Gafchromic film dosimetry: four years experience using FilmQA Pro software and Epson flatbed scanners.

M Mathot1, S Sobczak2, M-T Hoornaert2.   

Abstract

PURPOSE: s: To assess performance of FilmQA Pro software for pre-treatment patient-specific quality assurance (QA), using radiochromic films and two commercial flatbed scanners. To evaluate a novel multichannel approach compared to the classical red channel evaluation.
MATERIAL AND METHODS: Patient films (mostly EBT2 films, one box of EBT3) were digitalized using successively two flatbed scanners: the A4-size Epson V750 and the A3-size Epson 10000XL. Prior to patient dose verification, basic characteristics of films and scanners were investigated. Patient films were analyzed using FilmQA Pro software, which enables to use the signal from all three colour channels (Red, Green, Blue).
RESULTS: Compared to the red channel evaluation, multichannel evaluation presents better passing rates with regard to local gamma index. As expected, we obtained better results using A3-size scanner compared to A4-size scanner, especially when considering large region of interest. An observation of great interest was made for both scanners: after intensive use, a tilting in the blue transmittance profiles appeared in the lamp direction, making multichannel analysis unsuitable for accurate dose evaluation.
Copyright © 2014 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Flatbed scanners; GAFCHROMIC; Multichannel correction

Mesh:

Year:  2014        PMID: 25018050     DOI: 10.1016/j.ejmp.2014.06.043

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  8 in total

1.  SBRT/SRS patient-specific QA using GAFchromicTM EBT3 and FilmQATM Pro software.

Authors:  Giuseppe Stella; Nina Cavalli; Elisa Bonanno; Lucia Zirone; Giuseppina Rita Borzì; Martina Pace; Andrea Girlando; Anna M Gueli; Carmelo Marino
Journal:  J Radiosurg SBRT       Date:  2022

2.  Gafchromic EBT3 film dosimetry in electron beams - energy dependence and improved film read-out.

Authors:  Petri Sipilä; Jarkko Ojala; Sampsa Kaijaluoto; Ilkka Jokelainen; Antti Kosunen
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

Review 3.  Influence of the jaw tracking technique on the dose calculation accuracy of small field VMAT plans.

Authors:  Ans C C Swinnen; Michel C Öllers; Erik Roijen; Sebastiaan M Nijsten; Frank Verhaegen
Journal:  J Appl Clin Med Phys       Date:  2017-01-03       Impact factor: 2.102

4.  Calibration of Gafchromic EBT Film Using the Microtek ScanMaker 9800XL Plus Flatbed Scanner with a Modified One Red-Channel after Three-Channel Method.

Authors:  Liyun Chang; Sheng-Yow Ho; Hueisch-Jy Ding; Shyh-An Yeh; Pang-Yu Chen
Journal:  J Med Phys       Date:  2019 Jul-Sep

5.  Dose calibration of Gafchromic EBT3 film for Ir-192 brachytherapy source using 3D-printed PLA and ABS plastics.

Authors:  Courtney Oare; Christopher Wilke; Eric Ehler; Damien Mathew; David Sterling; Clara Ferreira
Journal:  3D Print Med       Date:  2019-02-06

6.  Multichannel Film Dosimetry for Quality Assurance of Intensity Modulated Radiotherapy Treatment Plans Under 0.35 T Magnetic Field.

Authors:  Gorkem Gungor; Latif Korkmaz; Namik Kayalilar; Gokhan Aydin; Bulent Yapici; Teuta Zoto Mustafayev; Banu Atalar; Enis Ozyar
Journal:  Cureus       Date:  2020-03-20

7.  The potential of an optical surface tracking system in non-coplanar single isocenter treatments of multiple brain metastases.

Authors:  Ans C C Swinnen; Michel C Öllers; Chin Loon Ong; Frank Verhaegen
Journal:  J Appl Clin Med Phys       Date:  2020-04-01       Impact factor: 2.102

8.  Calibration of the EBT3 Gafchromic Film Using HNN Deep Learning.

Authors:  Liyun Chang; Shyh-An Yeh; Sheng-Yow Ho; Hueisch-Jy Ding; Pang-Yu Chen; Tsair-Fwu Lee
Journal:  Biomed Res Int       Date:  2021-01-31       Impact factor: 3.411