CV using a LabVIEW program linked to the instrument control software. Ten mass spectra were averaged for each CV in the DIMS spectra. A static CV for the filter mode can also be selected with the LabVIEW program. The MS/MS spectra shown for the filter mode were averaged for 1 min at the static CV.
Figure 2

Simplified schematic of the addition of sinusoidal waves across the gap to form a bisinusoidal waveform.

Simplified schematic of the addition of sinusoidal waves across the gap to form a bisinusoidal waveform.

Results and Discussion

Spiked Peptide in a Model Peptide Pool

The model peptide pool that was generated was intended to simulate a complex mixture of very similar peptides (by molecular weight, hydrophobicity, and sequence) such as that which would be seen in a collection of antigen peptides obtained from a biological source. A DIMS scan of a solution of the pure “target” peptide, protonated GLR (amino acid sequence = FLSSANEHL), was performed to determine the optimum CV for selection of this peptide. The DIMS spectrum is shown in Figure 3. At a DV of 1200 V, the optimum CV was determined to be 6.5 V.
Figure 3

DIMS scan of protonated FLSSANEHL (also called GLR m/z 1017.5) at a DV of 1200 V.

DIMS scan of protonated FLSSANEHL (also called GLR m/z 1017.5) at a DV of 1200 V. The MS/MS spectrum obtained for pure GLR (m/z 1017.5) was compared to that obtained for the mixture of 94 peptides (Figure 4a). Without GLR spiked into the mixture, none of the product ions observed from pure GLR were present (Figure 4b). The GLR peptide was spiked into the 94-peptide pool at concentrations ranging from 1 μM to 1 pM. For each solution the MS/MS spectrum at the mass-to-charge ratio of the target protonated GLR peptide was taken with and without DIMS active. The five most intense product ions observed in the MS/MS spectrum of pure GLR, [M + H – H2O]+, [M + H – NH3]+, [M + H – H2O – NH3]+, b8, and b8-NH3 were used to evaluate whether GLR was detectable at each concentration. With 1 pM GLR spiked into the mixture, two of the five product ions were observable at a signal-to-background ratio greater than 3, where background is defined as the signal obtained for the mixture without GLR spiked in. However, the ratio of these product ions deviated too much from the reference spectrum to consider the peptide to be detected at this concentration. With 10 pM GLR spiked into the mixture, all five of the product ions were observable, with a contaminant ion present that was observed without GLR (Figure 4c). Without DIMS, none of the product ions observed at greater than 5% relative intensity match those observed from MS/MS of protonated GLR (Figure 4d). Using the detection of five product ions as indicative of the presence of the target ion (in typical biomarker detection, only two product ions are used to confirm the presence of an analyte), a comparison of the detection of GLR in the peptide mix with and without DIMS was made (Figure 5a,b). The red line represents the signal intensity at which the signal-to-background ratio is equal to 3. In this model peptide pool, the target antigen was readily detected at a concentration of 10 pM in a mixture of 94 other peptides possessing similar chemical properties. Each MS/MS scan was obtained using a volume of ∼2 μL; so the system is able to identify peptides at the low attomole level. This level of sensitivity (20 amol = 1.2 × 107 molecules) is important when considering applying this technique for use in human cancer samples. In the case of acute leukemia, most patient samples have ∼1 × 107 leukemia cells per mL of whole blood with each cell possessing ∼10,000 HLA molecules. A routine blood draw of 10 mL would therefore be expected to yield 1 × 1012 peptides, present at various frequencies, making the application of DIMS–MS/MS for leukemia antigen discovery feasible in most clinical samples.[31,32] In solid tumors, a 1 cm3 sample contains ∼1 × 108 cells; so again DIMS–MS/MS has the potential for use in readily accessible clinical specimens.[35]
Figure 4

MS/MS of protonated FLSSANEHL (GLR), where panel a was obtained for 10 μM FLSSANEHL in 50/49.9/0.1 acetonitrile/water/formic acid, panels b and c were obtained with DIMS at a DV of 1200 V and a CV of 6.5 V, where panel b was obtained for a mixture of 94 peptides without FLSSANEHL added, and panel c was obtained for the same mixture of 94 peptides, but with 10 pM FLSSANEHL in the mixture, and panel d was obtained without DIMS for the mixture of 94 peptides with 10 pM FLSSANEHL.

