Catheter displacement prior to the delivery of high-dose-rate brachytherapy in the treatment of prostate cancer patients.

Shogo Kawakami1, Hiromichi Ishiyama1, Tsuyoshi Terazaki2, Itaru Soda1, Takefumi Satoh3, Masashi Kitano1, Shinji Kurosaka3, Akane Sekiguchi1, Shouko Komori1, Masatsugu Iwamura3, Kazushige Hayakawa1.   

Abstract

PURPOSE: The purpose of this work was to report measured catheter displacement prior to the delivery of high-dose-rate brachytherapy (HDR) in the treatment of prostate cancer.
MATERIAL AND METHODS: Data from 30 prostate cancer patients treated with HDR brachytherapy were analyzed retrospectively. Eighteen transperineal hollow catheters were inserted under transrectal ultrasound guidance. Gold marker seeds were also placed transperineally into the base and apex of the prostate gland. Five treatment fractions of 7.5 Gy each were administered over 3 days. The patient underwent CT scanning prior to each treatment fraction. Catheter displacement was measured from the pre-treatment CT dataset reconstructed at 1.25 mm slice thickness.
RESULTS: Most of catheters were displaced in the caudal direction. Variations of 18 catheters for each patient were small (standard deviations < 1 mm for all but one patient). Mean displacements relative to the apex marker were 6 ± 4 mm, 12 ± 6 mm, 12 ± 6 mm, 12 ± 6 mm, and 12 ± 6 mm from plan to 1(st), 2(nd), 3(rd), 4(th), and 5(th) fractions, respectively.
CONCLUSIONS: Our results indicate that catheter positions must be confirmed and if required, adjusted, prior to every treatment fraction for the precise treatment delivery of HDR brachytherapy, and to potentially reduce over-dosage to the bulbo-membranous urethra.

Entities:  

Keywords:  brachytherapy; catheter displacement; high-dose-rate; prostate cancer

Year:  2014        PMID: 25097556      PMCID: PMC4105648          DOI: 10.5114/jcb.2014.43619

Source DB:  PubMed          Journal:  J Contemp Brachytherapy        ISSN: 2081-2841


Purpose

Among the modern radiotherapeutic techniques, brachytherapy is an effective approach for delivering high-dose radiation to a target. As a result, we have been treating prostate cancer patients with high-dose-rate (HDR) brachytherapy combined with hypofractionated external beam radiotherapy since 1999 [1, 2]. The dosimetric accuracy of HDR brachytherapy relies on catheter positions being accurately reproduced to match the original positions in the planning CT dataset for all fractions of treatment. However, caudal catheter displacement is often observed prior to the first fraction [3, 4] and between fractions [5-10]. This may lead to not only a significant risk of under-dosage at the target, but also over-dosage to the organs at risk. Several studies have demonstrated urethral stricture as the predominant late urinary toxicity of prostate HDR brachytherapy [2, 11]. With significant caudal displacement, a high dose is likely to be erroneously delivered to the bulbo-membranous urethra. Sullivan et al. reported that 92% of all strictures occurred in the bulbo-membranous urethra, which lies distal to the prostatic apex [11]. Bulbo-membranous urethral stricture formation may, therefore, represent a clinical manifestation of caudal catheter displacement during HDR brachytherapy. The purpose of this work was to determine catheter displacement prior to the delivery of HDR brachytherapy based on measurements from pre-treatment CT scan compared to the original planning CT scan with an implanted apex gold marker as representative of the bulbo-membranous urethra.

Material and methods

Patients

The institutional review board approved this retrospective study. Data from 30 prostate cancer patients treated with HDR brachytherapy were retrospectively analyzed. Patient characteristics are provided in Table 1.
Table 1

Patient characteristics

n 30
Age70 (52-81)
T stageT1c6
T2b3
T2c4
T3a9
T3b8
Initial PSA (ng/mL)23.1 (4.4-486)
Gleason score78
813
96
103
Volume (ml)10-2016
20-3012
30 +2

Values are number or median (range)

Patient characteristics Values are number or median (range)

