Literature DB >> 24993502

Evaluation of the need of elective implantable cardioverter-defibrillator generator replacement in primary prevention patients without prior appropriate ICD therapy.

Sing-Chien Yap1, Beat A Schaer2, Rohit E Bhagwandien1, Michael Kühne2, Lara Dabiri Abkenari1, Stefan Osswald2, Tamas Szili-Torok1, Christian Sticherling2, Dominic A M J Theuns1.   

Abstract

OBJECTIVE: It is not clear whether patients who received an implantable cardioverter-defibrillator (ICD) for primary prevention should undergo device replacement if they never experienced an appropriate ICD therapy during the first generator longevity. This study evaluated the incidence and predictors of appropriate ICD therapy after device replacement in this specific population.
METHODS: From two large prospective ICD registries, we identified all primary prevention patients who had a first ICD replacement without previous appropriate ICD therapy. Cox regression analysis was used to identify predictors of appropriate ICD therapy.
RESULTS: Of 403 primary prevention patients needing first ICD replacement, 275 patients (68%) had not received previous appropriate ICD therapy. Patients without previous appropriate ICD therapy before first ICD replacement (mean age at replacement 62 ± 12 years, 75% male) had a mean follow-up of 86 ± 24 months after the initial implantation and 30 ± 24 months after device replacement. Following replacement, 3-year cumulative incidence of appropriate ICD therapy was 13.7% (95% CI 8.6 to 18.8%). No predictive factors associated with appropriate ICD therapy after replacement could be identified in spite of including seven clinically relevant factors.
CONCLUSIONS: A considerable number of primary prevention patients without previous appropriate ICD therapy before first ICD replacement received appropriate ICD therapy after replacement. As there were no predictors of appropriate ICD therapy after replacement, replacing an ICD is still recommended in all primary prevention patients despite the lack of appropriate ICD therapy during first battery service life. Published by the BMJ Publishing Group Limited. For permission to use (where not already granted under a licence) please go to http://group.bmj.com/group/rights-licensing/permissions.

Entities:  

Mesh:

Year:  2014        PMID: 24993502     DOI: 10.1136/heartjnl-2014-305535

Source DB:  PubMed          Journal:  Heart        ISSN: 1355-6037            Impact factor:   5.994


  8 in total

1.  Implantable cardioverter-defibrillator elective generator replacement: a procedure for all?

Authors:  Sergio Barra; Manoj Goonewardene; Patrick Heck; David Begley; Munmohan Virdee; Simon Fynn; Andrew Grace; Sharad Agarwal
Journal:  J Interv Card Electrophysiol       Date:  2016-01-12       Impact factor: 1.900

2.  Outcomes After Implantable Cardioverter-Defibrillator Generator Replacement for Primary Prevention of Sudden Cardiac Death.

Authors:  Malini Madhavan; Jonathan W Waks; Paul A Friedman; Daniel B Kramer; Alfred E Buxton; Peter A Noseworthy; Ramila A Mehta; David O Hodge; Angela Y Higgins; Tracy L Webster; Chance M Witt; Yong-Mei Cha; Bernard J Gersh
Journal:  Circ Arrhythm Electrophysiol       Date:  2016-03

Review 3.  When Is It Safe Not to Reimplant an Implantable Cardioverter Defibrillator at the Time of Battery Depletion?

Authors:  Sana M Al-Khatib; Daniel J Friedman; Gillian D Sanders
Journal:  Card Electrophysiol Clin       Date:  2018-03

4.  Association between ventricular arrhythmias and myocardial mechanical dispersion assessed by strain analysis in patients with nonischemic cardiomyopathy.

Authors:  Jedrzej Kosiuk; Borislav Dinov; Andreas Bollmann; Emmanuel Koutalas; Andreas Mussigbrodt; Phillipp Sommer; Arash Arya; Sergio Richter; Gerhard Hindricks; Ole A Breithardt
Journal:  Clin Res Cardiol       Date:  2015-05-30       Impact factor: 5.460

5.  Reassessment of cardiovascular parameters and comorbidities in implantable cardioverter-defibrillator patients at the time of first replacement.

Authors:  Sandra Wuest; Raphael Twerenbold; Michael Kühne; Tobias Reichlin; Christian Sticherling; Stefan Osswald; Beat Schaer
Journal:  Clin Cardiol       Date:  2018-01-22       Impact factor: 2.882

6.  Outcomes of patients with heart failure after primary prevention ICD unit generator replacement.

Authors:  Khang-Li Looi; Andrew Gavin; Lisa Cooper; Liane Dawson; Debbie Slipper; Nigel Lever
Journal:  Heart Asia       Date:  2019-02-28

7.  Exhausted atrial reserve by tissue Doppler echocardiography: a risk marker in heart failure with reduced ejection fraction.

Authors:  Otto A Smiseth
Journal:  Eur Heart J Cardiovasc Imaging       Date:  2016-05-04       Impact factor: 6.875

8.  Longevity of implantable cardioverter-defibrillators in a single-center population.

Authors:  Joachim Seegers; Pascal Muñoz Expósito; Lars Lüthje; Thomas Fischer; Matthias Lueken; Hannes Wenk; Samuel T Sossalla; Gerd Hasenfuss; Markus Zabel
Journal:  J Interv Card Electrophysiol       Date:  2015-08-08       Impact factor: 1.900

  8 in total