Figure 5

Sum of the intensities of the five most intense product ions observed from MS/MS of protonated FLSSANEHL as a function of the concentration of FLSSANEHL (GLR) in the mixture of 94 peptides, where panel a is without DIMS and panel b is with DIMS under the optimum conditions for the selection of protonated GLR. The red line indicates the signal intensity at which the signal-to-background ratio is 3, where background is defined as the signal obtained for the mixture of 94 peptides without GLR spiked in.

MS/MS of protonated FLSSANEHL (GLR), where panel a was obtained for 10 μM FLSSANEHL in 50/49.9/0.1 acetonitrile/water/formic acid, panels b and c were obtained with DIMS at a DV of 1200 V and a CV of 6.5 V, where panel b was obtained for a mixture of 94 peptides without FLSSANEHL added, and panel c was obtained for the same mixture of 94 peptides, but with 10 pM FLSSANEHL in the mixture, and panel d was obtained without DIMS for the mixture of 94 peptides with 10 pM FLSSANEHL. Sum of the intensities of the five most intense product ions observed from MS/MS of protonated FLSSANEHL as a function of the concentration of FLSSANEHL (GLR) in the mixture of 94 peptides, where panel a is without DIMS and panel b is with DIMS under the optimum conditions for the selection of protonated GLR. The red line indicates the signal intensity at which the signal-to-background ratio is 3, where background is defined as the signal obtained for the mixture of 94 peptides without GLR spiked in.

Identification of a Known Cancer Associated Antigen, CG1, Using DIMS–MS/MS

A significant problem in cancer immunology is the identification of cancer-specific peptide antigens, which can then be developed into cancer immunotherapeutics, such as vaccines.[12,36,37] To demonstrate how DIMS–MS/MS could be applied to this problem, the device was used to detect three known cancer antigens from a model leukemia cell line, U937.A2. The three peptide antigens targeted, CG1 (FLLPTGAEA), UNC-CDK4-1 (ALTPVVVTL), and UNC-ANKRD17-1 (LLIERGASL), are (1) derived from proteins that are overexpressed or aberrantly processed in human acute myeloid leukemia (AML), (2) computationally predicted to bind with high affinity to HLA-A*0201 (the most common HLA type), and (3) have been previously shown to be presented by U937.A2 cells.[23,25,38] The optimum CV to select for sodiated CG1 (FLLPTGAEA + Na+, m/z 940.4) at a DV of 1200 V was determined to be 7.8 V from a DIMS scan of the pure peptide (Figure 6a). Note that, rather than the analysis being performed with the protonated peptide, the sodiated peptide was used because there was greater sensitivity with the latter. The sodiated peptide would likely not be seen in an LC/MS experiment because the LC effectively “desalts” the sample, and there would be no sodium present in the solvent to effect ionization when the peptide eluted. The MS/MS of the pure CG1 peptide is shown in Figure 6b. MS/MS of sodiated peptides has been extensively studied, and in general, the fragmentation efficiency is as good or better than that observed for protonated peptides.[39,40] With DIMS, the peptide cell extract MS/MS spectrum of m/z 940.4 exhibited five product ions that matched the MS/MS of the pure peptide as well as a contaminant peak (Figure 6c). Without DIMS, none of the product ions observed at greater than 5% relative intensity match those observed from the MS/MS of sodiated CG1 (Figure 6d).
Figure 6

(a) DIMS scan of sodiated FLLPTGAEA (m/z 940.6) and MS/MS obtained for sodiated FLLPTGAEA, where panel b was obtained for 10 μM FLLPTGAEA in 50/49.9/0.1 acetonitrile/water/formic acid, panel c was obtained for the peptide cell extract with DIMS with a DV of 1200 V and a CV of 7.8 V, and panel d was obtained for the peptide cell extract without DIMS.

(a) DIMS scan of sodiated FLLPTGAEA (m/z 940.6) and MS/MS obtained for sodiated FLLPTGAEA, where panel b was obtained for 10 μM FLLPTGAEA in 50/49.9/0.1 acetonitrile/water/formic acid, panel c was obtained for the peptide cell extract with DIMS with a DV of 1200 V and a CV of 7.8 V, and panel d was obtained for the peptide cell extract without DIMS. This approach was used for two other previously confirmed antigen peptides UNC-CDK4-1 (ALTPVVVTL) and UNC-ANKRD17-1 (LLIERGASL).[38] After determining the optimum CVs for UNC-CDK4-1 and UNC-ANKRD17-1, the same discovery experiment was performed as described above. In the case of UNC-CDK4-1, four identifiable product ions could be detected in the MS/MS with DIMS active and with DIMS transparent (Supplementary Figure S-1 a-c). For UNC-ANKRD17-1, one identifiable product ion could be assigned on the MS/MS with DIMS active, and no product ions could be identified without DIMS (Supplementary Figure S-2 a-c). With one observed product ion, we believe we have detected this peptide but cannot confirm its detection due to the lack of additional product ions and inability to perform MS3 due to low ion intensity. In all 3 peptides tested, DIMS–MS/MS performed as well or better than MS/MS alone. Multiple product ions from the MS/MS spectra could be identified with DIMS active for both CG1 and UNC-CDK4-1.

