Literature DB >> 25183359

Comparison of three commercial RT-PCR systems for the detection of respiratory viruses.

S A Butt1, V P Maceira1, M E McCallen1, K A Stellrecht2.   

Abstract

BACKGROUND: Due to the insensitivity of rapid tests for respiratory viruses, nucleic acid amplification tests are quickly becoming the standard of care. OBJECTIVES AND STUDY
DESIGN: The performance of the FilmArray Respiratory Panel (RP) and Verigene RV+ (RV+) were compared in a retrospective analysis of 89 clinical specimens previously determined to be positive for the following viruses by our test of record, Prodesse (Pro): influenza A (29, FluA), influenza B (13, FluB), respiratory syncytial virus (12, RSV), human metapneumovirus (10, hMPV), parainfluenza (14, PIV), and adenovirus (10, AdV). Samples positive for influenza A, B or RSV were tested by both methods, while the remainder were tested by RP only. True positives were defined as positive by two or more assays.
RESULTS: Limit of detection (LOD) analyses demonstrated Pro had the lowest LOD for all FluA strains tested, PIV1, PIV2 and AdV; RV+ had the lowest LOD for FluB; and RP had the lowest LOD for RSV, PIV3 and hMPV. Of the 55 samples tested by RV+, all 54 true positive samples were positive by RV+. Of the 89 samples tested by RP, 85 of the 88 true positive samples were positive by RP. From these results, the overall sensitivities for influenza A, B and RSV were 100% and 98% for RV+ and RP, respectively. The overall sensitivity of RP for all viruses was 97%.
CONCLUSIONS: In summary, these systems demonstrated excellent performance. Furthermore, each system has benefits which will ensure they will all have a niche in a clinical laboratory.
Copyright © 2014 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Influenza; Multiplex real-time PCR; Respiratory syncytial virus; Respiratory viruses

Mesh:

Year:  2014        PMID: 25183359      PMCID: PMC7172935          DOI: 10.1016/j.jcv.2014.08.010

Source DB:  PubMed          Journal:  J Clin Virol        ISSN: 1386-6532            Impact factor:   3.168


Background

Respiratory viruses are major contributors to morbidity and mortality in all age groups [1], [2], [3], [4]. Rapid identification is important for both therapeutic and infection control purposes. Traditional rapid viral diagnoses such as immunoassays produce quick results and are simple to perform; however, have sub-optimal sensitivity [5], [6], [7], [8]. Nucleic acid amplification assays (NAAT), which are relatively rapid and have greatly enhanced sensitivity, are becoming the method of choice [9]. More recently, less complex sample-to-result molecular platforms such as Xpert Flu Assay (Cepheid), RP (BioFire Diagnostics), RV+ (Nanosphere), eSensor Respiratory Viral Panel (GenMark Diagnostic) and Liat Influenza A/B (Iquum) have been introduced into the marketplace which have the potential to provide results in 1–4 h with random access capabilities [10], [11], [12], [13], [14], [15], [16], [17], [18], [19]. The main purpose of this study was to compare the performance of two new automated real-time PCR systems with a proven reliable comparator system [11], [13], [17], [20], [21].

Objectives

We compared the performance of two IVD-cleared rapid RT-PCR tests, RV+ and RP, to results obtained with Pro. Both rapid systems are highly automated and have random access capabilities.

Study design

Clinical samples

Clinical specimens were obtained from 89 patients (age range 45 d–86 yr, median 7 yr, 60% pediatric cases) who had respiratory specimens (90% nasopharyngeal, 8% lower respiratory tract) or stool (adenovirus, 2%) submitted for the detection of respiratory viruses and were previously determined to be positive by Pro. Specimens, collected between January 2008 and February 2012 and stored at −80 °C, were selected based on results from original testing. Samples were paired for differing viruses and tested as two-sample pools. Discrepant samples were retested individually, while samples discrepant for adenovirus were refereed by viral culture as previously described [22]. Specimens were defined as true positive if they were positive by two or more assays.

Viruses

The NATtrol™ RVP Verification Panel (ZeptoMetrix, Buffalo, NY) was used for limit of detection analysis by serial dilution (1:10) in normal human plasma for endpoint detection. Viral nucleic acid concentration determinations were based on quantified control viral RNA (Hologic Gen-Probe Inc., San Diego, CA) using Pro.

