Comparison between ADVIA Chemistry systems Enzymatic Creatinine_2 method and ADVIA Chemistry systems Creatinine method (kinetic Jaffe method) for determining creatinine.

Fernando J Hermida1, María J Lorenzo, Aida Pérez, María Fernández, Olatz Sagastagoia, Concepción Magadán.   

Abstract

OBJECTIVE: The aim of this study was two-fold: Firstly, to compare the serum creatinine concentration measured by enzymatic method and uncompensated kinetic Jaffe method, and secondly, to compare the effects of certain interfering substances such as glucose, bilirubin, proteins, triglycerides and hemoglobin on creatinine measurement.
METHODS: The determination of serum creatinine concentrations by enzymatic method and uncompensated kinetic Jaffe method was performed on ADVIA® 2400 analyzer. The interfering substances were tested by adding solutions of interference to serum pool with low (62 μmol/L), medium (221 μmol/L) and elevated (486 μmol/L) creatinine concentration.
RESULTS: In the method comparison study, despite the fact that the correlation between both methods for determining serum creatinine is very good (r = 0.998, p < 0.001), the regression analysis revealed that the results are not transferable, as indicated by the slope and intercept, which are significantly different from 1 and 0, respectively. A positive bias of + 14.1% in the determination of serum creatinine by uncompensated kinetic Jaffe method was found, and when the creatinine value is lower than ∼ 180 μmol/L this difference or bis progressively increases. We found a significant positive interference due to proteins and glucose and a significant negative interference due to bilirubin by kinetic Jaffe method, and no interferences by enzymatic method were found.
CONCLUSIONS: In conclusion, the enzymatic method is the best choice for determining serum creatinine with the ADVIA® 2400 analyzer.

Entities:  

Keywords:  Creatinine; analytical performance; enzymatic method; interfering substances; kinetic Jaffe method

Mesh:

Substances:

Year:  2014        PMID: 24988313     DOI: 10.3109/00365513.2014.928943

Source DB:  PubMed          Journal:  Scand J Clin Lab Invest        ISSN: 0036-5513            Impact factor:   1.713


  3 in total

1.  Effect of differences in serum creatinine estimation methodologies on estimated glomerular filtration rate.

Authors:  Su-Chu Lee; Lee-Moay Lim; Eddy-Essen Chang; Yi-Wen Chiu; Shang-Jyh Hwang; Hung-Chun Chen
Journal:  Singapore Med J       Date:  2019-09       Impact factor: 1.858

2.  Analytical performance evaluation of different test systems on serum creatinine assay.

Authors:  Lina He; Jinqi Yu; Guang Han; Di Huang; Liqiao Han; Qiaoxuan Zhang; Yunxiu Wang; Zemin Wan; Xian-Zhang Huang; Yujuan Xiong; Xiaobin Wu
Journal:  J Clin Lab Anal       Date:  2021-12-26       Impact factor: 2.352

3.  A new method for anti-negative interference of calcium dobesilate in serum creatinine enzymatic analysis.

Authors:  Hailan Shen; Kena Chen; Ju Cao
Journal:  J Clin Lab Anal       Date:  2021-07-30       Impact factor: 2.352

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.