Differential ion mobility spectrometry (DIMS) can be used as a filter to remove undesired background ions from reaching the mass spectrometer. The ability to use DIMS as a filter for known analytes makes DIMS coupled to tandem mass spectrometry (DIMS-MS/MS) a promising technique for the detection of cancer antigens that can be predicted by computational algorithms. In experiments using DIMS-MS/MS that were performed without the use of high-performance liquid chromatography (HPLC), a predicted model antigen, GLR (FLSSANEHL), was detected at a concentration of 10 pM (20 amol) in a mixture containing 94 competing model peptide antigens, each at a concentration of 1 μM. Without DIMS filtering, the GLR peptide was undetectable in the mixture even at 100 nM. Again, without using HPLC, DIMS-MS/MS was used to detect 2 of 3 previously characterized antigens produced by the leukemia cell line U937.A2. Because of its sensitivity, a targeted DIMS-MS/MS methodology can likely be used to probe for predicted cancer antigens from cancer cell lines as well as human tumor samples.
Differential ion mobility spectrometry (DIMS) can be used as a filter to remove undesired background ions from reaching the mass spectrometer. The ability to use DIMS as a filter for known analytes makes DIMS coupled to tandem mass spectrometry (DIMS-MS/MS) a promising technique for the detection of cancer antigens that can be predicted by computational algorithms. In experiments using DIMS-MS/MS that were performed without the use of high-performance liquid chromatography (HPLC), a predicted model antigen, GLR (FLSSANEHL), was detected at a concentration of 10 pM (20 amol) in a mixture containing 94 competing model peptide antigens, each at a concentration of 1 μM. Without DIMS filtering, the GLR peptide was undetectable in the mixture even at 100 nM. Again, without using HPLC, DIMS-MS/MS was used to detect 2 of 3 previously characterized antigens produced by the leukemia cell line U937.A2. Because of its sensitivity, a targeted DIMS-MS/MS methodology can likely be used to probe for predicted cancer antigens from cancer cell lines as well as humantumor samples.
Entities:
Keywords:
DIMS; Differential ion mobility spectrometry; FAIMS; cancer antigen; mass spectrometry
Differential ion mobility
spectrometry (DIMS) provides a gas-phase
separation of ions prior to mass analysis. DIMS has the ability to
separate isobaric and isomeric species and therefore can improve the
specificity and detection limits of mass spectrometry-based analyses.[1−3] Low electric field ion mobility separations take advantage of the
difference in the collisional cross-section of ions, which is directly
proportional to the low-field ion mobility.[4,5] At
low electric fields the mobility of an ion is independent of the field.
However, at high electric fields (>104 V/cm), the mobility
of an ion becomes dependent on the electric field in a complex function
that is still not understood.[4,5] DIMS takes advantage
of the change in ion mobility in high electric fields to separate
ions based on the difference in ion mobility between low and high
electric fields.Differential ion mobility spectrometers are
composed of two parallel
electrodes. The gas flow into the mass spectrometer carries ions through
the gap between the DIMS electrodes. In “transparent”
mode, both of the electrodes are held at the same potential, and all
ions are allowed to pass through the device without an ion mobility
separation. In “active” mode, an asymmetric waveform
is applied to the electrodes. The DIMS waveform alternates between
low and high electric fields of opposing polarity, such that the ions
zigzag through the device. The V0-P of the DIMS waveform measured during the high electric field portion
is called the dispersion voltage (DV). Because ions have a different
mobility in the high field versus low field, they are displaced toward
one or the other electrodes during transit through the device. To
direct ions of a selected differential ion mobility into the mass
spectrometer, a compensation voltage (CV) must be applied to one of
the electrodes, which counter balances the displacement due to the
differential mobility. Because high-field ion mobility is not yet
well understood, the optimum CV and DV for a given analyte are not
easily predicted. This makes DIMS currently most useful for targeted
analyses, where optimal DIMS conditions for the selection of the analyte
of interest can be determined prior to analysis of the real sample.
While there is a loss in overall ion signal using DIMS, the background
signal (chemical noise) is typically reduced to a much greater extent
than the analyte signal because background ions are eliminated through
collisions with the DIMS electrodes while target ions are not. The
net result is an overall increase in signal-to-noise (S/N).One potential application for DIMS is the identification of cancer
antigens. These antigens consist of a 9–10 amino acid peptide
fragment, derived from a cancer associated protein, which is complexed
to a class I human leukocyte antigen (HLA) molecule. The peptide/HLA
complexes are assembled in the endoplasmic reticulum and subsequently
transported to and presented on the cancer cell’s surface.