RV+

The Verigene® RV+ test (Nanosphere, Northbrook, IL) combines conventional PCR with a gold nanoparticle probes detection system. Initially, Test Cartridges, Extraction Trays, and Amplification Trays were loaded into Processor SP after thawing the Amplification Trays for 10–30 min. Specimen in viral transport (0.2 ml) was added to the extraction tray. Automated sample lysing and target amplification of approximately 10% of the lysate was performed on the Verigene Processor SP as a multiplex assay for the detection of FluA, 09H1, sH3, sH1, FluB, RSV-A and RSV-B. Within 5 min of completion of amplification, the RV+ test cartridge was removed from the processor, opened to release the hybridization slide, which was then inserted into the Verigene reader. Results were reported “detected” or “not detected” for each target, or a message of “No Call” (NC) was reported in the event of assay failure, such as an internal control failure. NC samples were retested to obtain a valid result.

RP

The FilmArray Respiratory Panel v1.6 (BioFire Diagnostics Inc., Salt Lake City, UT), which at the time of the clinical evaluation, was a comprehensive panel for the detection of 15 respiratory viral targets (AdV, coronaviruses [CoV] HKU1 and NL63, FluA, 09H1, sH3, sH1, FluB, hMPV, human rhinovirus/enterovirus (HRV), PIV1–4, and RSV) by real-time PCR. The closed system integrates extraction, amplification, and detection into a simple, hands free test. Briefly, 0.3 ml of specimen in viral transport was added to 0.5 ml sample buffer. The FilmArray pouch was injected with 0.5 ml of hydration solution and 0.3 ml of the diluted sample. The pouch was inserted into the FilmArray and within 70 min results were reported as detected, not detected or invalid for each target. Invalid samples were retested to obtain a valid result. In addition, a result of equivocal for FluA occurred when only one of the two FluA specific targets was amplified.

Pro

The Prodesse assays for respiratory viruses (Hologic|Gen-Probe, San Diego, CA) are a series of either multiplex or uniplex, real-time PCR assays, including ProFlu+ (detects FluA, FluB and RSV), ProFAST+ (differentiation of 09H1, sH3, and sH1), ProAdeno+, ProHMPV+ and ProParaflu+ (differentiates PIV1–3). Briefly, viral RNA was extracted from 0.2 ml of specimen in viral transport, along with a universal internal control, using NucliSens Magnetic Extraction Reagents and the easyMAG extractor (bioMerieux, Durham, NC) and eluted to a volume of 50 μl. RNA extract, 5 μl, was amplified on SmartCyclers (Cepheid, Carlsbad, CA) with the appropriate assay as requested for clinical purposes.

Statistical analysis

Probit analyses for the limit of detection with a 95% probability of detection were performed using SPSS version 8.0 (IBM, Armonk, NY). The chi-square test was performed using Microsoft Excel 2013 (Redmond, WA). P values of 0.05 were considered statistically significant. Confidence intervals (CI) were determined using Graph-Pad (La Jolla, CA).

Results

Limit of detection

Pro had the lowest LOD for all FluA viruses, while the LOD for RV+ was within 0.5 log copies/ml of Pro for the 09H1 and sH1, and more than 1 log greater for sH3 (Table 1 ). RV+ had the lowest LOD for FluB, but RP had the lowest LOD for RSV. The analytical sensitivity was mixed as to whether Pro or RP was more sensitive for PIV, hMPV and AdV.
Table 1

Limit of detection (copies/ml).

VirusProRPRV+
FluA 09H1N13.685.424.08
FluA sH1N13.726.034.01
FluA sH3N23.295.724.67
FluB5.294.934.22
RSV4.543.424.89
PIV14.014.92n.a.
PIV24.585.58n.a.
PIV32.962.43n.a.
hMPV4.093.76n.a.
AdV2.552.59n.a.

NATtrol panel members used: adenovirus 1, influenza A H1 Seasonal (A/NEWcAL/20/99; sH1), influenza A H3 Seasonal (A/Brisbane/10/07; sH3), influenza A H1N1 2009 (A/NY/02/2009; 09H1), influenza B (B/Florida/02/06), metapneumovirus 8 (Peru6-2003), parainfluenza 1–3, and respiratory syncytial virus A.

Limit of detection (copies/ml). NATtrol panel members used: adenovirus 1, influenza A H1 Seasonal (A/NEWcAL/20/99; sH1), influenza A H3 Seasonal (A/Brisbane/10/07; sH3), influenza A H1N1 2009 (A/NY/02/2009; 09H1), influenza B (B/Florida/02/06), metapneumovirus 8 (Peru6-2003), parainfluenza 1–3, and respiratory syncytial virus A.