Catheter insertion and planning procedure

Patients in the operating room were placed in a lithotomy position under epidural anesthesia. Multiple 24 cm long, closed-end, 6-F hollow plastic catheters were inserted transperineally using a Syed-Neblett plastic template (Alpha-Omega Services, Bellflower, CA, USA) under transrectal ultrasound guidance. Routinely, 18 catheters were implanted independently of prostate size. Twelve catheters were inserted in the peripheral portion, and six catheters were inserted in the central portion of the prostate. The needle tips were left within the urinary bladder 1.5 cm above the sonographically or cystoscopically defined base of the prostate. A CT scan was obtained for CT-based planning. The volumetric scans with reconstructed slice thickness of 1.25 mm were obtained using multi-detector row CT (Optima CT580, GE-Healthcare, WI, USA). The planning target volume (PTV) was defined as the prostate gland with or without proximal seminal vesicle, with a manually drawn margin between 3-5 mm in all directions. Reference points for the PTV were automatically distributed on the surface of the PTV. The dose limitation (maximum dose) was set as 8 Gy per fraction for the urethra, and 4 Gy per fraction for 5 mm behind the edge of the anterior rectal wall. A dose of 7.5 Gy per fraction to the PTV was prescribed, unless the dose limitation was violated, using inverse planning and geometric optimization. Five fractions of HDR treatment were administered. After CT-based planning performed on the Oncentra treatment planning system (Nucletron, Elekta AB, Stockholm, Sweden), the first treatment session of HDR brachytherapy was conducted using the Nucletron microSelectron HDR 192Ir remote afterloading system. The first treatment session was conducted on the day of implantation, with the subsequent four treatment sessions administered twice daily on subsequent days. Basically, the time differences between plan CT and 1st, 2nd, 3rd, 4th, and 5th fractions were 6 h, 24 h, 30 h, 48 h, and 54 h, respectively. Treatment duration of HDR was thus 3 days. Six days after completion of HDR brachytherapy, patients received EBRT using a dynamic-arc conformal technique, administered with high-energy photons comprising 10-MV X-rays (MHCL-15TP, Mitsubishi Electric, Tokyo, Japan) to a total dose of 30 Gy. Total dose was administered in 5 weekly fraction doses of 3 Gy. The radiation field was limited to the prostate gland with or without proximal seminal vesicles with a 7 mm leaf margin using multileaf collimators.

Measurement of each catheter displacement between original planning scan and 1st pre-treatment scan

During catheter insertion, gold markers (VISICOIL; Iba-Dosimetry, Schwarzenbruck, Germany) were also implanted in the apex and base of the prostate. The apex-marker was used as a representative reference point for prostate and bulbo-membranous urethral position (Fig. 1A). Obturator with 3 mm marker and 10 mm spacer was inserted into every catheter (Fig. 2). A CT scan was acquired at 1.25 mm slice thickness prior to 1st fraction in order to measure the catheters displacement relative to apex marker. The actual displacement was calculated by multiplying the thickness of the CT slice with the difference in number of CT slices between the slice of the apex marker and the marker of obturator of each catheter (Fig. 1D).
Fig. 1

Measurement of catheter displacement on CT images. Gold markers were implanted at the apex (A) and base (B) of the prostate gland. Catheter displacement was calculated by multiplying the thickness of the CT slice with the difference in number of CT slices between the slice of the apex gold marker (C) and the reference slice detected by the pattern of the obturator markers (D)

Fig. 2

Scheme of obturator

Measurement of catheter displacement on CT images. Gold markers were implanted at the apex (A) and base (B) of the prostate gland. Catheter displacement was calculated by multiplying the thickness of the CT slice with the difference in number of CT slices between the slice of the apex gold marker (C) and the reference slice detected by the pattern of the obturator markers (D) Scheme of obturator The slice of an apex marker and an obturator maker were defined as the most cranial slices that showed the top of marker without artifact.

Adjustment protocol

In this study, measurements of each catheter position were done for only the first pre-treatment scan. Other catheter displacements were calculated by “slice-specific pattern” as a representative for all 18 catheter positions. Figure 3 shows our catheter adjustment protocol used in clinical practice. The obturators in 18 catheters make slice-specific patterns on CT-images (high and low density in the catheter hollows; Fig. 1D), because no catheter was inserted in completely equal depth, and that make it possible to recognize a representative slice for 18 catheters positions. Therefore, this slice-specific pattern on CT-image was used as a representative for all catheter position instead of measuring each 18 needles, because measuring all catheter positions in all treatments was not practical.
Fig. 3

Flow chart of the catheter adjustment protocol

Flow chart of the catheter adjustment protocol Where there was enough space between the first dwell position and distal end of the catheter hollow; dwell position was adjusted by changing indexer length at the treatment console. Where there was no space and displacement > 2 mm, the catheter was manually advanced by Radiation Oncologist. After manual advancement, a second CT was acquired to confirm the catheter position. The positions of gold markers implanted at the apex and base of the prostate were also checked by comparing with soft tissue anatomy on the CT images.

