| Literature DB >> 25016166 |
Roberto Pontarolo1, Ana Carolina Gimenez2, Thais Martins Guimarães de Francisco2, Rômulo Pereira Ribeiro2, Flávia Lada Degaut Pontes2, João Cleverson Gasparetto2.
Abstract
The objective of this work was to develop and validate a HILIC-MS/MS method for the simultaneous determination of metformin and vildagliptin in human plasma. Chromatographic separation was achieved using an Atlantis HILIC Silica 150-mm × 2.1-mm, 3-μm particle size column maintained at 40°C. The isocratic mobile phase consisted of 20% water and 80% acetonitrile/water solution 95:5 (v/v), containing both 0.1% formic acid and 3mM ammonium formate. The flow rate was maintained at 400 μL min(-1). Data from validation studies demonstrated that the new method is highly selective, sensitive (limits of detection <1.5 ng mL(-1)) and free of matrix and residual effects. The new method was also precise (RSD<9.0%), accurate (RE<11.2%) and linear (r ≥ 0.99) over the ranges of 5-500 ng mL(-1) for each compound. The developed method was successfully applied to determine metformin and vildagliptin in plasma volunteers who orally received a single dose of metformin (850 mg), vildagliptin (50mg) or drug association (metformin 850 mg+vildagliptin 50mg). The new method can thus also be used as a tool for the clinical monitoring of metformin and vildagliptin.Entities:
Keywords: HILIC; Human plasma; LC–MS/MS; Metformin; Vildagliptin
Mesh:
Substances:
Year: 2014 PMID: 25016166 DOI: 10.1016/j.jchromb.2014.06.023
Source DB: PubMed Journal: J Chromatogr B Analyt Technol Biomed Life Sci ISSN: 1570-0232 Impact factor: 3.205