David J Browning1, Chong Lee2.
Abstract
PURPOSE: To determine measurement variability of N1P1 amplitudes and the R1/R2 ratio in normal subjects and hydroxychloroquine users without retinopathy.Entities:
Keywords: hydroxychloroquine; multifocal electroretinography; test-retest variability
Year: 2014 PMID: 25143706 PMCID: PMC4132257 DOI: 10.2147/OPTH.S66528
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Figure 110-2 visual fields over six years.
Abbreviation: FP, false positives.
Figure 2mfERGS for three consecutive years.
Notes: (A) Study of 15 August, 2011; (B) study of 20 August, 2012; (C), study of 19th August, 2013.
Demographic characteristics of subjects
| Characteristic | Controls | HC users |
|---|---|---|
| Age (years) | 38 (28–55) | 62 (54–72) |
| Women:men | 19:2 | 41:3 |
| Height (cm) | NA | 165 (160–168), n=41; not checked in 3 |
| Weight (kg) | NA | 77 (64–88), n=39; not checked in 5 |
| Daily dose (mg/day) | NA | 400 (n=22); 200 (n=14); 300 (n=6); not checked in 2 |
| Cumulative dose (grams) | NA | 706 (406–1,066); n=32; not checked in 12 |
Note: Data are presented as median (interquartile range) unless otherwise indicated.
Abbreviations: HC, hydroxychloroquine; NA, not applicable.
Values of multifocal electroretinogram measurements in normal subjects and hydroxychloroquine users without retinopathy
| Variable | Units | Normal subjects | HC users |
|---|---|---|---|
| R1 | nV/deg2 | 31.4±7.9 | 26.9±5.7 |
| R2 | nV/deg2 | 17.0±4.5 | 14.8±3.3 |
| R3 | nV/deg2 | 10.2±2.3 | 9.0±2.5 |
| R4 | nV/deg2 | 7.5±1.8 | 6.7±2.0 |
| R5 | nV/deg2 | 6.4±1.5 | 5.9±1.8 |
| R1/R2 | No unit | 1.84±0.15 | 1.83±0.23 |
| R5/R4 | No unit | 0.87±0.07 | 0.89±0.08 |
| R5/R3 | No unit | 0.64±0.08 | 0.66±0.09 |
Notes: R1 to R5 refer to ring averages of N1P1 amplitudes from the multifocal electroretinogram. R1/R2, R5/R4, and R5/R3 refer to ratios of these ring averages. Data are presented as means ± standard deviations.
Abbreviations: deg, degree; HC, hydroxychloroquine; nV, nanovolts.
Test-retest variability of multifocal electroretinogram measurements in normal subjects (n=21)
| Measurement | Units | Trial #1 | Trial #2 | COR, actual values | COR, normalized values, % |
|---|---|---|---|---|---|
| R1 | nV/deg2 | 32.4±8.5 | 30.4±9.3 | 16.1 | 52 |
| R2 | nV/deg2 | 17.6±4.3 | 16.4±4.5 | 7.2 | 43 |
| R3 | nV/deg2 | 10.5±2.6 | 9.8±2.5 | 4.3 | 43 |
| R4 | nV/deg2 | 7.8±1.9 | 7.2±2.0 | 3.2 | 43 |
| R5 | nV/deg2 | 6.7±1.6 | 6.2±1.8 | 2.9 | 46 |
| R1/R2 | No unit | 1.83±0.19 | 1.85±0.21 | 0.54 | 29 |
| R5/R4 | No unit | 0.87±0.07 | 0.86±0.06 | 0.11 | 13 |
| R5/R3 | No unit | 0.65±0.08 | 0.63±0.09 | 0.14 | 22 |
Notes: R1 to R5 refer to ring averages of N1P1 amplitudes from the multifocal electroretinogram. R1/R2, R5/R4, and R5/R3 refer to ratios of these ring averages.
Abbreviations: COR, coefficient of repeatability; deg, degree; nV, nanovolts; SD, standard deviation.
Figure 3Bland–Altman analysis of repeatability for normal subjects and hydroxychloroquine users.
Abbreviations: SD, standard deviation; HC, hydroxychloroquine.
Short-term variation of multifocal electroretinogram measurements in patients taking hydroxychloroquine without retinopathy (n=44)
| Measurement | Units | Trial #1 | Trial #2 | COR, actual values | COR, normalized values, % |
|---|---|---|---|---|---|
| R1 | nV/deg2 | 26.5±7.0 | 27.3±6.9 | 16.0 | 59 |
| R2 | nV/deg2 | 14.4±3.9 | 15.3±3.9 | 7.6 | 53 |
| R3 | nV/deg2 | 8.6±2.6 | 9.3±2.8 | 4.2 | 47 |
| R4 | nV/deg2 | 6.5±2.1 | 6.9±2.3 | 3.2 | 49 |
| R5 | nV/deg2 | 5.9±1.9 | 6.0±2.0 | 3.0 | 55 |
| R1/R2 | No unit | 1.88±0.36 | 1.80±0.25 | 0.89 | 45 |
| R5/R4 | No unit | 0.91±0.09 | 0.86±0.07 | 0.19 | 21 |
| R5/R3 | No unit | 0.69±0.12 | 0.64±0.09 | 0.24 | 36 |
Notes: R1 to R5 refer to ring averages from the multifocal electroretinogram. R1/R2, R5/R4, and R5/R3 refer to ratios of these ring averages.
Abbreviations: COR, coefficient of repeatability; deg, degree; nV, nanovolts; SD, standard deviation.