AIM: The aim of this in vivo study was to evaluate the accuracy and reliability of electronic apex locater and radiographic determination of root canal length in primary teeth.
METHODOLOGY: A total of 32 human primary molar teeth (96 roots) were selected. After endodontic access preparation, root canals were irrigated with physiological saline solution. The access cavities were dried with cotton pellets and, the roots were dried with paper points before performing the electronic measurement. The root canal length measurements were first taken with an apex locater (EndoMaster), and then, a size ♯ 15 K-file was inserted into the root canal, and radiography was taken to determine the working length measurements. The measurement data were recorded and compared by one-way anova and Kruskal-Wallis tests. P < 0.05 was accepted for the significance.
RESULTS: The mean values for radiography were 13.23 ± 1.92 mm and for EndoMaster were 13.08 ± 1.77 mm. The accuracy of EndoMaster was 80.2% in correct measurements ±1 mm (P < 0.001).
CONCLUSION: The electronic apex locators could be useful in determining working length and thereby decreasing the need for radiographs and exposure to ionizing radiation in pediatric dental patients.
© 2014 BSPD, IAPD and John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.

Entities:  

Mesh:

Year:  2014        PMID: 25073636     DOI: 10.1111/ipd.12128

Source DB:  PubMed          Journal:  Int J Paediatr Dent        ISSN: 0960-7439            Impact factor:   3.455


  5 in total

1.  In vitro comparison between apex locators, direct and radiographic techniques for determining the root canal length in primary teeth.

Authors:  L Z Adriano; J C Barasuol; M Cardoso; M Bolan
Journal:  Eur Arch Paediatr Dent       Date:  2019-01-02

2.  Evaluation of the Working Length Determination Accuracy by Cone-beam Computed Tomography in Primary Teeth.

Authors:  John Shibin; Prathima Gs; Suganya M; Nandhakumar S; Sanguida Adimoulame; Kavitha M
Journal:  Int J Clin Pediatr Dent       Date:  2022

3.  Comparative evaluation of root ZX Mini® apex locator and radiovisiography in determining the working length of primary molars: An In Vivo study.

Authors:  Rucha N Davalbhakta; Niraj S Gokhale; Shivayogi M Hugar; Chandrashekhar M Badakar; A Gowtham; Sanjana P Soneta
Journal:  J Oral Biol Craniofac Res       Date:  2021-02-17

4.  A Comparative Evaluation of Accuracy of New-generation Electronic Apex Locator with Conventional Radiography to determine Working Length in Primary Teeth: An in vivo Study.

Authors:  K Vidya Bhat; Prakashchandra Shetty; Latha Anandakrishna
Journal:  Int J Clin Pediatr Dent       Date:  2017-02-27

5.  Postoperative Pain with Hand, Reciprocating, and Rotary Instrumentation Techniques after Root Canal Preparation in Primary Molars: A Randomized Clinical Trial.

Authors:  Ganesh Jeevanandan; Vignesh Ravindran; Erulappan Mg Subramanian; Aravind S Kumar
Journal:  Int J Clin Pediatr Dent       Date:  2020 Jan-Feb
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.