Retrospective evaluation of the accuracy of Roche AccuChek Inform and Nova StatStrip glucose meters when used on critically ill patients.

Brad S Karon1, Chanse T Blanshan, Gayle R Deobald, Amy M Wockenfus.   

Abstract

BACKGROUND: More stringent accuracy guidelines for hospital-use glucose meters have recently been published, but it remains unclear whether glucose meters can meet these accuracy guidelines when measurement is performed on critically ill patients with fresh whole blood samples.
MATERIALS AND METHODS: We performed a retrospective evaluation of a conventional (Roche Diagnostics [Indianapolis, IN] AccuChek® Inform) and a newer-generation (Nova Biomedical [Waltham, MA] StatStrip®) glucose system by comparing paired (drawn within 5 min of each other) whole blood glucose meter and laboratory serum glucose values obtained from intensive care unit (ICU) patients. We also performed a prospective evaluation of the accuracy of the Nova StatStrip.
RESULTS: The median (interquartile range) bias between Roche AccuChek Inform and serum laboratory glucose measurements was 11 (6-18) mg/dL, compared with a median bias between the Nova StatStrip and serum glucose measurements of 1 (-5 to 5) mg/dL. StatStrip met International Organization for Standardization 15197 and Clinical and Laboratory Standards Institute (CLSI) POCT12-A3 accuracy guidelines using both retrospective and prospective datasets.
CONCLUSIONS: The newer-generation (StatStrip) glucose meter met more stringent CLSI POCT12-A3 accuracy criteria because of reduced bias compared with the previous-generation device. Reduced glucose meter bias led to fewer insulin dosing discrepancies when the insulin dose determined from serum glucose was compared with that determined from the glucose meter value using the institutional glycemic control protocol.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25093919     DOI: 10.1089/dia.2014.0074

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  9 in total

1.  Impact of Glucose Meter Error on Glycemic Variability and Time in Target Range During Glycemic Control After Cardiovascular Surgery.

Authors:  Brad S Karon; Jeffrey W Meeusen; Sandra C Bryant
Journal:  J Diabetes Sci Technol       Date:  2015-08-25

2.  Clinical Impact of Accurate Point-of-Care Glucose Monitoring for Tight Glycemic Control in Severely Burned Children.

Authors:  Nam K Tran; Zachary R Godwin; Amanda N Steele; Steven E Wolf; Tina L Palmieri
Journal:  Pediatr Crit Care Med       Date:  2016-09       Impact factor: 3.624

3.  Unintended Consequence of High-Dose Vitamin C Therapy for an Oncology Patient: Evaluation of Ascorbic Acid Interference With Three Hospital-Use Glucose Meters.

Authors:  Brooke M Katzman; Brandon R Kelley; Gayle R Deobald; Nikki K Myhre; Sean A Agger; Brad S Karon
Journal:  J Diabetes Sci Technol       Date:  2020-06-07

4.  Accuracy of Capillary and Arterial Whole Blood Glucose Measurements Using a Glucose Meter in Patients under General Anesthesia in the Operating Room.

Authors:  Brad S Karon; Leslie J Donato; Chelsie M Larsen; Lindsay K Siebenaler; Amy E Wells; Christina M Wood-Wentz; Mary E Shirk-Marienau; Timothy B Curry
Journal:  Anesthesiology       Date:  2017-09       Impact factor: 7.892

5.  Analytical Performance Evaluation of Infopia Element™ Auto-coding Blood Glucose Monitoring System for Self-Monitoring of Blood Glucose.

Authors:  Hae-Il Park; Seong-Su Lee; Jang-Won Son; Hee-Sun Kwon; Sung Rae Kim; Hyojin Chae; Myungshin Kim; Yonggoo Kim; Soonjib Yoo
Journal:  J Clin Lab Anal       Date:  2016-04-13       Impact factor: 2.352

6.  Accuracy of blood glucose measurements using the NOVA StatStrip® glucometer during cardiac surgery: a prospective observational study.

Authors:  Yosuke Nakadate; Hiroaki Sato; Patricia Roque; Tamaki Sato; Takashi Matsukawa; Linda Wykes; Akiko Kawakami; Thomas Schricker
Journal:  Can J Anaesth       Date:  2019-03-20       Impact factor: 5.063

7.  Glucose intolerance associated with hypoxia in people living at high altitudes in the Tibetan highland.

Authors:  Kiyohito Okumiya; Ryota Sakamoto; Yasuko Ishimoto; Yumi Kimura; Eriko Fukutomi; Motonao Ishikawa; Kuniaki Suwa; Hissei Imai; Wenling Chen; Emiko Kato; Masahiro Nakatsuka; Yoriko Kasahara; Michiko Fujisawa; Taizo Wada; Hongxin Wang; Qingxiang Dai; Huining Xu; Haisheng Qiao; Ri-Li Ge; Tsering Norboo; Norboo Tsering; Yasuyuki Kosaka; Mitsuhiro Nose; Takayoshi Yamaguchi; Toshihiro Tsukihara; Kazuo Ando; Tetsuya Inamura; Shinya Takeda; Masayuki Ishine; Kuniaki Otsuka; Kozo Matsubayashi
Journal:  BMJ Open       Date:  2016-02-23       Impact factor: 2.692

8.  Assessment of the performance of blood glucose monitoring systems for monitoring dysglycaemia in neonatal patients.

Authors:  Yin Ba; Jin Xu; Lin Yuan; Haiyan Zhu; Yipei Yang; Mei Mei Lam; Shulian Zhang; Mili Xiao; Jiayin Xu; Rong Zhang; Chao Chen
Journal:  BMJ Paediatr Open       Date:  2018-10-23

9.  Comparison of the FreeStyle Libre Pro Flash Continuous Glucose Monitoring (CGM) System and Point-of-Care Capillary Glucose Testing in Hospitalized Patients With Type 2 Diabetes Treated With Basal-Bolus Insulin Regimen.

Authors:  Rodolfo J Galindo; Alexandra L Migdal; Georgia M Davis; Maria A Urrutia; Bonnie Albury; Cesar Zambrano; Priyathama Vellanki; Francisco J Pasquel; Maya Fayfman; Limin Peng; Guillermo E Umpierrez
Journal:  Diabetes Care       Date:  2020-07-08       Impact factor: 17.152

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.