| Literature DB >> 25139037 |
T Schlebusch1, S Nienke, S Leonhardt, M Walter.
Abstract
Non-invasive estimation of bladder volume is required to progress from scheduled voiding to a demand-driven emptying scheme for patients with impaired bladder volume sensation. Electrical impedance tomography (EIT) is a promising candidate for the non-invasive monitoring of bladder volume. This article focuses on four estimation algorithms used to map recorded EIT data to a volume estimate. Two different approaches are presented: the tomographic algorithms (one based on global impedance, the other on equivalent circular diameter) rely on the reconstruction of a tomographic image and then extract a volume estimate, whereas the parametric algorithms (one based on neural networks, the other on the singular value difference method) directly map the raw data to a volume estimate. The four algorithms presented here are evaluated for volume estimation error, noise tolerance and suppression of varying urine conductivity based on finite element simulation data.Entities:
Mesh:
Year: 2014 PMID: 25139037 DOI: 10.1088/0967-3334/35/9/1813
Source DB: PubMed Journal: Physiol Meas ISSN: 0967-3334 Impact factor: 2.833