Allyson V Ritchie1, Ines Ushiro-Lumb2, Daniel Edemaga3, Hrishikesh A Joshi1, Annemiek De Ruiter4, Elisabeth Szumilin3, Isabelle Jendrulek4, Megan McGuire3, Neha Goel1, Pia I Sharma1, Jean-Pierre Allain5, Helen H Lee6.
Abstract
Routine viral-load (VL) testing of HIV-infected individuals on antiretroviral therapy (ART) is used to monitor treatment efficacy. However, due to logistical challenges, implementation of VL has been difficult in resource-limited settings. The aim of this study was to evaluate the performance of the SAMBA semi-Q (simple amplification-based assay semiquantitative test for HIV-1) in London, Malawi, and Uganda. The SAMBA semi-Q can distinguish between patients with VLs above and below 1,000 copies/ml. The SAMBA semi-Q was validated with diluted clinical samples and blinded plasma samples collected from HIV-1-positive individuals. SAMBA semi-Q results were compared with results from the Roche COBAS AmpliPrep/COBAS TaqMan HIV-1 test, v2.0. Testing of 96 2- to 10-fold dilutions of four samples containing HIV-1 subtype C as well as 488 samples from patients in the United Kingdom, Malawi, and Uganda yielded an overall accuracy for the SAMBA semi-Q of 99% (95% confidence interval [CI], 93.8 to 99.9%) and 96.9% (95% CI 94.9 to 98.3%), respectively, compared to to the Roche test. Analysis of VL data from patients in Malawi and Uganda showed that the SAMBA cutoff of 1,000 copies/ml appropriately distinguished treated from untreated individuals. Furthermore, analysis of the viral loads of 232 patients on ART in Malawi and Uganda revealed similar patterns for virological control, defined as either <1,000 copies/ml (SAMBA cutoff) or <5,000 copies/ml (WHO 2010 criterion; WHO, Antiretroviral Therapy for HIV Infection in Adults and Adolescents: Recommendations for a Public Health Approach, 2010). This study suggests that the SAMBA semi-Q has adequate concurrency with the gold standard measurements for viral load. This test can allow VL monitoring of patients on ART at the point of care in resource-limited settings.Entities:
Mesh:
Year: 2014 PMID: 25031444 PMCID: PMC4313165 DOI: 10.1128/JCM.00593-14
Source DB: PubMed Journal: J Clin Microbiol ISSN: 0095-1137 Impact factor: 5.948
FIG 1SAMBA semi-Q system. (A) SAMBAprep (right) and SAMBAamp (left) instruments. (B) SAMBAamp cartridge showing results for (i) >1,000 and (ii) <1,000 copies/ml and (iii) invalid results.
Validation of the SAMBA semi-Q cut-off with diluted plasma samples containing HIV-1 subtype C
| Sample | No. of HIV-1 RNA copies/ml | % with positive SAMBA HIV-1 semi-Q result (>1,000 copies/ml) |
|---|---|---|
| 1 | 151,114 | 100 |
| 15,111 | 100 | |
| 1,511 | 100 | |
| 756 | 0 | |
| 151 | 0 | |
| 15 | 0 | |
| 2 | 30,543 | 100 |
| 3,054 | 100 | |
| 1,527 | 100 | |
| 305 | 0 | |
| 31 | 0 | |
| 3 | 0 | |
| 3 | 222,238 | 100 |
| 22,224 | 100 | |
| 2,222 | 100 | |
| 1,111 | 100 | |
| 222 | 25 | |
| 22 | 0 | |
| 4 | 121,102 | 100 |
| 12,110 | 100 | |
| 1,211 | 75 | |
| 606 | 50 | |
| 121 | 0 | |
| 12 | 0 |
Four plasma samples were serially diluted to achieve concentrations of viral RNA ranging from 3 (0.48 log10) to 222,238 (5.35 log10) copies/ml according to quantification with Roche TaqMan v2. Four replicates of each dilution were tested with the SAMBA semi-Q.
Comparison of the SAMBA semi-Q and Roche TaqMan v2 results for 488 clinical samples from London (n = 134), Malawi (n = 200), and Uganda (n = 154)
| Sample source and SAMBA Semi-Q result (copies/ml) | No. with result with Roche TaqMan v2 (copies/ml) | |
|---|---|---|
| <1,000 | >1,000 | |
| London | ||
| <1,000 | 95 | 1 |
| >1,000 | 2 | 36 |
| Malawi | ||
| <1,000 | 146 | 0 |
| >1,000 | 4 | 50 |
| Uganda | ||
| <1,000 | 91 | 5 |
| >1,000 | 3 | 55 |
Concordance between the two tests was 96.9% (473/488) overall, 97.8% (131/134) in London, 98.0% (196/200) in Malawi, and 94.8% (146/154) in Uganda.
FIG 2Field testing algorithm for the SAMBA semi-Q with 354 samples collected in Malawi and Uganda and summary of results. All samples were tested with the SAMBA semi-Q and Roche TaqMan v2. Twelve samples were discrepant between SAMBA and Roche and were tested with Abbott RealTime. Ten of the 12 samples were discrepant between SAMBA and Abbott, and two were discrepant between Abbott and Roche.
FIG 3Distribution of VL among 284 patients receiving ART and 70 untreated individuals in Malawi and Uganda. VL was determined with Roche TaqMan v2.
FIG 4Virological suppression according to the SAMBA semi-Q cutoff and 2013 WHO guidelines (1,000 copies/ml [8]) or 2010 WHO guidelines (5,000 copies/ml [17]) in 232 patients on ART for various numbers of years in Uganda and Malawi.