Total catheter displacement in adjustment protocol

The length of total catheter displacement was evaluated as follows: where Dactual is actual displacement of the catheter calculated from difference between original planning CT and the first pre-treatment CT, Dmanual is actual length of manual catheter advancement calculated from difference between CTs scanned before and after manual advancement, and Dtotal is the sum of the two.

Results

Catheter displacement

In this manuscript, we used “mean ± 1 standard deviation” for all our results. Table 2 shows actual displacements between original planning scan and 1st pre-treatment CT scan of 18 needles for each patient. Standard deviations of 18 catheters for each patient (variation within one single patient) were < 1 mm except one. These small variations among 18 catheters permit our “slice-specific pattern” recognition. Regarding total catheter displacement in the adjustment protocol, mean Dtotal values were 6 ± 4 mm, 12 ± 6 mm, 12 ± 6 mm, 12 ± 6 mm, and 12 ± 6 mm from plan to 1st, 2nd, 3rd, 4th, and 5th fractions, respectively. After the 2nd fraction, catheter displacements were reduced (Fig. 4).
Table 2

Displacements (mm) between original planning CT and 1st pre-treatment CT of 18 needles for each patient

PatientsCatheter places on templateMeanSD
123456789101112131415161718
A86886661568868666861
B33331313133433133321
C0013113-11134-10013011
D88888888988889888880
E0000000-1000100000101
F66656555555656665561
G98866665666886666871
H151516151514141313131314151415131414141
I1111109991099910111010991010101
J45445444444544545541
K891098101010109109999991091
L65668656565566666661
M56555444565555566651
NI654133NANA465553335642
O141415151415141414141414141414141414140
P55555555665555665551
Q44445544433454444441
R89886666558886566671
S44444445554434455441
T31334155443434443331
U68668666555666666661
V1010910111311111099109101010910101
W00000000001000000000
X141415141414131313141414141313131414131
Y101011111111101011111110111011111111111
Z33334331433433343331
AA131314141514141413131413131414131411131
BB9999999109910910991010891
CC1099888688991088689981
DD11111110111010999101110101091010101

SD – standard deviation within each single patient

Fig. 4

Catheter displacement during 5 fractions of highdose- rate brachytherapy. Catheter displacements were seen especially in the 1st and 2nd fractions. Mean displacement was 6 mm between the plan and 1st, and 1st and 2nd fractions

Catheter displacement during 5 fractions of highdose- rate brachytherapy. Catheter displacements were seen especially in the 1st and 2nd fractions. Mean displacement was 6 mm between the plan and 1st, and 1st and 2nd fractions Displacements (mm) between original planning CT and 1st pre-treatment CT of 18 needles for each patient SD – standard deviation within each single patient

Manual advancement of catheter

Manual catheter adjustments were needed in 31 of 150 treatment fractions. Mean length of manual advancement was 10 ± 3 mm. After manual advancement, second pre-treatment CT was acquired (Fig. 3). On the second pre-treatment CT, however, actual advancement was 4 ± 3 mm.

Gold marker migration

No displacement of apex fiducial markers compared to soft tissue anatomy was seen on CT images. Meanwhile, two patients displayed base-marker migrations of 20 mm and 25 mm just after implantation, and we failed to implant a base marker in 1 patient. In the remaining 27 patients, distances between apex and base markers gradually increased over time. Mean increases of distance between the two gold markers were 0 ± 3 mm, 2 ± 4 mm, 3 ± 4 mm, 4 ± 2 mm, and 3 ± 4 mm from plan to 1st, 2nd, 3rd, 4th, and 5th fractions, respectively.