A cytotoxic CD8+ T-cell expressing a T-cell receptor with
high binding affinity for the specific peptide/HLA complex can kill
the antigen-presenting cells (i.e., cancer cell) through release of
cytotoxic granule proteins.[6−8] Because of the high activity and
specificity of cancer-antigen specific CD8+ T-cells and their ability
to persist in patients indefinitely, many forms of cancer immunotherapy
have been developed to harness this approach in various cancers including
melanoma, renal cell carcinoma, prostate cancer, and leukemia.[9−12] The identification of appropriate peptide antigens is a vital part
of the development of new T-cell therapies because of the complexity
and costs involved with the production and development of these therapeutics
and the risks involved with directing T-cells to noncancer (i.e.,
normal tissue) targets.[13,14]The majority
of cancer antigenic peptides have been confirmed using
nontargeted reverse phase high-performance liquid chromatography coupled
to tandem mass spectrometry (HPLC–MS) approaches on a complex
mixture of antigenic peptides obtained from a cellular sample.[15−20] While this approach has proven extremely useful for peptide discovery,
nontargeted HPLC–MS preferentially identifies the most prevalent
peptides, which are not necessarily the most immunogenic.[21] In recent years computational algorithms have
been developed to predict which peptides in a protein are likely to
bind specific HLA molecules.[22−26] These algorithms, when combined with the significant advances in
genomics and computational biology, can be used to predict the relatively
small number of antigenic peptides in tumor extracts that are likely
immunogenic and cancer-specific because they are produced as a result
of cancer-specific processes such as somatic gene mutations, fusions,
overexpression, and alternative splicing.[27,28]By employing a computationally predictive cancer-antigen discovery
approach, targeted MS methodologies aimed at detecting cancer-specific
peptide antigens would facilitate the discovery of high-value cancer
peptide antigens. In this study, a targeted antigen-discovery strategy
employing DIMS–MS/MS was used to identify known cancer-associated
antigenic peptides from a humanleukemia cell line without using HPLC.
Experimental
Section
Model Peptide Pool
A panel of 95 “Flashpure”
peptides (New England Peptide, Gardner MA) that had been computationally
predicted to bind with high affinity to HLA-A*0201 (Supplementary Table S-1) were dissolved in 100% DMSO to a
concentration of 20 mM and subsequently diluted to 1 μM using
50/49.9/0.1 ACN/H2O/FA (V/V/V) (Sigma-Aldrich, St. Louis,
MO) (Supplementary Table S-1). To create
a peptide pool, 94 of the 20 mM peptide solutions in DMSO were combined
in equal volumes and diluted to a final concentration of 1 μM
of each peptide in 50/49.9/0.1 ACN/H2O/FA (V/V/V). The
95th peptide, GLR (amino acid sequence FLSSANEHL), was further diluted
using 50/49.9/0.1 ACN/H2O/FA (V/V/V) and spiked into the
94 peptide pool to make 1 μM, 100 nM, 10 nM, 100 pM, 10 pM,
and 1 pM concentrations of GLR peptide solutions.
Leukemia Cell
Line Derived Peptide Mixture
The U937leukemia cell line stably transfected with HLA-A*0201 (called U937.A2)
were grown under sterile conditions at 37 °C with 5% CO2 in a medium consisting of RPMI 1640 supplemented with 10% heat inactivated
(56 °C for 30 min) fetal bovine serum (FBS), 1% penicillin–streptomycin,
and 1% l-glutamine.[29] Four liters
of U937.A2 cell culture suspension was incubated until the live cell
density reached ∼7.5 × 105 cells/mL. The cell
suspension (approximately 2 to 3 × 109 U937.A2 cells)
was centrifuged at 4 °C for 30 min at 1250 rpm to obtain a cell
pellet, and then the pelleted cells were lysed in 20 mM Tris-HCl,
pH 8, 150 mM NaCl, 1% (mass/volume) 3-[(3-cholamidopropyl) dimethylammonio]-1-propanesulfonate
(CHAPS), 1 mM (phenylmethylsulfonyl fluoride), 10 μg/mL pepstatin,
5 μg/mL aprotinin, and 10 μg/mL leupeptin (Sigma-Aldrich,
St. Louis, MO). The mixture was gently rocked in a laboratory shaker
for 15 min at 4 °C to break up the cells. The cell lysate was
subjected to ultracentrifugation at 40,000 rpm for 1 h to remove cell
debris.[30] All protein loading and elution
