| Literature DB >> 24932416 |
Sung Ho Hwang1, Hyun Hee Park2, Chung Sik Yoon3.
Abstract
BACKGROUND: The purpose of this study was to compare the concentration of total airborne bacteria (TAB) in biosafety cabinets (BSCs) at universities and hospital microbial laboratories to assess the performance of BSCs.Entities:
Keywords: biological safety cabinet; biosafety; microbial laboratory; total airborne bacteria
Year: 2014 PMID: 24932416 PMCID: PMC4048000 DOI: 10.1016/j.shaw.2014.01.001
Source DB: PubMed Journal: Saf Health Work ISSN: 2093-7911
Characteristics of the institutions and the biological safety cabinets (BSCs)
| Institution | Type of ventilation system | Size of the laboratory (m3) | Type of laboratory | Type of BSC (ID) | Manufacturing country |
|---|---|---|---|---|---|
| Hospital | HVAC | 444 | Diagnostic | Class II A2 (A) | USA |
| Hospital | HVAC | 810 | Diagnostic | Class II A2 (B-1) | USA |
| Class II A2 (B-2) | USA | ||||
| University | HVAC | 99 | Mouse | Class II A2 (I) | South Korea |
| University | General | 173 | Fish | Class II A2 (E) | South Korea |
| University | General | 528 | Microbial | Domestic BSC | South Korea |
| University | General | 75 | Microbial | Domestic BSC | South Korea |
| University | General | 372 | Microbial | Domestic BSC | South Korea |
| University | General | 399 | Microbial | Domestic BSC | South Korea |
| University | HVAC | 173 | Microbial | Class II A2 (H) | South Korea |
| University | HVAC | 49 | Microbial | Class II A2 (G) | South Korea |
HEPA, high efficiency particulate air; HVAC, heating, ventilating, and air conditioning.
Domestically identified as a type of BSC, but down-flow velocity and the efficiency of the HEPA filters could not be tested because of the faulty BSC design.
Concentrations of accredited and nonaccredited TAB facilities at different institutions when sampling was performed with the biosafety cabinets (BSCs) turned on or off
| Institution | BSC | TAB concentration (CFU/m3) | M/B ratio | Accreditation | ||||
|---|---|---|---|---|---|---|---|---|
| BSC turned off | BSC turned on | Mean (SD) | Background | |||||
| Open | Close | |||||||
| Hospital | A | 198 | 138 | 57 | 131 (70.8) | 86 | 1.5 | Yes (NSF) |
| University | E | 702 | 543 | 205 | 483 (253.8) | NA | NA | Yes (KS) |
| I | ND | ND | ND | ND | ND | ND | Yes (KS) | |
| Hospital | B-1 | 175 | 213 | 220 | 203 (24.2) | 115 | 1.8 | No |
| B-2 | 93 | 86 | 93 | 91 (4.0) | 0.8 | No | ||
| University | C | 182 | 93 | 266 | 180 (86.5) | 205 | 0.9 | No |
| F | 93 | 123 | 337 | 184 (133.1) | 153 | 1.2 | No | |
| G | 21 | 28 | 93 | 47 (39.7) | 14 | 3.4 | No | |
| D | 50 | 337 | 266 | 218 (149.5) | 386 | 0.6 | No | |
| H | 108 | 115 | 71 | 98 (23.6) | 108 | 0.9 | No | |
| J | 153 | 220 | 220 | 198 (38.7) | 251 | 0.8 | No | |
| Mean (SD) | 164 | 182 | 182 | 176 (10.4) | 146 | 1.2 | ||
NA, not applicable; ND, not determined; SD, standard deviation; TAB; total airborne bacteria.
The ratio of mean TAB concentrations for areas in front of BSCs/TAB concentrations for the background areas.
Fig. 1Example of the faulty design of biosafety cabinets due to the fluorescent light, which was disturbing the tests for the downflow velocity and efficiency of the HEPA filters. HEPA, high efficiency particulate air.