Literature DB >> 10725633

Needle displacement during HDR brachytherapy in the treatment of prostate cancer.

S J Damore1, A M Syed, A A Puthawala, A Sharma.   

Abstract

PURPOSE: We used clinical patient data to examine implant displacement between high dose rate (HDR) brachytherapy fractions for prostate cancer to determine its impact on treatment delivery.
MATERIALS AND METHODS: We analyzed the verification films taken prior to each fraction for 96 consecutive patients treated with HDR brachytherapy boosts as part of their radiation therapy for definitive treatment of organ-confined prostate cancer at our institution. Patients were treated with 18-24 Gy in 4 fractions of HDR delivered in 40 hours followed by 36-39.6 Gy external beam radiation to the prostate. We determined the mean and maximum displacement distances of marker seeds placed in the prostate and of the implanted needles between HDR fractions.
RESULTS: Mean and maximum displacement distances between fractions were documented up to 7.6 mm and 28.5 mm, respectively, for the implant needles and 3.6 mm and 11.4 mm, respectively, for the gold marker seeds. All displacement of implant needles occurred in the caudal direction. At least 1 cm caudal displacement of needles occurred prior to 15.5% all fractions. Manual adjustment of needles was required prior to 15% of fractions, and adjustment of the CLP only was required in 24%. Most of the displacement for both the marker seeds and needles occurred between the first and second fractions.
CONCLUSIONS: There is significant caudal displacement of interstitial implant needles between HDR fractions in our prostate cancer patients. Obtaining verification films and making adjustments in the treatment volume prior to each fraction is necessary to avoid significant inaccuracies in treatment delivery.

Entities:  

Mesh:

Year:  2000        PMID: 10725633     DOI: 10.1016/s0360-3016(99)00477-0

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  18 in total

1.  New ambulatory implant technique of high-dose-rate interstitial brachytherapy for prostate cancer.

Authors:  Ken Yoshida; Takayuki Nose; Hiroya Shiomi; Yasuo Yoshioka; Yuka Fujita; Satoru Kuroda; Mineo Yoshida; Tohru Takahashi; Masaya Kitamura; Hideyuki Akai; Toshitsugu Oka; Takuya Hosoki
Journal:  Radiat Med       Date:  2006-10

2.  Phase II trial of combined high-dose-rate brachytherapy and external beam radiotherapy for adenocarcinoma of the prostate: preliminary results of RTOG 0321.

Authors:  I-Chow Hsu; Kyounghwa Bae; Katsuto Shinohara; Jean Pouliot; James Purdy; Geoffrey Ibbott; Joycelyn Speight; Eric Vigneault; Robert Ivker; Howard Sandler
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-03-06       Impact factor: 7.038

3.  Intraoperative 360-deg three-dimensional transvaginal ultrasound during needle insertions for high-dose-rate transperineal interstitial gynecologic brachytherapy of vaginal tumors.

Authors:  Jessica Robin Rodgers; Jeffrey Bax; Kathleen Surry; Vikram Velker; Eric Leung; David D'Souza; Aaron Fenster
Journal:  J Med Imaging (Bellingham)       Date:  2019-04-08

4.  The impact of prostate volume changes during external-beam irradiation in consequence of HDR brachytherapy in prostate cancer treatment.

Authors:  Markus Karl Alfred Herrmann; Tammo Gsänger; Arne Strauss; Tereza Kertesz; Hendrik A Wolff; Hans Christiansen; Hilke Vorwerk; Clemens Friedrich Hess; Andrea Hille
Journal:  Strahlenther Onkol       Date:  2009-06-09       Impact factor: 3.621

5.  High-dose-rate interstitial brachytherapy for gynecologic malignancies--dosimetric changes during treatment period.

Authors:  Tsuyoshi Onoe; Takayuki Nose; Hideomi Yamashita; Minoru Yoshioka; Takashi Toshiyasu; Takuyo Kozuka; Masahiko Oguchi; Keiichi Nakagawa
Journal:  J Radiat Res       Date:  2013-01-07       Impact factor: 2.724

6.  Dosimetric comparison of inverse optimization with geometric optimization in combination with graphical optimization for HDR prostate implants.

Authors:  Swamidas V Jamema; Sherly Saju; Umesh M Shetty; Siddanna Pallad; D D Deshpande; S K Shrivastava
Journal:  J Med Phys       Date:  2006-04

7.  Evaluation of a robotic technique for transrectal MRI-guided prostate biopsies.

Authors:  Martijn G Schouten; Joyce G R Bomers; Derya Yakar; Henkjan Huisman; Eva Rothgang; Dennis Bosboom; Tom W J Scheenen; Sarthak Misra; Jurgen J Fütterer
Journal:  Eur Radiol       Date:  2011-09-29       Impact factor: 5.315

Review 8.  Review of advanced catheter technologies in radiation oncology brachytherapy procedures.

Authors:  Jun Zhou; Leonid Zamdborg; Evelyn Sebastian
Journal:  Cancer Manag Res       Date:  2015-07-16       Impact factor: 3.989

Review 9.  The emerging role of high-dose-rate (HDR) brachytherapy as monotherapy for prostate cancer.

Authors:  Yasuo Yoshioka; Ken Yoshida; Hideya Yamazaki; Norio Nonomura; Kazuhiko Ogawa
Journal:  J Radiat Res       Date:  2013-03-29       Impact factor: 2.724

10.  Effect of using different U/S probe Standoff materials in image geometry for interventional procedures: the example of prostate.

Authors:  Stefanos Diamantopoulos; Natasa Milickovic; Saeed Butt; Zaira Katsilieri; Vasiliki Kefala; Pawel Zogal; George Sakas; Dimos Baltas
Journal:  J Contemp Brachytherapy       Date:  2011-12-30
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.