steps were then performed using a Chemyx Fusion 100 syringe pump (Stafford,
TX). The cell pellet consisted of ∼2 × 109 cells
of U937.A2 with each cell estimated to have ∼10,000 HLA-A*0201
molecules.[31,32] For efficient isolation of nearly
all the HLA-A*0201 molecules in the lysate, 100× excess molecules
(3.3 × 10–9 moles) of the anti-HLA-A*0201 antibody,
BB7.2, were introduced into a 5 mL GE Healthcare HiTRAP recombinant
protein A column (Chicago, IL) for the immobilization. The BB7.2 antibody
molecules were reconstituted in 200 mM TrisHCl (pH 8) (Sigma-Aldrich,
St. Louis, MO). For effective antibody immobilization, first, the
column was washed with 50 mM Tris (pH 8) for 5 min at 1 mL/min flow
rate. Then, the reconstituted BB7.2 solution (1 mL) was injected into
the column at 0.2 mL/min flow rate for 5 min. Excess antibody was
washed out using 200 mM TrisHCl (pH 8) at 1 mL/min for 5 min.The HLA-A*0201/peptide complex containing supernatant was introduced
into the BB7.2 antibody immobilized column at 1 mL/min for immunoaffinity
capture of HLA-A*0201/peptide complexes. Then, the column was washed
using 50 mM Tris (pH 8) solution at 0.5 mL/min for 5 min to remove
excess cell lysate in the column. Finally, 0.2 M acetic acid solution
(pH 3) was introduced at a flow rate of 0.5 mL/min for 10 min to elute
the immobilized HLA-A*0201/peptide complexes from the column surface.
The output was collected into 1.5 mL low-protein binding microcentrifuge
tubes. Glacial acetic acid (17.5 M original) was added to the eluate
to increase the acetic acid percentage to 10% to dissociate the HLA-A*0201
from the peptides. Finally, the peptides were separated from the high
molecular weight proteins by centrifugation at 3220 rpm in 4 °C
for 40 min on an Amicon Ultra-3 kDa device (Millipore, Billerica,
MA). The peptide extract was lyophilized for ∼8 h at 40 °C.
The lyophilized peptide samples were stored in −80 °C
until they were analyzed by DIMS–MS/MS.
Mass Spectrometry
A Bruker HCT ion trap mass spectrometer
was used for these experiments. The lyophilized samples obtained from
the cell line were reconstituted for electrospray ionization (ESI)
in 60 μL of 50/49.9/0.1 ACN/H2O/FA (V/V/V) and were
directly infused for electrospray ionization (ESI) at a flow rate
of 2 μL/min. The scan rate was 26,000 (m/z)/s for all mass spectra. All MS/MS spectra represent an
average of ten to 11 spectra.
Differential Ion Mobility
Spectrometry
The ESI emitter
was held at ground potential, and a voltage of −4.25 kV was
applied to the custom built planar DIMS electrode assembly and housing
(Figure 1). The two parallel stainless steel
4 × 10 mm electrodes are separated by a 0.3 mm gap. The assembly
slides onto the glass transfer capillary of the source region of the
mass spectrometer. A housing is then oriented around the assembly
to reroute the desolvation gas, which also serves as the carrier gas
through the device. The temperature of the desolvation gas is set
to 300 °C in the instrument software. The residence time for
ions in the device is 0.2 ms.
Figure 1
AutoCAD drawing of DIMS assembly.
AutoCAD drawing of DIMS assembly.Ideally, a square wave should be used for DIMS,
alternating between
low and high electric fields of opposing polarity. However, because
of the high power requirements of high-voltage, high-frequency square
waves, most DIMS waveforms are bisinusoidal, approximating a square
wave (Figure 2).[33] In these experiments the waveform generator output a sinusoidal
wave at 2 MHz and a lower amplitude sinusoidal wave at 4 MHz, phase
shifted about 90°. Each sinusoidal wave is applied to one of
the electrodes, and thus, they are capacitatively summed across the
gap to form the bisinusoidal DIMS waveform at frequency of 2 MHz.[34] DIMS spectra were acquired by scanning the CV
using a LabVIEW program linked to the instrument control software.
Ten mass spectra were averaged for each CV in the DIMS spectra. A
static CV for the filter mode can also be selected with the LabVIEW
program. The MS/MS spectra shown for the filter mode were averaged
for 1 min at the static CV.
Figure 2
Simplified schematic of the addition of sinusoidal
waves across
the gap to form a bisinusoidal waveform.