Discussion

Because of the 1.25 mm CT slice thickness used, the lower limit of accuracy of our measurement was 1.25 mm. As a result, our measurement data were written as single figures, such as “1 mm”. Several investigators have already reported caudal catheter displacement during HDR brachytherapy (Table 3) [3–7, 9, 10]. Although they used 3-5 mm CT slice thickness, mean displacements resembles our results were reported.
Table 3

Reported data for catheter displacement during high-dose-rate brachytherapy

AuthorYear n ModalityReference pointMean displacementNumber of fractions/daysProstate volume
Takenaka2012303-mm CTCenter of 3 gold makers4.3 mm between 0-2 fractions7 fractions/4 daysna
Whitaker2011253-mm CT3 gold markers7.5 mm * between 0-1 fraction (median 254 min)2 fractions/ 2 weeks/2 implantationna
Simnor2009203-mm CTProstate image on CT7.9 mm between 1-23 fractions/2 daysna
Kim2007103-mm CTIschial bone Center of two gold markers2.7 mm between 1-2 fractions 5.5 mm between 1-2 fractions2 fractions/1 daysna
Mullokandov2004502-5-mm CTIschial bone2 mm between 1-24 fractions/24-28 hoursna
Hoskin2003205-mm CTProstate image on CT9.7 mm between 1-22 fractions/36 hoursna
Martinez200110FluoroscopyBony anatomy20 mm between 1-2 fractions 4 mm between 3-4 fractions4 fractions/2 days2.7% between 1 to 4 fraction
Damore200096Film3 gold markers7.6 mm between 1-2 fractions 2 mm between 3-4 fractions4 fractions/40 hoursBase-apex, mean 5.2-5.7 mm
Present study301.25-mm CTApex gold marker12 mm between 0-2 fractions5 fractions/3 days3 mm between 0-5 fractions

Median displacement

Reported data for catheter displacement during high-dose-rate brachytherapy Median displacement In our previous protocol, catheter displacements were checked by X-ray films and adjusted if displacement was > 5 mm relative to implanted markers. However, 10% of our patients suffered from grade 3 genitourinary toxicity including urethral stricture [2]. Because caudal catheter displacement could be one of the reasons for urethral stricture [8, 12], we changed our action level from 5 mm down to 2 mm with the aim of achieving a lower urethral dose. With this new protocol, exposure of the bulbo-membranous urethra to high-dose radiation could be avoided, and the urethral toxicity rate could be reduced. However, a longer follow-up time for the current patient cohort is required. Our results also confirmed that the distances between apex and base markers were increased, suggesting that the prostate gland was gradually swelling throughout the course of treatment [13]. Herrmann et al. [14] also reported displacement of markers after HDR brachytherapy, especially in the superior-inferior direction (mean: 3 mm). They discussed possible reasons for displacement such as localized bleeding into the prostate gland, or dislocation of prostate tissue inside the gland due to needle insertion. Although we added a 3- to 5-mm margin around the prostate, some parts of the PTV may not have received sufficient dose in some cases. Re-planning may be needed for patients with an excessively enlarged prostate after catheter insertion. Further investigation is required to resolve this problem. On the other hand, our results showed no significant displacement of apex markers. The apex is, in comparison to other structures around the prostate, less dependent on internal influences such as swelling of the prostate, therefore it may represent a suitable position for marker implantation compared to the base of the gland. Tiong et al. [15] examined the impact of catheter displacement on tumor control probability (TCP) in patients with prostate cancer receiving HDR, and advised that action levels to correct for catheter displacements should be ≤ 3 mm. According to their calculations, median relative TCP was 0.998, if catheter displacement was 3 mm. We therefore set our catheter displacement action level at 2 mm (Fig. 3) to include a safety margin. Another interesting point we found in this study was the difficulty of manually adjusting the catheter. Physically advancing the catheter 1 cm into the patient resulted in an actual advance of only approximately 0.4 cm relative to the apex marker, as the prostate itself was also pushed along with the catheter. This result suggests that a combination with computerized adjustment of dwell position in software is needed when manual catheter advancement is performed.

