Anatomical fit of seven different palmar distal radius plates.

Johannes Oppermann1, Max Wacker, Gregor Stein, Hans-Philipp Springorum, Wolfram Friedrich Neiss, Klaus J Burkhart, Peer Eysel, Jens Dargel.   

Abstract

INTRODUCTION: The purpose of this study was to compare the anatomical fit of different, precontoured palmar distal radius plates.
METHODS: The anatomical fit of seven different types of palmar distal radius plates [Königsee variable fixed-angle radius plate 7/3-hole, Königsee variable fixed-angle radius plate 5/3-hole (Allendorf, Germany), Medartis 2.5 Adaptive TriLock, Medartis 2.5 TriLock, Medartis 2.5 TriLock extraarticular, (Basel, Switzerland), Synthes VA-LCP distal two-column-radius, Synthes LCP extraarticular (Bettlach, Switzerland)] were investigated in 25 embalmed human cadaveric radii. An imprint of the space between the well-positioned plate and the distal radius was attained using a silicone mass and the maximum height of the silicone imprint was digitally measured. The mean maximum imprint height was compared between the seven plates using an analysis of variance with repeated measures and Bonferroni correction for multiple comparisons.
RESULTS: The mean maximum distance between the plates and the radial cortex was <2 mm for all plates. The greatest difference was found with the Medartis Adaptive (1.99 ± 0.45 mm) and the least difference with the Synthes two-column (1.56 ± 0.76 mm), this difference being statistically significant (p = 0.005).
CONCLUSION: Although there was no complete congruency between the plates and the radial cortex, all distal palmar radius plates investigated in this study presented a reasonable anatomical shape. The Synthes VA-LCP distal two-column-radius plate palmar showed the best anatomical fit. A low profile and optimized anatomical precontouring minimizes irritation of the surrounding soft tissues and should be considered with plate design and implant choice.

Entities:  

Mesh:

Year:  2014        PMID: 25108754     DOI: 10.1007/s00402-014-2072-y

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0936-8051            Impact factor:   3.067


  6 in total

1.  [Analysis of the accuracy of fit of five anatomically precontoured clavicle plate systems].

Authors:  Johannes Oppermann; Linda Ettmann; Carolin Meyer; Christian Ries; Gregor Stein; Christian Karl Spies; Lars Peter Müller; Christoph Faymonville
Journal:  Unfallchirurg       Date:  2018-06       Impact factor: 1.000

2.  The Watershed Line of the Distal Radius: Cadaveric and Imaging Study of Anatomical Landmarks.

Authors:  Minke Bergsma; Job N Doornberg; Annelise Borghorst; W A Kernkamp; R L Jaarsma; Gregory I Bain
Journal:  J Wrist Surg       Date:  2019-12-20

3.  Interpretations of the Term "Watershed Line" Used as Reference for Volar Plating.

Authors:  Minke Bergsma; Job N Doornberg; Laurent Hendrickx; Batur Hayat; Gino M M J Kerkhoffs; Bhavin Jhadav; Ruurd L Jaarsma; Gregory I Bain
Journal:  J Wrist Surg       Date:  2019-08-13

4.  Morphological Study of the Anterior Surface of the Distal Radius.

Authors:  Ho-Jung Cho; Soyeon Kim; Dai-Soon Kwak
Journal:  Biomed Res Int       Date:  2017-06-22       Impact factor: 3.411

5.  Analysis of the three-dimensional anatomical variance of the distal radius using 3D shape models.

Authors:  Sebastian F Baumbach; Jakob Binder; Alexander Synek; Fabian G Mück; Yan Chevalier; Ekkehard Euler; Georg Langs; Lukas Fischer
Journal:  BMC Med Imaging       Date:  2017-03-09       Impact factor: 1.930

6.  Morphometric Variations in the Volar Aspect of the Distal Radius.

Authors:  Bong Cheol Kwon; Joon Kyu Lee; Suk Yoon Lee; Jae Yeun Hwang; Jang-Hyeon Seo
Journal:  Clin Orthop Surg       Date:  2018-11-21
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.