Lora Dukić1, Ana-Maria Simundić.
Abstract
INTRODUCTION: The aim of this survey study was to assess the current practices and policies in use related to the various steps in the blood gas testing process, across hospital laboratories in Croatia.Entities:
Mesh:
Year: 2014 PMID: 24969922 PMCID: PMC4083580 DOI: 10.11613/BM.2014.031
Source DB: PubMed Journal: Biochem Med (Zagreb) ISSN: 1330-0962 Impact factor: 2.313
ISO 15189 accreditation status of the participating laboratories
| Accredited, N (ratio) | 4 (0.09) |
| In the process of accreditation, N (ratio) | 2 (0.04) |
| Preparing for accreditation, N (ratio) | 18 (0.38) |
| Laboratory is not under preparation for accreditation, N (ratio) | 23 (0.49) |
Question: Sample type tested on analyzer. Ratio of participants is assigned to each category of statement. In brackets is assigned number of participants in relation to the total number of participants who answered to the statement by the quoted categories.
| 1. | Arterial sample is preferred sample type for blood gas determination. | 0.07 (3/44) | 0.23 (10/44) | 0.59 (26/44) | 0.11 (5/44) | 0.099 |
| 2. | Capillary blood gas analysis is performed for neonates and children. | 0.03 (1/35) | 0.03 (1/35) | 0.14 (5/35) | 0.80 (28/35) | 0.489 |
| 3. | Capillary blood gas analysis is performed for adults. | 0.12 (5/41) | 0.27 (11/41) | 0.46 (19/41) | 0.15 (6/41) | 1.000 |
| 4. | Capillary blood gas sampling is performed by laboratory technician. | 0.02 (1/41) | 0.10 (4/41) | 0.22 (9/41) | 0.66 (27/41) | 0.343 |
| 5. | Capillary blood gas sampling is performed by specialist in laboratory medicine. | 0.71 (29/41) | 0.20 (8/41) | 0.07 (3/41) | 0.02 (1/41) | 0.343 |
| 6. | Processing of the samples on the blood gas analyzer is performed by laboratory technician. | 0.09 (4/44) | 0.07 (3/44) | 0.45 (20/44) | 0.39 (17/44) | 0.416 |
| 7. | Processing of the samples on the blood gas analyzer is performed by specialist in laboratory medicine. | 0.25 (11/44) | 0.39 (17/44) | 0.23 (10/44) | 0.14 (6/44) | 0.532 |
Fisher’s exact test for the difference relative to accreditation status.
Laboratories which are not under preparation for accreditation were considered as one group, while laboratories which are either preparing for accreditation, are in the process of accreditation or are already accredited were grouped as another group. Replies in a category “never” and “rare” were grouped in one group and compared with replies in categories “often” and “always” (also grouped as one group).
Question: Sampling procedure. Ratio of participants is assigned to each category of statement. In brackets is assigned number of participants in relation to the total number of participants who answered to the statement by the quoted categories.
| 1. | Arterial blood gas samples are labelled with patient name. | 0.00 (0/41) | 0.02 (1/41) | 0.07 (3/41) | 0.90 (37/41) | 1.000 |
| 2. | Capillary blood gas samples are labelled with patient name. | 0.27 (11/41) | 0.05 (2/41) | 0.00 (0/41) | 0.68 (28/41) | 0.744 |
| 3. | Blood gas samples are followed by order form. | 0.03 (1/40) | 0.03 (1/40) | 0.13 (5/40) | 0.83 (33/40) | 1.000 |
| 4. | Blood gas samples get bar-code label on the ward. | 0.76 (28/37) | 0.03 (1/37) | 0.08 (3/37) | 0.14 (5/37) | 1.000 |
| 5. | Blood gas samples get bar-code label in the central laboratory. | 0.19 (8/42) | 0.10 (4/42) | 0.02 (1/42) | 0.69 (29/42) | 0.742 |
| 6. | Request orders for blood gas analysis are delivered electronically by hospital information system. | 0.20 (8/40) | 0.05 (2/40) | 0.05 (2/40) | 0.70 (28/40) | 1.000 |
| 7. | Arterial blood sampling is performed by the physician. | 0.03 (1/29) | 0.14 (4/29) | 0.48 (14/29) | 0.34 (10/29) | 0.622 |