Performance with clinical samples

The determination of clinical performance involved 89 specimens previously positive for respiratory viruses by Pro including one sample originally positive for sH3 that became negative upon archiving (Table 2 ). The 54 samples positive for FluA, FluB or RSV and the negative sample were tested by both RP and RV+ as two sample pools. Initially, 24 sample pools were positive for the expected viruses by RV+, 1 was positive with 3 viruses, and 3 resulted as No Call. The NC pooled samples (6) were retested individually; five additional samples were positive for the expected virus while the sixth was positive for two viruses. The discrepant samples, which involved a pool of a sH1 sample and a sH3 sample additionally positive for 09H1 and a sH3 sample additionally positive for FluB, were retested individually. Upon repeat each sample was positive only for the expected virus.
Table 2

Comparison of RP and RV+ on clinical specimens.

VirusnRP
RV
TPTNFPFNInvalidTPTNFPFNInvalid
09H1N115141a1b151a
sH3N21212121
sH1N1222
Flu B1313131a3
RSV1111112
hMPV1010
PIVR155
PIVR244
PIVR355
AdV1091
RSV/AdV11c1c1
No virus111



Total89861130541206

Samples/pools positive for multiple viruses.

FluA equivocal.

Positive for RSV but negative for AdV.

Comparison of RP and RV+ on clinical specimens. Samples/pools positive for multiple viruses. FluA equivocal. Positive for RSV but negative for AdV. When testing the same samples with RP, 51 of the 54 positive specimens were positive for the expected virus. One sample from a patient with dual RSV and AdV infection was only positive for RSV and one 09H1 sample was equivocal for FluA. In addition, one pool was positive for three viruses (an AdV sample and a sH3 sample additionally positive for 09H1). Again, the samples from this discrepant pool were tested individually. Upon repeat each sample was positive only for the expected virus. The additional 33 specimens positive for PIV, hMPV or AdV were tested with RP, which detected the appropriate virus in all but one sample positive for AdV, for a total AdV detection rate of 9/11 AdV positive samples. In addition, AdV was originally detected in all positive specimens by viral culture. From these findings the sensitivities for FluA, B and RSV were 100% and 98% for RV+ and RP, respectively (95% CI = 92.1–100% and 89.3–100%, respectively). This difference in sensitivity was not statistical significance (two-tailed P value = 0.3129). The overall sensitivity of RP for all viruses was 97% (95% CI = 90.1–99.3%). The initial failure rates for these two assays with clinical samples were 10.7% and 0% for RV+ and RP. However, when including the LOD analysis, the failure rates were 9.7% and 1%, respectively. All failed results resolved with repeat testing.

Workflow and cost analysis

RP and RV+ both offer enhanced workflow, significantly reducing technical time and test-turn-around times as compared with Pro. Both systems are designed for on demand testing. The most cumbersome step for both was the test preparation, including sample addition. RP has an extra step of sample dilution, but the hands-on-time (HOT) was only 5 min (Table 3 ). RV+ had the additional steps of thawing for frozen reagents, loading the cartridges and trays, and one return to the instrument post-amplification for cartridge reading, for a total HOT of 21 min. Pro is more manual, including nucleic acid extraction and amplification reagent preparation, but has higher throughput capabilities. Cost analysis demonstrates that RP is the most expensive per test, however there are multiple targets reportable and billable per test, hence the cost/reportable is considerably lower than other methods.
Table 3

Comparison of labor and reagent costs based on list pricing.

RP
RV+
Pro
TestNRaCost/testTestNRCost/testTestNRCost/test
RP20$129.00RV+6$85.00EasyMAG$6.17
ProFlu+3$53.10
ProFAST+3$35.77
ProHMPV+1$35.77
ProParaflu+3$35.77
ProAdeno+1$35.77
Total20$129.00Total6$85.00Total11$202.35
Cost/reportable$6.45Cost/reportable$14.17Cost/reportable$18.40

Number of reportable.

Hands-on-time.

Test-turnaround-time.

One tenth of per run labor when testing10 specimens.

Five tests analyses for each sample.

Minutes per reportable.

Comparison of labor and reagent costs based on list pricing. Number of reportable. Hands-on-time. Test-turnaround-time. One tenth of per run labor when testing10 specimens. Five tests analyses for each sample. Minutes per reportable.

