Literature DB >> 20225947

A rapid and automated sample-to-result HIV load test for near-patient application.

Sultan Tanriverdi1, Lingjun Chen, Shuqi Chen.   

Abstract

BACKGROUND: Current viral load tests for human immunodeficiency virus (HIV) can only be performed in laboratory environments, require highly trained operators and expensive equipment, and have a turnaround time of several hours to days. The Liat HIV Quant Assay proves that such nucleic acid testing can be performed rapidly and easily, allowing application at the point of care.
METHODS: The Liat Analyzer automates the entire assay, including sample preparation, amplification, and detection, in a self-contained and closed-tube system. Dynamic range, limit of detection, and subtype specificity were evaluated. Clinical samples were tested retrospectively to correlate the result of this assay with those of commercial HIV load assays.
RESULTS: The Liat assay demonstrated linearity of >6 logs (R(2), 0.98) and a limit of detection of 57 copies/mL. HIV-1 group M (clades A-H), group O, and HIV-2 were detected. Testing of clinical samples showed a high degree of concordance between the copy numbers detected by the Liat assay and those detected by the Siemens and Roche assays (92.0% and 88% correlation coefficient of the log copy number, respectively). The time from sample collection to result was 88 min.
CONCLUSIONS: Results suggest that the Liat HIV Quant Assay has performance equivalent to that of commercial HIV load assays and significantly reduces assay time, simplifies test operation, and provides biocontainment to allow operation in nonlaboratory settings.

Entities:  

Mesh:

Year:  2010        PMID: 20225947     DOI: 10.1086/650387

Source DB:  PubMed          Journal:  J Infect Dis        ISSN: 0022-1899            Impact factor:   5.226


  40 in total

1.  Point-of-care testing.

Authors:  David A Anderson; Suzanne M Crowe; Mary Garcia
Journal:  Curr HIV/AIDS Rep       Date:  2011-03       Impact factor: 5.071

2.  HIV testing updates and challenges: when regulatory caution and public health imperatives collide.

Authors:  Bernard M Branson
Journal:  Curr HIV/AIDS Rep       Date:  2015-03       Impact factor: 5.071

3.  Comparison of the GenMark Diagnostics eSensor respiratory viral panel to real-time PCR for detection of respiratory viruses in children.

Authors:  Virginia M Pierce; Richard L Hodinka
Journal:  J Clin Microbiol       Date:  2012-08-08       Impact factor: 5.948

4.  A two-magnet strategy for improved mixing and capture from biofluids.

Authors:  Thomas F Scherr; Hayley B Ryskoski; Andrew B Doyle; Frederick R Haselton
Journal:  Biomicrofluidics       Date:  2016-04-11       Impact factor: 2.800

5.  Laboratory evaluation of the Liat HIV Quant (IQuum) whole-blood and plasma HIV-1 viral load assays for point-of-care testing in South Africa.

Authors:  Lesley Scott; Natasha Gous; Sergio Carmona; Wendy Stevens
Journal:  J Clin Microbiol       Date:  2015-03-04       Impact factor: 5.948

6.  Diagnostic tools for tackling febrile illness and enhancing patient management.

Authors:  Konstantinos Mitsakakis; Valérie D'Acremont; Sebastian Hin; Felix von Stetten; Roland Zengerle
Journal:  Microelectron Eng       Date:  2018-10-05       Impact factor: 2.523

Review 7.  Developments in CD4 and viral load monitoring in resource-limited settings.

Authors:  Christopher F Rowley
Journal:  Clin Infect Dis       Date:  2013-11-11       Impact factor: 9.079

Review 8.  Diagnosis of Human Immunodeficiency Virus Infection.

Authors:  Bharat S Parekh; Chin-Yih Ou; Peter N Fonjungo; Mireille B Kalou; Erin Rottinghaus; Adrian Puren; Heather Alexander; Mackenzie Hurlston Cox; John N Nkengasong
Journal:  Clin Microbiol Rev       Date:  2018-11-28       Impact factor: 26.132

Review 9.  Nucleic acid testing for tuberculosis at the point-of-care in high-burden countries.

Authors:  Angelika Niemz; David S Boyle
Journal:  Expert Rev Mol Diagn       Date:  2012-09       Impact factor: 5.225

10.  Low-cost tools for diagnosing and monitoring HIV infection in low-resource settings.

Authors:  Grace Wu; Muhammad H Zaman
Journal:  Bull World Health Organ       Date:  2012-10-19       Impact factor: 9.408

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.