| Literature DB >> 25012356 |
Jonas Josef Peter Mark1, Paolo Piccinelli, Frank-Michael Matysik.
Abstract
A method for conducting fast and efficient capillary electrophoresis (CE) based on short separation capillaries in vertical alignment was developed. The strategy enables for high-throughput analysis from small sample vials (low microliter to nanoliter range). The system consists of a lab-made miniaturized autosampling unit and an amperometric end-column detection (AD) cell. The device enables a throughput of up to 200 separations per hour. CE-AD separations of a dye model system in capillaries of only 4 to 7.5 cm length with inner diameters (ID) of 10 or 15 μm were carried out under conditions of very high electric field strengths (up to 3.0 kV/cm) with high separation efficiency (half peak widths below 0.2 s) in less than 3.5 s migration time. A non-aqueous background electrolyte, consisting of 10 mM ammonium acetate and 1 M acetic acid in acetonitrile, was used. The practical suitability of the system was evaluated by applying it to the determination of dyes in overhead projector pens.Entities:
Mesh:
Substances:
Year: 2014 PMID: 25012356 DOI: 10.1007/s00216-014-7993-3
Source DB: PubMed Journal: Anal Bioanal Chem ISSN: 1618-2642 Impact factor: 4.142