| 8. | Arterial blood sampling is performed by the nurse. | 0.29 (8/28) | 0.54 (15/28) | 0.14 (4/28) | 0.04 (1/28) | 1.000 |
| 9. | Arterial blood is taken into the syringes which contain lyophilised anticoagulant without adding liquid heparin. | 0.16 (6/38) | 0.08 (3/38) | 0.13 (5/38) | 0.63 (24/38) | 1.000 |
| 10. | Arterial blood samples are visually inspected on presence of air bubbles after delivery to central laboratory. | 0.00 (0/41) | 0.00 (0/41) | 0.07 (3/41) | 0.93 (38/41) | N/A |
| 11. | Arterial blood samples are visually inspected on presence of clots after delivery to central laboratory. | 0.00 (0/41) | 0.00 (0/41) | 0.02 (1/41) | 0.98 (40/41) | N/A |
| 12. | After delivery of the arterial sample to central laboratory visual inspection of the integrity of the syringe is performed. | 0.02 (1/41) | 0.15 (6/41) | 0.02 (1/41) | 0.80 (33/41) | 0.679 |
| 13. | If the quality of the sample is not satisfactory, ward personnel is contacted and new blood sampling is requested. | 0.00 (0/41) | 0.07 (3/41) | 0.05 (2/41) | 0.88 (36/41) | 0.572 |
| 14. | Capillary blood samples for blood gas analysis are taken into capillary tubes containing lyophilised anticoagulant without adding of liquid heparin. | 0.00 (0/41) | 0.02 (1/41) | 0.05 (2/41) | 0.93 (38/41) | 0.488 |
| 15. | Before capillary blood gas sampling, arterialisation of the sampling area is performed. | 0.27 (10/37) | 0.38 (14/37) | 0.19 (7/37) | 0.16 (6/37) | 1.000 |
| 16. | During sampling procedure for capillary blood gas sample “milking” of sampling area is performed. | 0.24 (9/37) | 0.46 (17/37) | 0.27 (10/37) | 0.03 (1/37) | 0.728 |
| 17. | During sampling procedure air contamination is avoided. | 0.03 (1/40) | 0.00 (0/40) | 0.18 (7/40) | 0.80 (32/40) | 0.475 |
| 18. | Immediately after capillary blood gas sampling, capillary tube is tightly closed. | 0.05 (2/40) | 0.03 (1/40) | 0.03 (1/40) | 0.90 (36/40) | 1.000 |
| 19. | Capillary blood gas sample is mixed with anticoagulant using metal stirrer and magnet. | 0.08 (3/40) | 0.10 (4/40) | 0.03 (1/40) | 0.80 (32/40) | 1.000 |
| 20. | Time from sampling until delivery to central lab for arterial blood samples is recorded. | 0.36 (14/39) | 0.18 (7/39) | 0.08 (3/39) | 0.38 (15/39) | 1.000 |
| 21. | Time from delivery to central lab until sample processing on analyzer for arterial blood samples is recorded. | 0.15 (6/40) | 0.00 (0/40) | 0.03 (1/40) | 0.83 (33/40) | 1.000 |
| 22. | Time from sampling until delivery to central lab for capillary blood samples is recorded. | 0.37 (15/41) | 0.07 (3/41) | 0.02 (1/41) | 0.54 (22/41) | 0.756 |
| 23. | Time from delivery to central lab until sample processing on analyzer for capillary blood samples is recorded. | 0.18 (7/40) | 0.03 (1/40) | 0.00 (0/40) | 0.80 (32/40) | 0.698 |
Fisher’s exact test for the difference relative to accreditation status; N/A – not applicable - categories “never” and “rare” were too small for statistical analysis
Laboratories which are not under preparation for accreditation were considered as one group, while laboratories which not accreditation, are in the process of accreditation or are already accredited were grouped as another group. Replies in a category “never” and “rare” were grouped in one group and compared with replies in categories “often” and “always” (also grouped as one group).
Question: Blood gas sample analysis. Ratio of participants is assigned to each category of statement. In brackets is assigned number of participants in relation to the total number of participants who answered to the statement by the quoted categories.