Conclusions

Many human tumors are susceptible to T-cell mediated killing; however, to avoid off-target toxicity, T-cell immunotherapeutics such as vaccines or adoptive cellular therapy must target specific cancer-associated antigens.[12,36,41−43] Nontargeted HPLC–MS has typically been used to identify these antigens.[15−17,19,20] In this work we demonstrate that cancer antigens can be identified using a targeted MS approach, aimed at identifying specific antigens, and that DIMS significantly improved our ability to confirm the presence of antigens by improving the signal-to-noise of the MS experiments. Cancer antigen discovery using nontargeted MS methods have typically required extremely large cell populations of >1 × 109 cells per analysis, which makes analysis of most clinical samples impossible.[19] While we originally expanded >1 × 109 U937.A2 cells, the isolated peptides were divided into 10 aliquots so that the antigen discovery experiments were performed on peptides from ∼1 × 108 cells, a number of cells that can be obtained in many more clinical specimens, particularly AML samples.[35] The ability to detect antigens using these small quantities of cellular material is a result of our targeted approach to antigen discovery and the improved signal-to-noise of the DIMS device, which we observed in all experiments, when used in a targeted antigen discovery method. Over the past two decades, computational algorithms have greatly enhanced our ability to predict cancer antigens.[22−26] With this advancement, targeted MS approaches, as described in this article, can now be used to screen for and identify high value cancer antigens. This approach is particularly exciting given the dramatic expansion of cancer genome sequencing.[44] With these types of genomic data, it is possible to probe patient tumor genomes and predict candidate antigens that result from genetic lesions in the tumor itself.[27] DIMS–MS/MS could then be applied to specific patient samples to confirm or deny the presence of the predicted antigen, which could lead to more informed decisions regarding possible cancer immunotherapies. Modifications to our current device could include further optimization of the DIMS electrodes to improve focusing of the ion stream in the DIMS prior to entry into the MS and the use of nanospray ionization.
  41 in total

1.  Review of applications of high-field asymmetric waveform ion mobility spectrometry (FAIMS) and differential mobility spectrometry (DMS).

Authors:  Beata M Kolakowski; Zoltán Mester
Journal:  Analyst       Date:  2007-06-26       Impact factor: 4.616

2.  Online nanoelectrospray/high-field asymmetric waveform ion mobility spectrometry as a potential tool for discovery pharmaceutical bioanalysis.

Authors:  Panos Hatsis; Gary Valaskovic; Jing-Tao Wu
Journal:  Rapid Commun Mass Spectrom       Date:  2009-12       Impact factor: 2.419

3.  C-terminal peptide sequencing via multistage mass spectrometry.

Authors:  T Lin; G L Glish
Journal:  Anal Chem       Date:  1998-12-15       Impact factor: 6.986

4.  Selection and generation of waveforms for differential mobility spectrometry.

Authors:  Evgeny V Krylov; Stephen L Coy; John Vandermey; Bradley B Schneider; Thomas R Covey; Erkinjon G Nazarov
Journal:  Rev Sci Instrum       Date:  2010-02       Impact factor: 1.523

5.  Transferred WT1-reactive CD8+ T cells can mediate antileukemic activity and persist in post-transplant patients.

Authors:  Aude G Chapuis; Gunnar B Ragnarsson; Hieu N Nguyen; Colette N Chaney; Jeffrey S Pufnock; Thomas M Schmitt; Natalie Duerkopp; Ilana M Roberts; Galina L Pogosov; William Y Ho; Sebastian Ochsenreither; Matthias Wölfl; Merav Bar; Jerald P Radich; Cassian Yee; Philip D Greenberg
Journal:  Sci Transl Med       Date:  2013-02-27       Impact factor: 17.956

6.  NetMHCpan, a method for MHC class I binding prediction beyond humans.

Authors:  Ilka Hoof; Bjoern Peters; John Sidney; Lasse Eggers Pedersen; Alessandro Sette; Ole Lund; Søren Buus; Morten Nielsen
Journal:  Immunogenetics       Date:  2008-11-12       Impact factor: 2.846

7.  Natural splice variant of MHC class I cytoplasmic tail enhances dendritic cell-induced CD8+ T-cell responses and boosts anti-tumor immunity.

