Fuminari Tatsugami1, Toru Higaki, Masao Kiguchi, So Tsushima, Akira Taniguchi, Yoko Kaichi, Takuji Yamagami, Kazuo Awai.
Abstract
We evaluated the accuracy of the electron densities and effective atomic numbers determined by raw data-based dual-energy analysis on a 320-detector computed tomography scanner. The mean (SD) errors between the measured and true electron densities and between the measured and true effective atomic numbers were 1.3% (1.5%) and 3.1% (3.2%), respectively. Electron densities and effective atomic numbers can be determined with high accuracy, which may help to improve accuracy in radiotherapy treatment planning.Mesh:
Year: 2014 PMID: 24983439 DOI: 10.1097/RCT.0000000000000129