| 1. | Except blood gas results, electrolyte results from the blood gas analyzer are reported. | 0.48 (21/44) | 0.52 (23/44) | 0.003 |
| 2. | Except blood gas results, metabolite (glucose, lactate) results from the blood gas analyzer are reported. | 0.26 (11/42) | 0.74 (31/42) | 0.004 |
| 3. | Internal quality control with the commercial quality control material is performed once a day on blood gas analyzer. | 0.57 (25/44) | 0.43 (19/44) | 0.076 |
| 4. | Internal quality control with the commercial quality control material is performed few times a day on blood gas analyzer. | 0.36 (16/44) | 0.64 (28/44) | 1.000 |
| 5. | Laboratory participates in CROQALM (national external quality control) program for blood gas analysis. | 0.98 (43/44) | 0.02 (1/44) | 1.000 |
| 6. | Laboratory participates in EQA (External Quality Assessment) quality control programme for blood gas analysis. | 0.16 (7/44) | 0.84 (37/44) | 1.000 |
| 7. | There are clear standard operating instructions for blood gas sampling in Croatian language available to laboratory personnel. | 0.91 (39/43) | 0.09 (4/43) | 1.000 |
| 8. | There are clear standard operating instructions for blood gas analyzer handling in Croatian language available to laboratory personnel. | 0.98 (43/44) | 0.02 (1/44) | 1.000 |
| 9. | All nonconformities in operation of the blood gas analyzer are recorded. | 0.91 (40/44) | 0.09 (4/44) | 0.335 |
| 10. | There is a record on education for blood gas sampling for each new employee. | 0.55 (23/42) | 0.45 (19/42) | 0.352 |
| 11. | There is a record on education for sample processing on blood gas analyzer for each new employee. | 0.59 (26/44) | 0.41 (18/44) | 0.065 |
Fisher’s exact test for the difference relative to accreditation status.
Laboratories which are not under preparation for accreditation were considered as one group, while laboratories which are either preparing for accreditation, are in the process of accreditation or are already accredited were grouped as another group. Replies in a category “never” and “rare” were grouped in one group and compared with replies in categories “often” and “always” (also grouped as one group).
Questions related to POCT analyzers. Ratio of participants is assigned to each category of statement. In brackets is assigned number of participants in relation to the total number of participants who answered to the statement by the quoted categories.
| 1. | Laboratory personnel trained ward personnel on sampling and processing of blood gas samples. | 0.55 (12/22) | 0.45 (10/22) | 0.172 |
| 2. | Laboratory issues certificate on training for sampling and blood gas analysis to the ward personnel. | 0.14 (3/22) | 0.86 (19/22) | 0.523 |
| 3. | Daily internal quality control is performed with commercial quality control material on POCT blood gas analyzer. | 0.30 (6/20) | 0.70 (14/20) | 1.000 |
| 4. | Laboratory personnel perform quality control on POCT analyzer. | 0.55 (11/20) | 0.45 (9/20) | 0.049 |
| 5. | Ward personnel performs quality control on POCT analyzer. | 0.40 (8/20) | 0.60 (12/20) | 1.000 |
| 6. | Are there standard operating instructions approved by laboratory on Croatian language for blood gas capillary sampling for POCT analyzer? | 0.53 (9/17) | 0.47 (8/17) | 0.049 |
| 7. | Are there standard operating instructions approved by laboratory on Croatian language for sample processing on POCT analyzer? | 0.53 (10/19) | 0.47 (9/19) | 0.057 |
| 8. | CROQALM external quality control for blood gases is performed for POCT analyzer. | 0.58 (11/19) | 0.42 (8/19) | 0.319 |
| 9. | EQA external quality control programme for blood gases is performed for POCT analyzer. | 0.16 (3/19) | 0.84 (16/19) | 1.000 |
| 10. | POCT analyzer maintenance is performed by laboratory personnel. | 0.45 (9/20) | 0.55 (11/20) | 0.157 |
Fisher’s exact test for the difference relative to accreditation status.
Laboratories which are not under preparation for accreditation were considered as one group, while laboratories are either preparing for accreditation, are in the process of accreditation or are already accredited were grouped as another group. Replies in a category “never” and “rare” were grouped in one group and compared with replies in categories “often” and “always” (also grouped as one group).
Average number of blood gas samples per day.
| 1–10 | 8 (0.24) |
| 11–20 | 7 (0.21) |
| 21–30 | 7 (0.21) |
| 31–40 | 2 (0.06) |
| 41–50 | 2 (0.06) |
| 70–80 | 1 (0.03) |
| 80 | 2 (0.06) |
Percentage of capillary blood specimens. In brackets is assigned ratio of laboratories in relation to the total number (N = 33).
| 0 | 7 (0.21) |
| 10 | 2 (0.06) |
| 20 | 2 (0.06) |
| 30 | 4 (0.12) |
| 50 | 6 (0.18) |
| 70 | 2 (0.06) |
| 75 | 2 (0.06) |
| 80 | 1 (0.03) |
| 85 | 1 (0.03) |
| 95 | 1 (0.03) |
| 100 | 5 (0.15) |