Conclusions

In conclusion, frequent catheter displacements relative to the apex of the prostate and bulbo-membranous urethra were confirmed by measurement prior to each treatment fraction on pre-treatment CT images with 1.25 mm slice thickness. Our results indicate that catheter positions must be confirmed and if required, adjusted, prior to every treatment fraction for the precise treatment delivery of HDR brachytherapy, and to potentially reduce over-dosage to the bulbo-membranous urethra.
  15 in total

1.  Analysis of serial CT scans to assess template and catheter movement in prostate HDR brachytherapy.

Authors:  Eduard Mullokandov; Glen Gejerman
Journal:  Int J Radiat Oncol Biol Phys       Date:  2004-03-15       Impact factor: 7.038

2.  High dose rate afterloading brachytherapy for prostate cancer: catheter and gland movement between fractions.

Authors:  Peter J Hoskin; Peter J Bownes; Peter Ostler; Ken Walker; Linda Bryant
Journal:  Radiother Oncol       Date:  2003-09       Impact factor: 6.280

3.  Justification for inter-fraction correction of catheter movement in fractionated high dose-rate brachytherapy treatment of prostate cancer.

Authors:  Tania Simnor; Sonia Li; Gerry Lowe; Peter Ostler; Linda Bryant; Caroline Chapman; Dave Inchley; Peter J Hoskin
Journal:  Radiother Oncol       Date:  2009-10-23       Impact factor: 6.280

4.  A small tolerance for catheter displacement in high-dose rate prostate brachytherapy is necessary and feasible.

Authors:  Albert Tiong; Sean Bydder; Martin Ebert; Nikki Caswell; David Waterhouse; Nigel Spry; Peter Camille; David Joseph
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-07-18       Impact factor: 7.038

5.  Prostate HDR brachytherapy catheter displacement between planning and treatment delivery.

Authors:  May Whitaker; George Hruby; Aimee Lovett; Nitya Patanjali
Journal:  Radiother Oncol       Date:  2011-08-31       Impact factor: 6.280

6.  Phase II prospective study of the use of conformal high-dose-rate brachytherapy as monotherapy for the treatment of favorable stage prostate cancer: a feasibility report.

Authors:  A A Martinez; I Pataki; G Edmundson; E Sebastian; D Brabbins; G Gustafson
Journal:  Int J Radiat Oncol Biol Phys       Date:  2001-01-01       Impact factor: 7.038

7.  Toxicity and health-related quality of life during and after high dose rate brachytherapy followed by external beam radiotherapy for prostate cancer.

Authors:  S Egawa; S Shimura; A Irie; M Kitano; I Nishiguchi; S Kuwao; K Hayakawa; S Baba
Journal:  Jpn J Clin Oncol       Date:  2001-11       Impact factor: 3.019

8.  Gold marker displacement due to needle insertion during HDR-brachytherapy for treatment of prostate cancer: a prospective cone beam computed tomography and kilovoltage on-board imaging (kV-OBI) study.

Authors:  Markus K A Herrmann; Tereza Kertesz; Tammo Gsänger; Eugen Bloch; Gerhard Pollul; Mohamed Bouabdallaoui; Arne Strauss; Mareike Herrmann; Hans Christiansen; Hendrik A Wolff; Clemens F Hess; Andrea Hille
Journal:  Radiat Oncol       Date:  2012-02-20       Impact factor: 3.481

9.  Case series analysis of post-brachytherapy prostate edema and its relevance to post-implant dosimetry. Post-implant prostate edema and dosimetry.

Authors:  Ajay Tejwani; Eva Bieniek; Lindsay Puckett; Amir Lavaf; Adel Guirguis; Aaron Bennish; Hani Ashamalla
Journal:  J Contemp Brachytherapy       Date:  2012-06-30

10.  Measurement of craniocaudal catheter displacement between fractions in computed tomography-based high dose rate brachytherapy of prostate cancer.

Authors:  Yongbok Kim; I-Chow J Hsu; Jean Pouliot
Journal:  J Appl Clin Med Phys       Date:  2007-09-17       Impact factor: 2.102

View more
  1 in total

1.  Focal MRI-Guided Salvage High-Dose-Rate Brachytherapy in Patients With Radiorecurrent Prostate Cancer.

Authors:  Metha Maenhout; Max Peters; Marco van Vulpen; Marinus A Moerland; Richard P Meijer; Maurice A A J van den Bosch; Paul L Nguyen; Steven J Frank; Jochem R N van der Voort van Zyp
Journal:  Technol Cancer Res Treat       Date:  2017-12-05
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.