Mechanical evaluation of four narrow-diameter implant systems.

Ronaldo Hirata, Estevam A Bonfante, Lucas S Machado, Nick Tovar, Paulo G Coelho.   

Abstract

PURPOSE: This study aimed to evaluate the survival probability of four narrow-diameter implant systems when subjected to fatigue loading.
MATERIALS AND METHODS: Seventy-two narrow-diameter implants to be restored with single-unit crowns were divided into four groups (n = 18): Astra Tech (3.5-mm diameter), with a standard connection (ASC); BioHorizon (3.4-mm diameter), with a standard connection (BSC); Intra-Lock (3.4-mm diameter), with a standard multilobular connection (ISC); and Intra-Lock (3.4-mm diameter), with a modified square connection (IMC). The corresponding abutments were screwed onto the implants, and standardized metal crowns (maxillary central incisors) were cemented and subjected to step-stress accelerated life testing in water. Use-level probability Weibull curves and reliability for 100,000 cycles at 150 and 200 N (90% two-sided confidence intervals) were calculated. Polarized light and scanning electron microscopes were used to access the failure modes.
RESULTS: The calculated survival probability for 100,000 cycles at 150 N was approximately 93% in group ASC, 98% in group BSC, 94% in group ISC, and 99% in group IMC. At 200 N, the survival rate was estimated to be approximately < 0.1% for ASC, 77% for BSC, 34% for ISC, and 93% for IMC. Abutment screw fracture was the main failure mode for all groups.
CONCLUSIONS: Although the probability of survival was not significantly different among systems at a load of 150 N, a significant decrease was observed at 200 N for all groups except IMC.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25010880     DOI: 10.11607/ijp.3926

Source DB:  PubMed          Journal:  Int J Prosthodont        ISSN: 0893-2174            Impact factor:   1.681


  2 in total

1.  Reliability and failure modes of narrow implant systems.

Authors:  Ronaldo Hirata; Estevam A Bonfante; Rodolfo B Anchieta; Lucas S Machado; Gileade Freitas; Vinicius P Fardin; Nick Tovar; Paulo G Coelho
Journal:  Clin Oral Investig       Date:  2015-11-03       Impact factor: 3.573

2.  Fracture and Fatigue of Titanium Narrow Dental Implants: New Trends in Order to Improve the Mechanical Response.

Authors:  Eugenio Velasco-Ortega; Antonio Flichy-Fernández; Miquel Punset; Alvaro Jiménez-Guerra; José María Manero; Javier Gil
Journal:  Materials (Basel)       Date:  2019-11-12       Impact factor: 3.623

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.