Discussion

Since their introduction in 2011, there have been numerous studies regarding the performance of RV+ [10], [15], [23] and RP [11], [12], [14], [24]; however, only one compared the systems side by side [25] Interestingly, in this study both RP and Pro demonstrated significantly superior sensitivity for the detection of FluA as compared to RV+. Hence, it is surprising that in our study the LOD for all FluA strains tested were much lower with RV+ than RP; and both systems demonstrated excellent sensitivity with clinical specimens. Furthermore, Van Wesenbeeck et al. [25] evaluated the LOD of the different assays based on the virus titers in the clinical samples and were able to determine an LOD for RP and Pro for FluA in the range of 3.5–4.5 log RNA copies/ml, which is similar to our findings for Pro but not for RP. One potential reason for the discrepancy between studies is that all specimens in the Belgium study were collected between February and March 2012 and almost all FluA specimens subtyped as sH3. Hence, RV+ may perform suboptimally with the strain of sH3 circulating in Belgium during the winter of 2012. Our clinical samples would have been representative of the typical genetic diversity seen over a period of 4 years. In addition, these authors reported a higher than typical percentage of invalid results with RV+ (15.8%), which may indicate a system problem or improper handling of reagents which is critical for optimal performance. Other samples in which the expected virus was not detected were two containing AdV. In our study, the LOD with RP for the virus strain in the NATtrol™ RVP Verification Panel was similar to Pro. However, it has been established that RP version 1.6 had low sensitivity for most group C AdV as well as select serotypes with all other groups [11], [15], whereas the version 1.7 kits demonstrated significant improvements in the range of AdV detected [26]. Interestingly, these two samples were still negative with version 1.7 kits; however, one was now negative with Pro, and the other had a Ct value of more than 37 which is very close to the LOD. One limitation of these studies is that the clinical performance of Pro cannot be assessed as sample selection was based on Pro test results. In addition, assay specificity cannot be accurately assessed because the population was artificially skewed toward positive specimens rather than the annual respiratory virus positivity rate of 29% that we typically see at our institution. However, false positive results were seen with both RP and RV+. It is possible that the false positive results were an artifact of testing samples in pools. Contamination is another possible source for false positive results. RP is a closed, real-time PCR; hence there is no opportunity for contamination with residual amplicon. Of course, carry over contamination between samples is possible, but unlikely since the samples were all tested individually on one instrument. Although RV+ does involve post-amplification processing, the manufacturer has indicated that amplicons are not released from their cartridges. In addition to the enhanced workflow, both RP & RV+ have many advantages and disadvantages (Table 4 ). RV+ allows for a relatively simple, highly sensitive option for detecting FluA, B and RSV at a more economical price. But since this system does not detect other respiratory viruses, a second system would be required if a full diagnostic profile is needed. RP has room temperature reagent storage and a total test time of little over an hour, making this system conducive for STAT testing. RP also detects a broad range of organisms, including HRV and CoV. After our studies were performed, the manufacturer added two more strains of CoV, as well as three bacterial targets. Both systems only test one specimen at a time per instrument; hence, multiple instruments are necessary for increased throughput.
Table 4

Advantages and disadvantages of the automated systems.

AdvantageDisadvantage
RPHighly automated with little HOTaMultiple instruments cannot be linked
TAT ∼ 1 h for STAT testing capabilitiesNot as sensitive as other molecular tests
Low cost per targetHigh cost per patient
On demand testing
Room temp storage for all components
Subtypes influenza A virus strains
Detects all respiratory viruses previously detected by culture
Low percentage of invalid results
Small foot print (1.94 sq. ft. per system)



RV+Highly automated with minimal HOT.Return to instrument required with strict timing
Multiple instruments can be linkedOne-sample/instrument
Processors are stackablePost-amplification processing
Good sensitivitySplit storage conditions for kit components (refrigerated and frozen)
Subtypes influenza A virus strainsOnly detects influenza A, B and RSV
Small foot print (2.87 sq. ft. per system, 1.21 sq. ft. each additional processer)

Hands-on-time.

Advantages and disadvantages of the automated systems. Hands-on-time. Lastly, Pro offers the advantage of high throughput testing and a broad menu with the ability to select different analytes. It is difficult to compare the labor costs for this system with the two on-demand systems as the labor per specimen needed for Pro is highly dependent on the number of specimens tested. None-the-less, all three systems offer excellent sensitivity and each system offers different benefits to fit different laboratory needs. It is apparent that these highly automated assays demonstrate similar performance and greatly reduce the hands-on time.

