Impact of fiducial arrangement and registration sequence on target accuracy using a phantom frameless stereotactic navigation model.

Timothy R Smith1, Divakar S Mithal2, James A Stadler2, Camelia Asgarian2, Kenji Muro3, Joshua M Rosenow2.   

Abstract

Modern frameless stereotactic techniques utilize scalp fiducial markers for registration. Anecdotal reports from surgeons indicate a variety of methods for improving accuracy using different fiducial arrangements and registration sequences. The few published studies on registration accuracy do not provide a simple and systematic method for determining target accuracy. Nine different arrangements of ten fiducial markers were attached to a model. Ten separate markers were designated as targets for evaluation of registration accuracy. We systematically registered each of the arrangements over multiple trials, in one of four sequences, and then measured the targets. The target coordinates were compared against the established target values, and a root-mean-square deviation (RMSD) was derived. A systematic multivariate analysis determined the effects of different variables on the RMSD. We found no correlation between the "Registration Accuracy" provided by Medtronic (Medtronic Navigation, Louisville, CO, USA) and our RMSD representing targeting accuracy (R=0.008). RMSD did vary for different fiducial arrangements. We found no significant difference between the various sequences of fiducial arrangement. Thus, regardless of fiducial arrangement, registration sequence has no impact on accuracy. Fiducial arrangements distributed optimally across the skull, however, allowed for significantly improved accuracy. Further studies are required to determine which different arrangements of fiducials are relevant for specific procedures. Published by Elsevier Ltd.

Keywords:  Accuracy; Brain mapping; Frameless stereotaxy; Functional neurosurgery; Image-guided neurosurgery; Neuronavigation; Stereotactic surgery

Mesh:

Year:  2014        PMID: 24957630     DOI: 10.1016/j.jocn.2014.04.006

Source DB:  PubMed          Journal:  J Clin Neurosci        ISSN: 0967-5868            Impact factor:   1.961


  3 in total

1.  The comparative accuracy of the ROSA stereotactic robot across a wide range of clinical applications and registration techniques.

Authors:  Nicholas J Brandmeir; Sandip Savaliya; Pratik Rohatgi; Michael Sather
Journal:  J Robot Surg       Date:  2017-05-08

2.  How precise are oral splints for frameless stereotaxy in guided ear, nose, throat, and maxillofacial surgery: a cadaver study.

Authors:  Manfred Nilius; Minou Hélène Nilius
Journal:  Eur Radiol Exp       Date:  2021-06-30

3.  Stereotactic Neuro-Navigation Phantom Designs: A Systematic Review.

Authors:  Marko Švaco; Ivan Stiperski; Domagoj Dlaka; Filip Šuligoj; Bojan Jerbić; Darko Chudy; Marina Raguž
Journal:  Front Neurorobot       Date:  2020-10-23       Impact factor: 2.650

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.