Literature DB >> 25128851

Hysteroscopic morcellation: review of the manufacturer and user facility device experience (MAUDE) database.

Karina Haber1, Eleanor Hawkins2, Mark Levie2, Scott Chudnoff2.   

Abstract

STUDY
OBJECTIVE: To investigate the number and type of adverse events associated with hysteroscopic morcellation of intrauterine disease.
DESIGN: Systematic review of Manufacturer and User Device Experience (MAUDE) database from 2005 to June 2014 (Canadian Task Force classification III). SETTING N/A PATIENTS: Women undergoing hysteroscopic surgery for removal of intrauterine polyps or myomas with use of a reciprocating morcellator.
INTERVENTIONS: The MAUDE database was searched for the key words "Hysteroscope," "Hysteroscopic reciprocating morcellator," "Interlace," "MyoSure," "Smith & Nephew," and "TRUCLEAR," to identify reported incidences of device malfunction, injury, or death. A total of 119 adverse events were analyzed. Reports were reviewed individually and categorized by date of occurrence, type of morcellation device, type of complication, and a brief description. Each company was contacted to provide an estimate of the number of procedures performed or units sold to date.
MEASUREMENTS AND MAIN RESULTS: From 2005 to June 2014, 119 adverse events were reported to the MAUDE database. On the basis of severity, adverse events were categorized as major or minor complications. Major events included intubation/admission to an intensive care unit (n = 14), bowel damage (n = 12), hysterectomy (n = 6), and death (n = 2). Minor events included uterine perforation requiring no other treatment (n = 29), device failure (n = 25), uncomplicated fluid overload (n = 19), postoperative bleeding controlled using noninvasive measures (n = 6), and pelvic infection (n = 4). These events were then categorized according to manufacturer. The number of adverse events reported to the MAUDE database was divided by the total units sold as a surrogate for the estimated number of procedures performed. Understanding the limitation of the numbers used as a numerator and denominator, we concluded that adverse events complicated hysteroscopic morcellation in <0.1% cases.
CONCLUSIONS: The suction-based, mechanical energy, rotating tubular cutting system was developed to overcome adverse events that occur during traditional resectoscopy. On the basis of acknowledged limited information from the MAUDE database, it seems that life-threatening complications such as fluid overload, uterine perforation, and bleeding do occur with hysteroscopic morcellation but less frequently than with traditional electrocautery.
Copyright © 2015 AAGL. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Hysteroscope; Interlace; MyoSure; Smith & Nephew; TRUCLEAR; hysteroscopic reciprocating morcellator

Mesh:

Year:  2014        PMID: 25128851     DOI: 10.1016/j.jmig.2014.08.008

Source DB:  PubMed          Journal:  J Minim Invasive Gynecol        ISSN: 1553-4650            Impact factor:   4.137


  10 in total

1.  Laparoscopic management of uncommon benign uterine tumors: a systematic review.

Authors:  Zaki Sleiman; Sara Hussein; Amani Mohsen; Aline Khazzaka; Alessandro Tropea; Antonio Biondi
Journal:  Updates Surg       Date:  2019-04-02

2.  Device Malfunctions and Complications Associated with Benign Prostatic Hyperplasia Surgery: Review of the Manufacturer and User Facility Device Experience Database.

Authors:  Neel H Patel; Nikil Uppaluri; Michael Iorga; Ariel Schulman; Jonathan B Bloom; John Phillips; Sean Fullerton; Sensuke Konno; Muhammad Choudhury; Majid Eshghi
Journal:  J Endourol       Date:  2019-05-24       Impact factor: 2.942

3.  Hysteroscopic management versus ultrasound-guided evacuation for women with first-trimester pregnancy loss, a randomised controlled trial.

Authors:  Hadeer Meshaal; Emad Salah; Eman Fawzy; Mazen Abdel-Rasheed; Ahmed Maged; Hany Saad
Journal:  BMC Womens Health       Date:  2022-05-25       Impact factor: 2.742

4.  Hysteroscopic myomectomy without anesthesia.

Authors:  Nuria-Laia Rodríguez-Mias; Montserrat Cubo-Abert; Laura Gomila-Villalonga; Juanjo Gómez-Cabeza; Jose Luis Poza-Barrasús; Antonio Gil-Moreno
Journal:  Obstet Gynecol Sci       Date:  2019-04-04

5.  Crosslinked Hyaluronic Acid Gels for the Prevention of Intrauterine Adhesions after a Hysteroscopic Myomectomy in Women with Submucosal Myomas: A Prospective, Randomized, Controlled Trial.

Authors:  Chen-Yu Huang; Wen-Hsun Chang; Min Cheng; Hsin-Yi Huang; Huann-Cheng Horng; Yi-Jen Chen; Wen-Ling Lee; Peng-Hui Wang
Journal:  Life (Basel)       Date:  2020-05-15

6.  Mechanical hysteroscopic tissue removal or hysteroscopic morcellator: understanding the past to predict the future. A narrative review.

Authors:  M Franchini; O Ceci; P Casadio; J Carugno; G Giarrè; G Gubbini; U Catena; M Chiara de Angelis; A Di Spiezio Sardo
Journal:  Facts Views Vis Obgyn       Date:  2021-06-10

7.  Comparison of Hysteroscopic Morcellation Versus Resectoscopy in Treatment of Patients with Endometrial Lesions: A Meta-Analysis.

Authors:  Fangying Ren; Guannan Huang; Xue Wang; Xuehui Li; Jianning Cai
Journal:  Med Sci Monit       Date:  2022-07-18

Review 8.  Focus on the Primary Prevention of Intrauterine Adhesions: Current Concept and Vision.

Authors:  Wen-Ling Lee; Chia-Hao Liu; Min Cheng; Wen-Hsun Chang; Wei-Min Liu; Peng-Hui Wang
Journal:  Int J Mol Sci       Date:  2021-05-13       Impact factor: 5.923

9.  Improved adequacy of endometrial tissue sampled from postmenopausal women using the MyoSure Lite hysteroscopic tissue removal system versus conventional curettage.

Authors:  Peter Rosenblatt; Sara Barcia; Anthony DiSciullo; Hussein Warda
Journal:  Int J Womens Health       Date:  2017-10-27

10.  Endometrial Polyp Removed by a Manual Hysteroscopic Tissue Removal Device.

Authors:  Menelik Man Hin Lee
Journal:  Gynecol Minim Invasive Ther       Date:  2020-01-23
  10 in total