Effect of reduction of the pulse rates of fluoroscopy and CINE-acquisition on x-ray dose and angiographic image quality during invasive cardiovascular procedures.

Christopher T Pyne1, Gautam Gadey2, Cathy Jeon2, Thomas Piemonte2, Sergio Waxman2, Frederic Resnic2.   

Abstract

BACKGROUND: Reducing digital pulse rates (PR) are known to reduce total energy during invasive cardiovascular procedures, which likely has benefits for patients and staff. Physicians may be reluctant to reduce these parameters because they fear a decline in image quality that could affect procedural outcomes. We sought to assess the effect of default rates of fluoroscopy (Fluoro) and CINE-acquisition (CINE) on total x-ray dose and image quality during invasive cardiovascular procedures. METHODS AND
RESULTS: We retrospectively reviewed procedures done with 2 default PRs: a standard dose cohort (PR, 15 for Fluoro and CINE), and a reduced dose cohort (PR, 10 for Fluoro and CINE). Total x-ray dose, Fluoro time, and contrast use were compared between groups. A blinded angiographic image quality assessment was then performed using an objective 10-point angiographic quality score. There were no significant differences between cohorts for fluoroscopy time or contrast use. The reduced dose cohort has a significant reduction in mean total x-ray dose (PR 15, 1763.1 mGy; PR 10, 1179.1 mGy; P<0.0001). When adjusted for potential confounders, a 38% reduction in total x-ray dose was identified (P<0.0001). There was no difference in adjusted angiographic quality score between the cohorts (PR 15, 7.90; PR 10, 8.00; P=0.67), indicating no decline in image quality with PR reduction.
CONCLUSIONS: Reducing default PRs during invasive cardiovascular procedures yields large and significant reductions in total x-ray energy with no decline in angiographic image quality.
© 2014 American Heart Association, Inc.

Entities:  

Keywords:  angiography; fluoroscopy; image quality enhancement; x-rays

Mesh:

Year:  2014        PMID: 25006174     DOI: 10.1161/CIRCINTERVENTIONS.114.001479

Source DB:  PubMed          Journal:  Circ Cardiovasc Interv        ISSN: 1941-7640            Impact factor:   6.546


  5 in total

1.  Generative learning approach for radiation dose reduction in X-ray guided cardiac interventions.

Authors:  Fariba Azizmohammadi; Iñaki Navarro Castellanos; Joaquim Miró; Paul Segars; Ehsan Samei; Luc Duong
Journal:  Med Phys       Date:  2022-04-18       Impact factor: 4.506

2.  Diagnostic accuracy of quantitative flow ratio (QFR) and vessel fractional flow reserve (vFFR) estimated retrospectively by conventional radiation saving X-ray angiography.

Authors:  Chongying Jin; Anantharaman Ramasamy; Hannah Safi; Yakup Kilic; Vincenzo Tufaro; Retesh Bajaj; Guosheng Fu; Anthony Mathur; Christos V Bourantas; Andreas Baumbach
Journal:  Int J Cardiovasc Imaging       Date:  2021-01-16       Impact factor: 2.357

Review 3.  Interventional Radiology Suite: A Primer for Trainees.

Authors:  Bedros Taslakian; Ross Ingber; Eric Aaltonen; Jeremy Horn; Ryan Hickey
Journal:  J Clin Med       Date:  2019-08-30       Impact factor: 4.241

4.  Radiation Exposure Reduction and Patient Outcome by Using Very Low Frame Rate Fluoroscopy Protocol (3.8 + 7.5 fps) During Percutaneous Coronary Intervention.

Authors:  Ankush Gupta; Sanya Chhikara; Rajesh Vijayvergiya; Parag Barwad; Krishna Prasad; Rajat Datta; Nalin K Mahesh; Preetika Maurya; Navreet Singh
Journal:  Front Cardiovasc Med       Date:  2021-02-09

5.  A System for Real-Time, Online Mixed-Reality Visualization of Cardiac Magnetic Resonance Images.

Authors:  Dominique Franson; Andrew Dupuis; Vikas Gulani; Mark Griswold; Nicole Seiberlich
Journal:  J Imaging       Date:  2021-12-14
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.