Abstract

PURPOSE: Catheter ablation is a common treatment option for atrial fibrillation (AF). Interventional C-arm X-ray systems are used for guiding AF procedures, employing standard view positions. Since the projection angles are not adapted to the individual patient anatomy, standard projections do not necessarily offer the best views of important anatomical structures. Using a pre-procedural 3D data set acquired with MRI or CT, suitable ablation sites (lines) can be identified in advance so an ablation plan can be superimposed on fluoroscopic images to guide the procedure.
METHODS: A method was developed to estimate optimized projection views for biplane X-ray C-arm systems based on planning data for AF ablation procedures. The estimated viewing angles were compared to standard angulations using an objective quality metric, the length of the planned ablation line as seen under X-ray. This method was tested using 35 clinical datasets annotated with planned ablation lines for ipsilateral pulmonary vein isolation.
RESULTS: The optimized views computed using the new method yielded 28 % less foreshortening of pre-planned ablation lines on average. In one case, anatomy-based view calculation lead to a 69 % reduction in foreshortening.
CONCLUSION: The commonly used standard views provide reasonable a priori choices, and some improvement is possible by switching among common angulations depending on the treatment region. Further gains are possible by using anatomy-optimized biplane C-arm angulations.

Entities:  

Mesh:

Year:  2014        PMID: 25081738     DOI: 10.1007/s11548-014-1103-z

Source DB:  PubMed          Journal:  Int J Comput Assist Radiol Surg        ISSN: 1861-6410            Impact factor:   2.924


  14 in total

1.  3-D reconstruction of coronary arterial tree to optimize angiographic visualization.

Authors:  S J Chen; J D Carroll
Journal:  IEEE Trans Med Imaging       Date:  2000-04       Impact factor: 10.048

2.  Multislice computed tomography for prediction of optimal angiographic deployment projections during transcatheter aortic valve implantation.

Authors:  Ronen Gurvitch; David A Wood; Jonathon Leipsic; Edgar Tay; Mark Johnson; Jian Ye; Fabian Nietlispach; Namal Wijesinghe; Anson Cheung; John G Webb
Journal:  JACC Cardiovasc Interv       Date:  2010-11       Impact factor: 11.195

3.  An augmented reality system for patient-specific guidance of cardiac catheter ablation procedures.

Authors:  Stijn De Buck; Frederik Maes; Joris Ector; Jan Bogaert; Steven Dymarkowski; Hein Heidbüchel; Paul Suetens
Journal:  IEEE Trans Med Imaging       Date:  2005-11       Impact factor: 10.048

4.  Determination of optimal angiographic viewing angles: basic principles and evaluation study.

Authors:  A M Dumay; J C Reiber; J J Gerbrands
Journal:  IEEE Trans Med Imaging       Date:  1994       Impact factor: 10.048

5.  Electrophysiological end point for catheter ablation of atrial fibrillation initiated from multiple pulmonary venous foci.

Authors:  M Haïssaguerre; P Jaïs; D C Shah; S Garrigue; A Takahashi; T Lavergne; M Hocini; J T Peng; R Roudaut; J Clémenty
Journal:  Circulation       Date:  2000-03-28       Impact factor: 29.690

Review 6.  Catheter-based left atrial appendage occlusion procedure: role of echocardiography.

Authors:  Gila Perk; Simon Biner; Itzhak Kronzon; Muhamed Saric; Larry Chinitz; Keith Thompson; Takahiro Shiota; Asma Hussani; Roberto Lang; Robert Siegel; Saibal Kar
Journal:  Eur Heart J Cardiovasc Imaging       Date:  2011-09-08       Impact factor: 6.875

7.  Angiographic views used for percutaneous coronary interventions: a three-dimensional analysis of physician-determined vs. computer-generated views.

Authors:  Nathan E Green; S-Y James Chen; Adam R Hansgen; John C Messenger; Bertron M Groves; John D Carroll
Journal:  Catheter Cardiovasc Interv       Date:  2005-04       Impact factor: 2.692

Review 8.  Balloon catheters for pulmonary vein isolation.

Authors:  Boris Schmidt; Kyoung Ryul Julian Chun; Andreas Metzner; Feifan Ouyang; Karl-Heinz Kuck
Journal:  Herz       Date:  2008-12       Impact factor: 1.443

9.  Optimal fluoroscopic projections for angiographic imaging of the pulmonary vein ostia: lessons learned from the intraprocedural reconstruction of the left atrium and pulmonary veins.

Authors:  Min Tang; Jin-Hong Gerds-Li; Sotirios Nedios; Mattias Roser; Eckart Fleck; Charalampos Kriatselis
Journal:  Europace       Date:  2010-01       Impact factor: 5.214

10.  Pulmonary vein isolation supported by MRI-derived 3D-augmented biplane fluoroscopy: a feasibility study and a quantitative analysis of the accuracy of the technique.

Authors:  Felix Bourier; Dejan Vukajlovic; Alexander Brost; Joachim Hornegger; Norbert Strobel; Klaus Kurzidim
Journal:  J Cardiovasc Electrophysiol       Date:  2012-11-06
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.