Literature DB >> 24972358

Fiber-optic raster scanning two-photon endomicroscope using a tubular piezoelectric actuator.

Dukho Do1, Hongki Yoo2, Dae-Gab Gweon1.   

Abstract

A nonresonant, fiber-optic raster scanning endomicroscope was developed using a quarter-tubular piezoelectric (PZT) actuator. A fiber lever mechanism was utilized to enhance the small actuation range of the tubular PZT actuator and to increase its field-of-view. Finite element method simulation of the endoscopic probe was conducted for various conditions to maximize its scanning range. After fabricating the probe using a double clad fiber, we obtained two-photon fluorescence images using raster beam scanning of the fiber. The outer diameter of the probe was 3.5 mm and its rigid distal length was 30 mm including a high numerical aperture gradient index lens. These features are sufficient for input into the instrumental channel of a commercial colonoscope or gastroscope to obtain high resolution images in vivo.

Mesh:

Year:  2014        PMID: 24972358     DOI: 10.1117/1.JBO.19.6.066010

Source DB:  PubMed          Journal:  J Biomed Opt        ISSN: 1083-3668            Impact factor:   3.170


  7 in total

1.  Ultrahigh speed en face OCT capsule for endoscopic imaging.

Authors:  Kaicheng Liang; Giovanni Traverso; Hsiang-Chieh Lee; Osman Oguz Ahsen; Zhao Wang; Benjamin Potsaid; Michael Giacomelli; Vijaysekhar Jayaraman; Ross Barman; Alex Cable; Hiroshi Mashimo; Robert Langer; James G Fujimoto
Journal:  Biomed Opt Express       Date:  2015-03-05       Impact factor: 3.732

2.  Compact fiber-based multi-photon endoscope working at 1700 nm.

Authors:  Farhad Akhoundi; Yukun Qin; N Peyghambarian; Jennifer K Barton; Khanh Kieu
Journal:  Biomed Opt Express       Date:  2018-04-25       Impact factor: 3.732

3.  Development of tunable miniature piezoelectric-based scanners validated by the combination of two scanners in a direct image relay technique.

Authors:  Adam Harbi Shadfan; Michal Emanuel Pawlowski; Tomasz S Tkaczyk
Journal:  Opt Eng       Date:  2016-01-18

4.  Label-free highly multimodal nonlinear endoscope.

Authors:  D Septier; V Mytskaniuk; R Habert; D Labat; K Baudelle; A Cassez; G Brévalle-Wasilewski; M Conforti; G Bouwmans; H Rigneault; A Kudlinski
Journal:  Opt Express       Date:  2022-07-04       Impact factor: 3.833

5.  Lissajous Scanning Two-photon Endomicroscope for In vivo Tissue Imaging.

Authors:  Daniel Youngsuk Kim; Kyungmin Hwang; Jinhyo Ahn; Yeong-Hyeon Seo; Jae-Beom Kim; Soyoung Lee; Jin-Hui Yoon; Eunji Kong; Yong Jeong; Sangyong Jon; Pilhan Kim; Ki-Hun Jeong
Journal:  Sci Rep       Date:  2019-03-05       Impact factor: 4.379

6.  Frequency selection rule for high definition and high frame rate Lissajous scanning.

Authors:  Kyungmin Hwang; Yeong-Hyeon Seo; Jinhyo Ahn; Pilhan Kim; Ki-Hun Jeong
Journal:  Sci Rep       Date:  2017-10-26       Impact factor: 4.379

Review 7.  Single-Pixel MEMS Imaging Systems.

Authors:  Guangcan Zhou; Zi Heng Lim; Yi Qi; Guangya Zhou
Journal:  Micromachines (Basel)       Date:  2020-02-20       Impact factor: 2.891

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.