Simplified schematic of the addition of sinusoidal
waves across
the gap to form a bisinusoidal waveform.
Results and Discussion
Spiked Peptide in a Model Peptide Pool
The model peptide
pool that was generated was intended to simulate a complex mixture
of very similar peptides (by molecular weight, hydrophobicity, and
sequence) such as that which would be seen in a collection of antigen
peptides obtained from a biological source. A DIMS scan of a solution
of the pure “target” peptide, protonated GLR (amino
acid sequence = FLSSANEHL), was performed to determine the optimum
CV for selection of this peptide. The DIMS spectrum is shown in Figure 3. At a DV of 1200 V, the optimum CV was determined
to be 6.5 V.
Figure 3
DIMS scan of protonated FLSSANEHL (also called GLR m/z 1017.5) at a DV of 1200 V.
DIMS scan of protonated FLSSANEHL (also called GLR m/z 1017.5) at a DV of 1200 V.The MS/MS spectrum obtained for pure GLR (m/z 1017.5) was compared to that obtained
for the mixture
of 94 peptides (Figure 4a). Without GLR spiked
into the mixture, none of the product ions observed from pure GLR
were present (Figure 4b). The GLR peptide was
spiked into the 94-peptide pool at concentrations ranging from 1 μM
to 1 pM. For each solution the MS/MS spectrum at the mass-to-charge
ratio of the target protonated GLR peptide was taken with and without
DIMS active. The five most intense product ions observed in the MS/MS
spectrum of pure GLR, [M + H – H2O]+,
[M + H – NH3]+, [M + H – H2O – NH3]+, b8, and
b8-NH3 were used to evaluate whether GLR was
detectable at each concentration. With 1 pM GLR spiked into the mixture,
two of the five product ions were observable at a signal-to-background
ratio greater than 3, where background is defined as the signal obtained
for the mixture without GLR spiked in. However, the ratio of these
product ions deviated too much from the reference spectrum to consider
the peptide to be detected at this concentration. With 10 pM GLR spiked
into the mixture, all five of the product ions were observable, with
a contaminant ion present that was observed without GLR (Figure 4c). Without DIMS, none of the product ions observed
at greater than 5% relative intensity match those observed from MS/MS
of protonated GLR (Figure 4d). Using the detection
of five product ions as indicative of the presence of the target ion
(in typical biomarker detection, only two product ions are used to
confirm the presence of an analyte), a comparison of the detection
of GLR in the peptide mix with and without DIMS was made (Figure 5a,b). The red line represents the signal intensity
at which the signal-to-background ratio is equal to 3. In this model
peptide pool, the target antigen was readily detected at a concentration
of 10 pM in a mixture of 94 other peptides possessing similar chemical
properties. Each MS/MS scan was obtained using a volume of ∼2
μL; so the system is able to identify peptides at the low attomole
level. This level of sensitivity (20 amol = 1.2 × 107 molecules) is important when considering applying this technique
for use in humancancer samples. In the case of acute leukemia, most
patient samples have ∼1 × 107 leukemia cells
per mL of whole blood with each cell possessing ∼10,000 HLA
molecules. A routine blood draw of 10 mL would therefore be expected
to yield 1 × 1012 peptides, present at various frequencies,
making the application of DIMS–MS/MS for leukemia antigen discovery
feasible in most clinical samples.[31,32] In solid tumors,
a 1 cm3 sample contains ∼1 × 108 cells; so again DIMS–MS/MS has the potential for use in readily
accessible clinical specimens.[35]
Figure 4
MS/MS of protonated
FLSSANEHL (GLR), where panel a was obtained
for 10 μM FLSSANEHL in 50/49.9/0.1 acetonitrile/water/formic
acid, panels b and c were obtained with DIMS at a DV of 1200 V and
a CV of 6.5 V, where panel b was obtained for a mixture of 94 peptides
without FLSSANEHL added, and panel c was obtained for the same mixture
of 94 peptides, but with 10 pM FLSSANEHL in the mixture, and panel
d was obtained without DIMS for the mixture of 94 peptides with 10
pM FLSSANEHL.
Figure 5
Sum of the intensities
of the five most intense product ions observed
from MS/MS of protonated FLSSANEHL as a function of the concentration
of FLSSANEHL (GLR) in the mixture of 94 peptides, where panel a is
without DIMS and panel b is with DIMS under the optimum conditions
for the selection of protonated GLR. The red line indicates the signal
intensity at which the signal-to-background ratio is 3, where background
is defined as the signal obtained for the mixture of 94 peptides without
GLR spiked in.