Authors:  Tania G Rodríguez-Cruz; Shujuan Liu; Jahan S Khalili; Mayra Whittington; Minying Zhang; Willem Overwijk; Gregory Lizée
Journal:  PLoS One       Date:  2011-08-10       Impact factor: 3.240

8.  Phosphorylation-dependent interaction between antigenic peptides and MHC class I: a molecular basis for the presentation of transformed self.

Authors:  Fiyaz Mohammed; Mark Cobbold; Angela L Zarling; Mahboob Salim; Gregory A Barrett-Wilt; Jeffrey Shabanowitz; Donald F Hunt; Victor H Engelhard; Benjamin E Willcox
Journal:  Nat Immunol       Date:  2008-10-05       Impact factor: 25.606

9.  The MHC class I peptide repertoire is molded by the transcriptome.

Authors:  Marie-Hélène Fortier; Etienne Caron; Marie-Pierre Hardy; Grégory Voisin; Sébastien Lemieux; Claude Perreault; Pierre Thibault
Journal:  J Exp Med       Date:  2008-02-25       Impact factor: 14.307

10.  NetMHC-3.0: accurate web accessible predictions of human, mouse and monkey MHC class I affinities for peptides of length 8-11.

Authors:  Claus Lundegaard; Kasper Lamberth; Mikkel Harndahl; Søren Buus; Ole Lund; Morten Nielsen
Journal:  Nucleic Acids Res       Date:  2008-05-07       Impact factor: 16.971

View more
  7 in total

1.  Differential mobility spectrometry: a valuable technology for analyzing challenging biological samples.

Authors:  J Larry Campbell; J C Yves Le Blanc; Richard G Kibbey
Journal:  Bioanalysis       Date:  2015       Impact factor: 2.681

2.  Improved Differential Ion Mobility Separations Using Linked Scans of Carrier Gas Composition and Compensation Field.

Authors:  Brandon G Santiago; Rachel A Harris; Samantha L Isenberg; Mark E Ridgeway; Alice L Pilo; Desmond A Kaplan; Gary L Glish
Journal:  J Am Soc Mass Spectrom       Date:  2015-07-07       Impact factor: 3.109

3.  Variables Affecting the Internal Energy of Peptide Ions During Separation by Differential Ion Mobility Spectrometry.

Authors:  Brandon G Santiago; Matthew T Campbell; Gary L Glish
Journal:  J Am Soc Mass Spectrom       Date:  2017-06-26       Impact factor: 3.109

4.  Increased Ion Transmission for Differential Ion Mobility Combined with Mass Spectrometry by Implementation of a Flared Inlet Capillary.

Authors:  Matthew T Campbell; Gary L Glish
Journal:  J Am Soc Mass Spectrom       Date:  2016-10-17       Impact factor: 3.109

5.  Computational modeling and confirmation of leukemia-associated minor histocompatibility antigens.

Authors:  Jefferson L Lansford; Udara Dharmasiri; Shengjie Chai; Sally A Hunsucker; Dante S Bortone; James E Keating; Ian M Schlup; Gary L Glish; Edward J Collins; Gheath Alatrash; Jeffrey J Molldrem; Paul M Armistead; Benjamin G Vincent
Journal:  Blood Adv       Date:  2018-08-28

6.  Alkali Metal Cationization of Tumor-associated Antigen Peptides for Improved Dissociation and Measurement by Differential Ion Mobility-Mass Spectrometry.

Authors:  James E Keating; Chris Chung; Shengjie Chai; Jans F Prins; Benjamin G Vincent; Sally A Hunsucker; Paul M Armistead; Gary L Glish
Journal:  J Proteome Res       Date:  2020-07-20       Impact factor: 5.370

7.  To What Extent is FAIMS Beneficial in the Analysis of Proteins?

Authors:  Helen J Cooper
Journal:  J Am Soc Mass Spectrom       Date:  2016-02-02       Impact factor: 3.109

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.