Accuracy of preoperative templating in total hip arthroplasty.

Andrew Riddick1, Adam Smith2, D Phill Thomas3.   

Abstract

PURPOSE: To assess the accuracy of templating in total hip arthroplasty (THA).
METHODS: Anteroposterior pelvic radiographs of 20 men and 33 women aged 17 to 80 (mean, 60) years who underwent uncemented THA of the right (n=24) and left (n=29) sides using a single prosthesis were reviewed. A 30-mm-diameter scaling ball was placed between the patient's legs for calculating the magnification factor (MF) of the radiographs. The planned THA was templated using computer software. The positioning of the scaling ball was measured. The preoperative MF was calculated as the size of the ball on radiograph divided by the true size of the ball times 100. The true MF was calculated as the size of the cup on radiograph divided by the true size of the cup times 100. The accuracy of the preoperative MF was calculated as the preoperative MF divided by the true MF times 100.
RESULTS: The cup size and stem size were predicted correctly (within one size error) in 87% and 92% of the patients, respectively; no template size was greater than 3 size errors. The overall mean pre- and post-operative MFs differed significantly (122% vs. 127%, p<0.001), as did the mean postoperative MFs between males and females (126% vs. 128%, p=0.006). Assuming the MF was constant, and the postoperative measurement was the most accurate, the mean accuracy of preoperative MF was 96% (range, 89.6-99.9%), whereas the relative mean height of the scaling ball was 30 (range, 1-90) mm away from the correct height. The accuracy of the preoperative MF did not correlate with body mass index, the scaling ball position, and the accuracy of cup and stem templates.
CONCLUSION: If the MF of 127% is constant across different patients, using a MF of 127% is more accurate than using the scaling ball technique.

Entities:  

Keywords:  arthroplasty, replacement, hip; radiographiczzm321990magnification

Mesh:

Year:  2014        PMID: 25163949     DOI: 10.1177/230949901402200211

Source DB:  PubMed          Journal:  J Orthop Surg (Hong Kong)        ISSN: 1022-5536            Impact factor:   1.118


  4 in total

1.  Digital templating without a calibration marker is accurate at predicting implant size for hip hemiarthroplasty.

Authors:  Mina Derias; Muhammad A Khan; Jamie Buchanan
Journal:  Eur J Trauma Emerg Surg       Date:  2018-09-12       Impact factor: 3.693

2.  Digital templating in hip hemiarthroplasty: Is it possible to accurately predict femoral head size from magnification alone?

Authors:  Daniel Thurston; Saad El-Ashry; Sreenadh Gella; Kanthan Theivendran
Journal:  J Clin Orthop Trauma       Date:  2022-07-16

3.  A Novel Method of Determining Acetabular Component Size to Guide Explant in Revision Hip Arthroplasty.

Authors:  Fiachra O Deasmhunaigh; Jeremy Abetz; Shaun English; David Mitchell; Biswadev Mitra
Journal:  Arthroplast Today       Date:  2020-04-24

Review 4.  The accuracy and reliability of preoperative digital 2D templating in prosthesis size prediction in uncemented versus cemented total hip arthroplasty: a systematic review and meta-analysis.

Authors:  Joshua B V Smith; Habeeb Bishi; Chao Wang; Vipin Asopa; Richard E Field; David H Sochart
Journal:  EFORT Open Rev       Date:  2021-11-19
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.