MS/MS of protonated
FLSSANEHL (GLR), where panel a was obtained
for 10 μM FLSSANEHL in 50/49.9/0.1 acetonitrile/water/formic
acid, panels b and c were obtained with DIMS at a DV of 1200 V and
a CV of 6.5 V, where panel b was obtained for a mixture of 94 peptides
without FLSSANEHL added, and panel c was obtained for the same mixture
of 94 peptides, but with 10 pM FLSSANEHL in the mixture, and panel
d was obtained without DIMS for the mixture of 94 peptides with 10
pM FLSSANEHL.Sum of the intensities
of the five most intense product ions observed
from MS/MS of protonated FLSSANEHL as a function of the concentration
of FLSSANEHL (GLR) in the mixture of 94 peptides, where panel a is
without DIMS and panel b is with DIMS under the optimum conditions
for the selection of protonated GLR. The red line indicates the signal
intensity at which the signal-to-background ratio is 3, where background
is defined as the signal obtained for the mixture of 94 peptides without
GLR spiked in.
Identification of a Known
Cancer Associated Antigen, CG1, Using
DIMS–MS/MS
A significant problem in cancer immunology
is the identification of cancer-specific peptide antigens, which can
then be developed into cancer immunotherapeutics, such as vaccines.[12,36,37] To demonstrate how DIMS–MS/MS
could be applied to this problem, the device was used to detect three
known cancer antigens from a model leukemia cell line, U937.A2. The
three peptide antigens targeted, CG1 (FLLPTGAEA), UNC-CDK4-1 (ALTPVVVTL),
and UNC-ANKRD17-1 (LLIERGASL), are (1) derived from proteins that
are overexpressed or aberrantly processed in humanacute myeloid leukemia
(AML), (2) computationally predicted to bind with high affinity to
HLA-A*0201 (the most common HLA type), and (3) have been previously
shown to be presented by U937.A2 cells.[23,25,38]The optimum CV to select for sodiated CG1 (FLLPTGAEA
+ Na+, m/z 940.4) at
a DV of 1200 V was determined to be 7.8 V from a DIMS scan of the
pure peptide (Figure 6a). Note that, rather
than the analysis being performed with the protonated peptide, the
sodiated peptide was used because there was greater sensitivity with
the latter. The sodiated peptide would likely not be seen in an LC/MS
experiment because the LC effectively “desalts” the
sample, and there would be no sodium present in the solvent to effect
ionization when the peptide eluted. The MS/MS of the pure CG1 peptide
is shown in Figure 6b. MS/MS of sodiated peptides
has been extensively studied, and in general, the fragmentation efficiency
is as good or better than that observed for protonated peptides.[39,40] With DIMS, the peptide cell extract MS/MS spectrum of m/z 940.4 exhibited five product ions that matched
the MS/MS of the pure peptide as well as a contaminant peak (Figure 6c). Without DIMS, none of the product ions observed
at greater than 5% relative intensity match those observed from the
MS/MS of sodiated CG1 (Figure 6d).
Figure 6
(a) DIMS scan
of sodiated FLLPTGAEA (m/z 940.6)
and MS/MS obtained for sodiated FLLPTGAEA, where
panel b was obtained for 10 μM FLLPTGAEA in 50/49.9/0.1 acetonitrile/water/formic
acid, panel c was obtained for the peptide cell extract with DIMS
with a DV of 1200 V and a CV of 7.8 V, and panel d was obtained for
the peptide cell extract without DIMS.
(a) DIMS scan
of sodiated FLLPTGAEA (m/z 940.6)
and MS/MS obtained for sodiated FLLPTGAEA, where
panel b was obtained for 10 μM FLLPTGAEA in 50/49.9/0.1 acetonitrile/water/formic
acid, panel c was obtained for the peptide cell extract with DIMS
with a DV of 1200 V and a CV of 7.8 V, and panel d was obtained for
the peptide cell extract without DIMS.This approach was used for two other previously confirmed
antigen
peptides UNC-CDK4-1 (ALTPVVVTL) and UNC-ANKRD17-1 (LLIERGASL).[38] After determining the optimum CVs for UNC-CDK4-1
and UNC-ANKRD17-1, the same discovery experiment was performed as
described above. In the case of UNC-CDK4-1, four identifiable product
ions could be detected in the MS/MS with DIMS active and with DIMS
transparent (Supplementary Figure S-1 a-c). For UNC-ANKRD17-1, one identifiable product ion could be assigned
on the MS/MS with DIMS active, and no product ions could be identified
without DIMS (Supplementary Figure S-2 a-c). With one observed product ion, we believe we have detected this
peptide but cannot confirm its detection due to the lack of additional
product ions and inability to perform MS3 due to low ion
intensity. In all 3 peptides tested, DIMS–MS/MS performed as
well or better than MS/MS alone. Multiple product ions from the MS/MS
spectra could be identified with DIMS active for both CG1 and UNC-CDK4-1.