Funding

We acknowledge the kind generosity of Nanosphere and BioFire Diagnostics Inc., for providing the reagents for this study.

Competing interests

The authors declare no competing interests.

Ethical approval

Obtained from Albany Medical Center.
  26 in total

1.  Evaluation of the Cepheid Xpert Flu Assay for rapid identification and differentiation of influenza A, influenza A 2009 H1N1, and influenza B viruses.

Authors:  S M Novak-Weekley; E M Marlowe; M Poulter; D Dwyer; D Speers; W Rawlinson; C Baleriola; C C Robinson
Journal:  J Clin Microbiol       Date:  2012-02-29       Impact factor: 5.948

2.  Low sensitivity of rapid diagnostic test for influenza.

Authors:  Timothy M Uyeki; Ramakrishna Prasad; Charles Vukotich; Samuel Stebbins; Charles R Rinaldo; Yu-Hui Ferng; Stephen S Morse; Elaine L Larson; Allison E Aiello; Brian Davis; Arnold S Monto
Journal:  Clin Infect Dis       Date:  2009-05-01       Impact factor: 9.079

3.  Comparison of two multiplex methods for detection of respiratory viruses: FilmArray RP and xTAG RVP.

Authors:  Kenneth H Rand; Howard Rampersaud; Herbert J Houck
Journal:  J Clin Microbiol       Date:  2011-04-20       Impact factor: 5.948

4.  Comparison of the Biofire FilmArray RP, Genmark eSensor RVP, Luminex xTAG RVPv1, and Luminex xTAG RVP fast multiplex assays for detection of respiratory viruses.

Authors:  Elena B Popowitch; Stacey S O'Neill; Melissa B Miller
Journal:  J Clin Microbiol       Date:  2013-03-13       Impact factor: 5.948

5.  Evaluation of three influenza A and B real-time reverse transcription-PCR assays and a new 2009 H1N1 assay for detection of influenza viruses.

Authors:  Suresh B Selvaraju; Rangaraj Selvarangan
Journal:  J Clin Microbiol       Date:  2010-09-15       Impact factor: 5.948

6.  The clinical utility of a near patient care rapid microarray-based diagnostic test for influenza and respiratory syncytial virus infections in the pediatric setting.

Authors:  Akihiro Nakao; Ken Hisata; Nobuaki Matsunaga; Makoto Fujimori; Naomi Yoshikawa; Mitsutaka Komatsu; Ken Kikuchi; Hiroshi Takahashi; Toshiaki Shimizu
Journal:  Diagn Microbiol Infect Dis       Date:  2013-11-15       Impact factor: 2.803

7.  Influenza- and respiratory syncytial virus-associated mortality and hospitalisations.

Authors:  A G S C Jansen; E A M Sanders; A W Hoes; A M van Loon; E Hak
Journal:  Eur Respir J       Date:  2007-08-22       Impact factor: 16.671

8.  Influenza and respiratory syncytial virus detection in clinical specimens without nucleic acid extraction using FOCUS direct disc assay is substantially equivalent to the traditional methods and the FOCUS nucleic acid extraction-dependent RT-PCR assay.

Authors:  Suresh B Selvaraju; David Tierney; Amy L Leber; Anami Patel; Amalia K Earley; Dipeshkumar Jaiswal; Marilyn A Menegus
Journal:  Diagn Microbiol Infect Dis       Date:  2013-11-27       Impact factor: 2.803

9.  Clinical performance of the 3M Rapid Detection Flu A+B Test compared to R-Mix culture, DFA and BinaxNOW Influenza A&B Test.

Authors:  Christine C Ginocchio; Madhavi Lotlikar; Leon Falk; Suman Arora; Margaret Kowerska; Mark Bornfreund; Ryhana Manji
Journal:  J Clin Virol       Date:  2009-05-05       Impact factor: 3.168

10.  Evaluation and implementation of FilmArray version 1.7 for improved detection of adenovirus respiratory tract infection.

Authors:  Christopher D Doern; Damon Lacey; Rong Huang; Crissie Haag
Journal:  J Clin Microbiol       Date:  2013-09-25       Impact factor: 5.948

View more
  18 in total

1.  Practical Guidance for Clinical Microbiology Laboratories: Viruses Causing Acute Respiratory Tract Infections.

Authors:  Carmen L Charlton; Esther Babady; Christine C Ginocchio; Todd F Hatchette; Robert C Jerris; Yan Li; Mike Loeffelholz; Yvette S McCarter; Melissa B Miller; Susan Novak-Weekley; Audrey N Schuetz; Yi-Wei Tang; Ray Widen; Steven J Drews
Journal:  Clin Microbiol Rev       Date:  2018-12-12       Impact factor: 26.132

2.  Depletion of Human DNA in Spiked Clinical Specimens for Improvement of Sensitivity of Pathogen Detection by Next-Generation Sequencing.

