Literature DB >> 24924280

The Handy Eye Check: a mobile medical application to test visual acuity in children.

Keri N Toner1, Michael J Lynn2, T Rowan Candy3, Amy K Hutchinson4.   

Abstract

PURPOSE: To compare visual acuity results obtained with the Handy Eye Chart to results obtained using the Handy Eye Check, a mobile medical application that electronically presents isolated Handy Eye Chart optotypes according the Amblyopia Treatment Study (ATS) protocol.
METHODS: Consecutive patients 6-18 years of age presenting for eye examinations between May 30, 2012, and June 26, 2012, were invited to participate. Monocular visual acuity testing was performed on the subject's poorer-seeing eye using both the Handy Eye Check and the Handy Eye Chart under the same conditions. Visual acuity was first tested using the mobile application, then using the chart, followed by repeated application testing. Patients were excluded if they were unable to undergo the required visual acuity testing or if visual acuity in the worse-seeing eye was less than 20/200 (for validity testing, but not reliability testing).
RESULTS: There was a strong linear correlation (r = 0.92) and a mean difference in acuity of -0.005 logMAR, or less than one letter (95% CI, -0.03 to 0.02), between the two tests. The 95% limits of agreement were ± 2 lines. Test-retest reliability was high, with 81% of retest scores within 0.1 logMAR (5 letters) and 100% within 0.2 logMAR (10 letters), an intraclass correlation coefficient of 0.93, and a standard error of measurement of 0.08.
CONCLUSIONS: The Handy Eye Check mobile application compares similarly to the Handy Eye Chart as a valid and reliable test of visual acuity in children age 6-18 years. Published by Mosby, Inc.

Entities:  

Mesh:

Year:  2014        PMID: 24924280     DOI: 10.1016/j.jaapos.2014.01.011

Source DB:  PubMed          Journal:  J AAPOS        ISSN: 1091-8531            Impact factor:   1.220


  3 in total

1.  A new visual acuity test on touchpad for vision screening in children.

Authors:  Elsa Di Foggia; Noémie Stoll; Hélène Meunier; Adam Rimelé; Pascal Ance; Pierre-Henri Moreau; Claude Speeg-Schatz; Arnaud Sauer
Journal:  Int J Ophthalmol       Date:  2020-09-18       Impact factor: 1.779

2.  Comparing distance visual acuity measurement with a novel eye chart and the Landolt C chart in a population of children aged 6-18 years.

Authors:  John P Gorham; Beau B Bruce; Amy K Hutchinson
Journal:  Int Ophthalmol       Date:  2016-11-08       Impact factor: 2.031

3.  Undermining a common language: smartphone applications for eye emergencies.

Authors:  Jennifer M Charlesworth; Myriam A Davidson
Journal:  Med Devices (Auckl)       Date:  2019-01-15
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.