Conclusions
Many humantumors are susceptible to T-cell
mediated killing; however,
to avoid off-target toxicity, T-cell immunotherapeutics such as vaccines
or adoptive cellular therapy must target specific cancer-associated
antigens.[12,36,41−43] Nontargeted HPLC–MS has typically been used to identify these
antigens.[15−17,19,20] In this work we demonstrate that cancer antigens can be identified
using a targeted MS approach, aimed at identifying specific antigens,
and that DIMS significantly improved our ability to confirm the presence
of antigens by improving the signal-to-noise of the MS experiments.Cancer antigen discovery using nontargeted MS methods have typically
required extremely large cell populations of >1 × 109 cells per analysis, which makes analysis of most clinical samples
impossible.[19] While we originally expanded
>1 × 109 U937.A2 cells, the isolated peptides were
divided into 10 aliquots so that the antigen discovery experiments
were performed on peptides from ∼1 × 108 cells,
a number of cells that can be obtained in many more clinical specimens,
particularly AML samples.[35] The ability
to detect antigens using these small quantities of cellular material
is a result of our targeted approach to antigen discovery and the
improved signal-to-noise of the DIMS device, which we observed in
all experiments, when used in a targeted antigen discovery method.Over the past two decades, computational algorithms have greatly
enhanced our ability to predict cancer antigens.[22−26] With this advancement, targeted MS approaches, as
described in this article, can now be used to screen for and identify
high value cancer antigens. This approach is particularly exciting
given the dramatic expansion of cancer genome sequencing.[44] With these types of genomic data, it is possible
to probe patienttumor genomes and predict candidate antigens that
result from genetic lesions in the tumor itself.[27] DIMS–MS/MS could then be applied to specific patient
samples to confirm or deny the presence of the predicted antigen,
which could lead to more informed decisions regarding possible cancer
immunotherapies. Modifications to our current device could include
further optimization of the DIMS electrodes to improve focusing of
the ion stream in the DIMS prior to entry into the MS and the use
of nanospray ionization.
Authors: Evgeny V Krylov; Stephen L Coy; John Vandermey; Bradley B Schneider; Thomas R Covey; Erkinjon G Nazarov Journal: Rev Sci Instrum Date: 2010-02 Impact factor: 1.523
Authors: Aude G Chapuis; Gunnar B Ragnarsson; Hieu N Nguyen; Colette N Chaney; Jeffrey S Pufnock; Thomas M Schmitt; Natalie Duerkopp; Ilana M Roberts; Galina L Pogosov; William Y Ho; Sebastian Ochsenreither; Matthias Wölfl; Merav Bar; Jerald P Radich; Cassian Yee; Philip D Greenberg Journal: Sci Transl Med Date: 2013-02-27 Impact factor: 17.956
Authors: Fiyaz Mohammed; Mark Cobbold; Angela L Zarling; Mahboob Salim; Gregory A Barrett-Wilt; Jeffrey Shabanowitz; Donald F Hunt; Victor H Engelhard; Benjamin E Willcox Journal: Nat Immunol Date: 2008-10-05 Impact factor: 25.606
Authors: Brandon G Santiago; Rachel A Harris; Samantha L Isenberg; Mark E Ridgeway; Alice L Pilo; Desmond A Kaplan; Gary L Glish Journal: J Am Soc Mass Spectrom Date: 2015-07-07 Impact factor: 3.109
Authors: Jefferson L Lansford; Udara Dharmasiri; Shengjie Chai; Sally A Hunsucker; Dante S Bortone; James E Keating; Ian M Schlup; Gary L Glish; Edward J Collins; Gheath Alatrash; Jeffrey J Molldrem; Paul M Armistead; Benjamin G Vincent Journal: Blood Adv Date: 2018-08-28
Authors: James E Keating; Chris Chung; Shengjie Chai; Jans F Prins; Benjamin G Vincent; Sally A Hunsucker; Paul M Armistead; Gary L Glish Journal: J Proteome Res Date: 2020-07-20 Impact factor: 5.370