| Literature DB >> 25086594 |
Andrea Aebischer, Martin Beer, Bernd Hoffmann1.
Abstract
BACKGROUND: In order to control and eradicate transboundary animal diseases, early diagnosis and reaction is essential for the implementation of control activities. Thus, mobile diagnostic units which allow analytical testing close to the site of occurrence could provide valuable support for centralized laboratories. Consequently, the availability of diagnostic tests using mobile amplification and detection technologies has been increasing over the past years. However, methods enabling rapid and simple nucleic acid extraction also under resource-limited settings are still scarce.Entities:
Mesh:
Substances:
Year: 2014 PMID: 25086594 PMCID: PMC4124660 DOI: 10.1186/1743-422X-11-137
Source DB: PubMed Journal: Virol J ISSN: 1743-422X Impact factor: 4.099
Speed-optimized extraction protocols for the KF Duo and the BS 15 instrument
| Binding | VXL + ACB + Beads | 5 | 2 | 2 |
| Wash 1 | AW1 | 3 | 12 s | 1 |
| Wash 2 | AW2 | 2 | 12 s | 1 |
| Wash 3 | Ethanol 100% | 2 | 12 s | - |
| Evaporation | Water | 5 | 0 | 0 |
| Elution | AVE | 2 | 1 | 0.5 |
| Total time | 24:00 | 8:00 | 7:50 | |
Each step of the original protocol was shortened to the maximum for the respective extraction platform.
Figure 1Analytical sensitivity and linear dynamic range of manual and rapid automated extraction methods. Dilution series of serum samples positive for SBV-RNA (A) or BVDV-RNA (B) as well as dilution series of EDTA-blood samples positive for SBV-RNA (C) or BVDV-RNA (D) were used to evaluate the extraction efficiency of the rapid extraction protocols in comparison to the standard manual extraction protocol. Each dilution step was extracted in duplicate and the extracted RNA from each panel was tested in the same RT-qPCR run. R2 values as defined by regression analysis are indicated for each method.
Reproducibility of manual and rapid automated extraction techniques
| Manual | 23.83 | 0.19 | 0.80 | 0.14 | 0.58 |
| EZ1 | 24.10 | 0.16 | 0.65 | 0.12 | 0.51 |
| BS 15 | 25.12 | 0.34 | 1.36 | 0.28 | 1.10 |
| KF Duo | 25.31 | 0.25 | 1.00 | 0.20 | 0.79 |
SD = standard deviation, CV = coefficient of variation.
Mean Cq values, standard deviations of replicates as well as intra- and inter-assay coefficients of variation were defined using 6 replicates of an SBV-positive blood sample in 3 independent runs.
PCR efficiency after manual and automated extraction
| SBV | Blood | 99.6 | 97.0 | 86.1 | 101.2 |
| Serum | 97.8 | 121.5 | 87.1 | 112.7 | |
| BVDV | Blood | 112.4 | 99.3 | 88.8 | 84.2 |
| Serum | 95.0 | 98.2 | 92.8 | 110.8 | |
10-fold dilution series of SBV- and BVDV-positive blood and serum samples which were extracted in duplicate; PCR efficiencies of the respective RNA extracts were calculated from the slope of the standard curves.
Figure 2Analytical performance of rapid automated extraction using clinical samples. Direct comparison of standard manual extraction with automated extraction procedures using the EZ1 Advanced XL (A), the BS 15 (B), and the KF Duo (C). For this purpose, blood and serum samples positive for SBV-RNA (black circles) or BVDV-RNA (white circles) were used. Regression is indicated by solid lines for SBV-positive samples and by dashed lines for BVDV-positive samples.
Operational analysis of manual and automated extraction procedures
| Filling of reagents | 5
| 15 | 5 | 5 |
| Mix sample + lysis buffer | 10 | 5 | 10 | 10 |
| Incubation | 10 | 0 | 0 | 0 |
| Load instrument | 0 | 5 | 1 | 1 |
| Extraction/Instrument run | 30 | 16 | 7.5 | 8 |
| Total hands-on time | 40 - 45 | 25 | 16 | 16 |
| Total processing time | 50 - 55 | 41 | 24 | 24 |
dissolving carrier RNA.
The time required for each operator step was measured for the extraction of 8 samples, respectively.