Authors:  Mohammad R Hasan; Arun Rawat; Patrick Tang; Puthen V Jithesh; Eva Thomas; Rusung Tan; Peter Tilley
Journal:  J Clin Microbiol       Date:  2016-01-13       Impact factor: 5.948

3.  Trends in Bronchiolitis ICU Admissions and Ventilation Practices: 2010-2019.

Authors:  Jonathan H Pelletier; Alicia K Au; Dana Fuhrman; Robert S B Clark; Christopher Horvat
Journal:  Pediatrics       Date:  2021-05-10       Impact factor: 9.703

4.  Molecular point-of-care testing for respiratory viruses versus routine clinical care in adults with acute respiratory illness presenting to secondary care: a pragmatic randomised controlled trial protocol (ResPOC).

Authors:  Nathan J Brendish; Ahalya K Malachira; Tristan W Clark
Journal:  BMC Infect Dis       Date:  2017-02-06       Impact factor: 3.090

5.  Multicenter Clinical Evaluation of the Luminex Aries Flu A/B & RSV Assay for Pediatric and Adult Respiratory Tract Specimens.

Authors:  Stefan Juretschko; James Mahony; Richard S Buller; Ryhana Manji; Sherry Dunbar; Kimberly Walker; Arundhati Rao
Journal:  J Clin Microbiol       Date:  2017-05-24       Impact factor: 5.948

6.  Improved diagnostic policy for respiratory tract infections essential for patient management in the emergency department.

Authors:  Randy Poelman; Johan van der Meer; Corina van der Spek; Annelies Riezebos-Brilman; Marjolein Knoester; Coretta Van Leer-Buter; Alexander W Friedrich; Hubert G Niesters
Journal:  Future Microbiol       Date:  2020-06-03       Impact factor: 3.165

7.  Automated Real-Time Collection of Pathogen-Specific Diagnostic Data: Syndromic Infectious Disease Epidemiology.

Authors:  Mark A Poritz; Lindsay Meyers; Christine C Ginocchio; Aimie N Faucett; Frederick S Nolte; Per H Gesteland; Amy Leber; Diane Janowiak; Virginia Donovan; Jennifer Dien Bard; Silvia Spitzer; Kathleen A Stellrecht; Hossein Salimnia; Rangaraj Selvarangan; Stefan Juretschko; Judy A Daly; Jeremy C Wallentine; Kristy Lindsey; Franklin Moore; Sharon L Reed; Maria Aguero-Rosenfeld; Paul D Fey; Gregory A Storch; Steve J Melnick; Christine C Robinson; Jennifer F Meredith; Camille V Cook; Robert K Nelson; Jay D Jones; Samuel V Scarpino; Benjamin M Althouse; Kirk M Ririe; Bradley A Malin
Journal:  JMIR Public Health Surveill       Date:  2018-07-06

Review 8.  Sample-to-result molecular infectious disease assays: clinical implications, limitations and potential.

Authors:  Stacy G Beal; Naziheh Assarzadegan; Kenneth H Rand
Journal:  Expert Rev Mol Diagn       Date:  2016-01-11       Impact factor: 5.225

9.  Multicenter Evaluation of the ePlex Respiratory Pathogen Panel for the Detection of Viral and Bacterial Respiratory Tract Pathogens in Nasopharyngeal Swabs.

Authors:  N Esther Babady; Matthew R England; Kristen L Jurcic Smith; Taojun He; Dona Saumya Wijetunge; Yi-Wei Tang; Robin R Chamberland; Marilyn Menegus; Ella M Swierkosz; Robert C Jerris; Wallace Greene
Journal:  J Clin Microbiol       Date:  2018-01-24       Impact factor: 5.948

10.  Incidence of respiratory viral infection in infants with respiratory symptoms evaluated for late-onset sepsis.

Authors:  J B Cerone; R P Santos; D Tristram; D M Lamson; K A Stellrecht; K St George; M J Horgan; A Rios
Journal:  J Perinatol       Date:  2017-05-18       Impact factor: 2.521

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.