Literature DB >> 25046981

Seizure detection with automated EEG analysis: a validation study focusing on periodic patterns.

Alba Sierra-Marcos1, Mark L Scheuer2, Andrea O Rossetti3.   

Abstract

OBJECTIVE: To evaluate an automated seizure detection (ASD) algorithm in EEGs with periodic and other challenging patterns.
METHODS: Selected EEGs recorded in patients over 1year old were classified into four groups: A. Periodic lateralized epileptiform discharges (PLEDs) with intermixed electrical seizures. B. PLEDs without seizures. C. Electrical seizures and no PLEDs. D. No PLEDs or seizures. Recordings were analyzed by the Persyst P12 software, and compared to the raw EEG, interpreted by two experienced neurophysiologists; Positive percent agreement (PPA) and false-positive rates/hour (FPR) were calculated.
RESULTS: We assessed 98 recordings (Group A=21 patients; B=29, C=17, D=31). Total duration was 82.7h (median: 1h); containing 268 seizures. The software detected 204 (=76.1%) seizures; all ictal events were captured in 29/38 (76.3%) patients; in only in 3 (7.7%) no seizures were detected. Median PPA was 100% (range 0-100; interquartile range 50-100), and the median FPR 0/h (range 0-75.8; interquartile range 0-4.5); however, lower performances were seen in the groups containing periodic discharges.
CONCLUSION: This analysis provides data regarding the yield of the ASD in a particularly difficult subset of EEG recordings, showing that periodic discharges may bias the results. SIGNIFICANCE: Ongoing refinements in this technique might enhance its utility and lead to a more extensive application.
Copyright © 2014 International Federation of Clinical Neurophysiology. Published by Elsevier Ireland Ltd. All rights reserved.

Entities:  

Keywords:  Electroencephalogram; Ictal; LPDs; PLEDs; Persyst; Seizure

Mesh:

Year:  2014        PMID: 25046981     DOI: 10.1016/j.clinph.2014.06.025

Source DB:  PubMed          Journal:  Clin Neurophysiol        ISSN: 1388-2457            Impact factor:   3.708


  4 in total

Review 1.  Recent applications of quantitative electroencephalography in adult intensive care units: a comprehensive review.

Authors:  Sung-Min Cho; Eva K Ritzl; Jaeho Hwang
Journal:  J Neurol       Date:  2022-08-19       Impact factor: 6.682

Review 2.  Big data sharing and analysis to advance research in post-traumatic epilepsy.

Authors:  Dominique Duncan; Paul Vespa; Asla Pitkänen; Adebayo Braimah; Niina Lapinlampi; Arthur W Toga
Journal:  Neurobiol Dis       Date:  2018-06-01       Impact factor: 5.996

3.  Improved epileptic seizure detection combining dynamic feature normalization with EEG novelty detection.

Authors:  J G Bogaarts; D M W Hilkman; E D Gommer; V H J M van Kranen-Mastenbroek; J P H Reulen
Journal:  Med Biol Eng Comput       Date:  2016-04-06       Impact factor: 2.602

4.  Validation of a smartphone-based EEG among people with epilepsy: A prospective study.

Authors:  Erica D McKenzie; Andrew S P Lim; Edward C W Leung; Andrew J Cole; Alice D Lam; Ani Eloyan; Damber K Nirola; Lhab Tshering; Ronald Thibert; Rodrigo Zepeda Garcia; Esther Bui; Sonam Deki; Liesly Lee; Sarah J Clark; Joseph M Cohen; Jo Mantia; Kate T Brizzi; Tali R Sorets; Sarah Wahlster; Mia Borzello; Arkadiusz Stopczynski; Sydney S Cash; Farrah J Mateen
Journal:  Sci Rep       Date:  2017-04-03       Impact factor: 4.379

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.